A Monte Carlo study on the PTW 60019 microDiamond detector

被引:12
作者
Hartmann, Guenther H. [1 ]
Zink, Klemens [2 ,3 ,4 ]
机构
[1] German Canc Res Ctr, D-69120 Heidelberg, Germany
[2] Univ Appl Sci Giessen, Inst Med Phys & Radiat Protect IMPS, D-35390 Giessen, Germany
[3] Univ Med Ctr Giessen Marburg, Dept Radiotherapy & Radiooncol, D-35043 Marburg, Germany
[4] Goethe Univ, FIAS, D-60438 Frankfurt, Germany
关键词
microdiamond detector; Monte Carlo calculation; output correction factors; perturbation factors; small beam dosimetry; QUALITY CORRECTION FACTORS; THIMBLE IONIZATION CHAMBERS; SYNTHETIC MICRODIAMOND; DIAMOND DETECTORS; DOSE-RESPONSE; ION CHAMBERS; DOSIMETRY; PERTURBATION; DIODE; K(Q);
D O I
10.1002/mp.13721
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Data on the output correction factor for small photon beam dosimetry of the microDiamond detector manufactured by the company PTW can be found in a variety of papers. Referring either to measurements or to Monte Carlo (MC) calculations, they show substantial disagreements particularly at very small fields. This work reports results of a further MC study aiming at a better understanding of how specific properties of the microDiamond detector are influencing its output correction factor and whether this can explain at least some of the disagreements. Methods In this study the method of a fluence-based decomposition of the dose conversion factor was used which is considered as a useful tool to understand the response of a detector in nonreference conditions. This decomposition method yields the following three factors: (a) the stopping power ratio water to diamond, (b) a perturbation factor p(int) taking into account all fluence changes in the transition from a small water voxel at the point of dose determination to the bare diamond detector, and (c) a perturbation factor p(ext) taking into account all additional fluence changes in the fully simulated diamond detector caused by the material and design details outside the sensitive volume. Results Monte Carlo calculated output correction factors were obtained for Co-60, 6 MV and 10 MV photon beams showing that the maximum variation with field size remained in the order of 2% for quadratic field sizes larger than about 0.3 cm. For field sizes smaller than about 0.5 cm a clear under-response is obtained at all three radiation qualities in agreement with all known MC calculations, however, in contrast to some measured result. The shape of the output correction factor can be well explained by an opposite mode of action between under-response expressed by the perturbation factor p(int) and over-response expressed by the perturbation factor p(ext) where the first one is mainly influenced by volume averaging, and the second one by a back scatter effect of electrons from the diamond substrate into the sensitive volume. Conclusion The response of microDiamond detector can be well described under various measuring conditions by the dose conversion factor and the dependency of its fluence-based subfactors on detector characteristics. Monte Carlo simulations offer an improvement in the understanding particularly of small-field effects by relating the output correction factor to spectral fluence changes in the sensitive volume of the detector. The most significant influence factors are the finite size of the active volume and the presence of the high-density diamond substrate causing a field size-dependent backscattering. These perturbations are opposite in their effects. The diamond in the sensitive volume itself and in particular its density has almost no influence. Scattering of results at very small field sizes can be explained by different gradients of dose profiles around the beam axis at identical full width half maximum (FWHM) field size parameters and by possible deviations of the radius of the sensitive volume from the nominal radius. The backscattering effect also has an influence on the determination of profiles and for very small field sizes on the response at different rotation angles.
引用
收藏
页码:5159 / 5172
页数:14
相关论文
共 50 条
  • [41] Monte Carlo study of magnetic behavior of core-shell nanoribbon
    Drissi, L. B.
    Zriouel, S.
    Bahmad, L.
    [J]. JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2015, 374 : 639 - 646
  • [42] A feasibility study for in vivo treatment verification of IMRT using Monte Carlo dose calculation and deep learning-based modelling of EPID detector response
    Zhang, Jun
    Cheng, Zhibiao
    Fan, Ziting
    Zhang, Qilin
    Zhang, Xile
    Yang, Ruijie
    Wen, Junhai
    [J]. RADIATION ONCOLOGY, 2022, 17 (01)
  • [43] A Monte Carlo study to evaluate and optimise the angular dependence of the Octa-A 2D silicon array detector used for dosimetry in stereotactic radiotherapy
    Filipev, Ilia
    Biasi, Giordano
    Poder, Joel
    Lerch, Michael
    Petasecca, Marco
    Guatelli, Susanna
    Kron, Tomas
    Rosenfeld, Anatoly
    [J]. RADIATION MEASUREMENTS, 2023, 166
  • [44] Measured and Monte Carlo calculated kQ factors: Accuracy and comparison
    Muir, B. R.
    McEwen, M. R.
    Rogers, D. W. O.
    [J]. MEDICAL PHYSICS, 2011, 38 (08) : 4600 - 4609
  • [45] A Fano cavity test for Monte Carlo proton transport algorithms
    Sterpin, Edmond
    Sorriaux, Jefferson
    Souris, Kevin
    Vynckier, Stefaan
    Bouchard, Hugo
    [J]. MEDICAL PHYSICS, 2014, 41 (01)
  • [46] Determination of the dose rate around a HDR 192Ir brachytherapy source with the microDiamond and the microSilicon detector
    Rossi, Giulio
    Failing, Thomas
    Gainey, Mark
    Kollefrath, Michael
    Hensley, Frank
    Zink, Klemens
    Baltas, Dimos
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (04): : 463 - 478
  • [47] Determination of the KQclin,Qmsrfclin,fmsr correction factors for detectors used with an 800 MU/min CyberKnife® system equipped with fixed collimators and a study of detector response to small photon beams using a Monte Carlo method
    Moignier, C.
    Huet, C.
    Makovicka, L.
    [J]. MEDICAL PHYSICS, 2014, 41 (07)
  • [48] Monte Carlo Calculations for Radiosurgery of the Clivus
    Ho, A.
    Gibbs, I.
    Chang, S.
    Soltys, S.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 277 - 278
  • [49] A Monte Carlo based radiotherapy simulator
    Bliznakova, K
    Kolitsi, Z
    Pallikarakis, N
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2004, 222 (3-4) : 445 - 461
  • [50] Differential sampling for the Monte Carlo practitioner
    Peplow, DE
    Verghese, K
    [J]. PROGRESS IN NUCLEAR ENERGY, 2000, 36 (01) : 39 - 75