A Monte Carlo study on the PTW 60019 microDiamond detector

被引:12
|
作者
Hartmann, Guenther H. [1 ]
Zink, Klemens [2 ,3 ,4 ]
机构
[1] German Canc Res Ctr, D-69120 Heidelberg, Germany
[2] Univ Appl Sci Giessen, Inst Med Phys & Radiat Protect IMPS, D-35390 Giessen, Germany
[3] Univ Med Ctr Giessen Marburg, Dept Radiotherapy & Radiooncol, D-35043 Marburg, Germany
[4] Goethe Univ, FIAS, D-60438 Frankfurt, Germany
关键词
microdiamond detector; Monte Carlo calculation; output correction factors; perturbation factors; small beam dosimetry; QUALITY CORRECTION FACTORS; THIMBLE IONIZATION CHAMBERS; SYNTHETIC MICRODIAMOND; DIAMOND DETECTORS; DOSE-RESPONSE; ION CHAMBERS; DOSIMETRY; PERTURBATION; DIODE; K(Q);
D O I
10.1002/mp.13721
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Data on the output correction factor for small photon beam dosimetry of the microDiamond detector manufactured by the company PTW can be found in a variety of papers. Referring either to measurements or to Monte Carlo (MC) calculations, they show substantial disagreements particularly at very small fields. This work reports results of a further MC study aiming at a better understanding of how specific properties of the microDiamond detector are influencing its output correction factor and whether this can explain at least some of the disagreements. Methods In this study the method of a fluence-based decomposition of the dose conversion factor was used which is considered as a useful tool to understand the response of a detector in nonreference conditions. This decomposition method yields the following three factors: (a) the stopping power ratio water to diamond, (b) a perturbation factor p(int) taking into account all fluence changes in the transition from a small water voxel at the point of dose determination to the bare diamond detector, and (c) a perturbation factor p(ext) taking into account all additional fluence changes in the fully simulated diamond detector caused by the material and design details outside the sensitive volume. Results Monte Carlo calculated output correction factors were obtained for Co-60, 6 MV and 10 MV photon beams showing that the maximum variation with field size remained in the order of 2% for quadratic field sizes larger than about 0.3 cm. For field sizes smaller than about 0.5 cm a clear under-response is obtained at all three radiation qualities in agreement with all known MC calculations, however, in contrast to some measured result. The shape of the output correction factor can be well explained by an opposite mode of action between under-response expressed by the perturbation factor p(int) and over-response expressed by the perturbation factor p(ext) where the first one is mainly influenced by volume averaging, and the second one by a back scatter effect of electrons from the diamond substrate into the sensitive volume. Conclusion The response of microDiamond detector can be well described under various measuring conditions by the dose conversion factor and the dependency of its fluence-based subfactors on detector characteristics. Monte Carlo simulations offer an improvement in the understanding particularly of small-field effects by relating the output correction factor to spectral fluence changes in the sensitive volume of the detector. The most significant influence factors are the finite size of the active volume and the presence of the high-density diamond substrate causing a field size-dependent backscattering. These perturbations are opposite in their effects. The diamond in the sensitive volume itself and in particular its density has almost no influence. Scattering of results at very small field sizes can be explained by different gradients of dose profiles around the beam axis at identical full width half maximum (FWHM) field size parameters and by possible deviations of the radius of the sensitive volume from the nominal radius. The backscattering effect also has an influence on the determination of profiles and for very small field sizes on the response at different rotation angles.
引用
收藏
页码:5159 / 5172
页数:14
相关论文
共 50 条
  • [1] Performance of a PTW 60019 microDiamond detector in a 1.5 T MRI-linac
    Woodings, S. J.
    Wolthaus, J. W. H.
    van Asselen, B.
    de Vries, J. H. W.
    Kok, J. G. M.
    Lagendijk, J. J. W.
    Raaymakers, B. W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (05):
  • [2] Experimental determination of the PTW 60019 microDiamond dosimeter active area and volume
    Marinelli, Marco
    Prestopino, G.
    Verona, C.
    Verona-Rinati, G.
    MEDICAL PHYSICS, 2016, 43 (09) : 5205 - 5212
  • [3] Dose rate dependence of the PTW 60019 microDiamond detector in high dose-per-pulse pulsed beams
    Brualla-Gonzalez, Luis
    Gomez, Faustino
    Pombar, Miguel
    Pardo-Montero, Juan
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (01): : N11 - N19
  • [4] Under-response of a PTW-60019 microDiamond detector in the Bragg peak of a 62 MeV/n carbon ion beam
    Rossomme, S.
    Hopfgartner, J.
    Vynckier, S.
    Palmans, H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (12): : 4551 - 4563
  • [5] Characterizing a PTW microDiamond detector in kilovoltage radiation beams
    Khan, Ahtesham Ullah
    Culberson, Wesley S.
    DeWerd, Larry A.
    MEDICAL PHYSICS, 2020, 47 (09) : 4553 - 4562
  • [6] The dose response of PTW microDiamond and microSilicon in transverse magnetic field under small field conditions
    Blum, Isabel
    Tekin, Tuba
    Delfs, Bjorn
    Schonfeld, Ann-Britt
    Kapsch, Ralf-Peter
    Poppe, Bjorn
    Looe, Hui Khee
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (15):
  • [7] Properties of a commercial PTW-60019 synthetic diamond detector for the dosimetry of small radiotherapy beams
    Manuel Larraga-Gutierrez, Jose
    Ballesteros-Zebadua, Paola
    Rodriguez-Ponce, Miguel
    Amanda Garcia-Garduno, Olivia
    Galvan de la Cruz, Olga Olinca
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (02): : 905 - 924
  • [8] Use of PTW-microDiamond for relative dosimetry of unflattened photon beams
    Reggiori, Giacomo
    Stravato, Antonella
    Pimpinella, Maria
    Lobefalo, Francesca
    De Coste, Vanessa
    Fogliata, Antonella
    Mancosu, Pietro
    De Rose, Fiorenza
    Palumbo, Valentina
    Scorsetti, Marta
    Tomatis, Stefano
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 38 : 45 - 53
  • [9] An EGS4 Monte Carlo examination of the response of a PTW-diamond radiation detector in megavoltage electron beams
    Mobit, PN
    Sandison, GA
    MEDICAL PHYSICS, 1999, 26 (05) : 839 - 844
  • [10] Determination of output factor for 6 MV small photon beam: comparison between Monte Carlo simulation technique and microDiamond detector
    Krongkietlearts, K.
    Tangboonduangjit, P.
    Paisangittisakul, N.
    13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694