Intraocular lens power calculation for plus and minus lenses in high myopia using partial coherence interferometry

被引:7
作者
Fuest, Matthias [1 ]
Plange, Niklas [1 ]
Kuerten, David [1 ]
Schellhase, Hannah [1 ]
Mazinani, Babac A. E. [1 ]
Walter, Peter [1 ]
Kohnen, Stephan [2 ]
Widder, Randolf A. [3 ,4 ]
Roessler, Gernot [1 ,3 ]
机构
[1] Rhein Westfal TH Aachen, Dept Ophthalmol, Pauwelsstr 30, D-52074 Aachen, Germany
[2] ACD Augen Ctr Dreilandereck, Aachen, Germany
[3] St Martinus Krankenhaus Dusseldorf, Dept Ophthalmol, Dusseldorf, Germany
[4] Univ Cologne, Dept Ophthalmol, Cologne, Germany
关键词
Cataract; Biometry; Pathologic; Myopia; IOL; Calculation;
D O I
10.1007/s10792-020-01684-y
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose We assessed the accuracy of lens power calculation in highly myopic patients implanting plus and minus intraocular lenses (IOL). Methods We included 58 consecutive, myopic eyes with an axial length (AL) > 26.0 mm, undergoing phacoemulsification and IOL implantation following biometry using the IOLMaster 500. For lens power calculation, the Haigis formula was used in all cases. For comparison, refraction was back-calculated using the Barrett Universal II (Barrett), Holladay I, Hill-RBF (RBF) and SRK/T formulae. Results The mean axial length was 30.17 +/- 2.67 mm. Barrett (80%), Haigis (87%) and RBF (82%) showed comparable numbers of IOLs within 1 diopter (D) of target refraction. Visual acuity (BSCVA) improved (p < 0.001) from 0.60 +/- 0.35 to 0.29 +/- 0.29 logMAR (> 28-days postsurgery). The median absolute error (MedAE) of Barrett 0.49 D, Haigis 0.38, RBF 0.44 and SRK/T 0.44 did not differ. The MedAE of Haigis was significantly smaller than Holladay (0.75 D; p = 0.01). All median postoperative refractive errors (MedRE) differed significantly with the exception of Haigis to SRK/T (p = 0.6): Barrett - 0.33 D, Haigis 0.25, Holladay 0.63, RBF 0.04 and SRK/T 0.13. Barrett, Haigis, Holladay and RBF showed a tendency for higher MedAEs in their minus compared to plus IOLs, which only reached significance for SRK/T (p = 0.001). Barrett (p < 0.001) and RBF (p = 0.04) showed myopic, SRK/T (p = 002) a hyperopic shift in their minus IOLs. Conclusions In highly myopic patients, the accuracies of Barrett, Haigis and RBF were comparable with a tendency for higher MedAEs in minus IOLs. Barrett and RBF showed myopic, SRK/T a hyperopic shift in their minus IOLs.
引用
收藏
页码:1585 / 1592
页数:8
相关论文
共 22 条
  • [1] Intraocular lens power calculation for eyes with an axial length greater than 26.0 mm: Comparison of formulas and methods
    Abulafia, Adi
    Barrett, Graham D.
    Rotenberg, Michael
    Kleinmann, Guy
    Levy, Adi
    Reitblat, Olga
    Koch, Douglas D.
    Wang, Li
    Assia, Ehud I.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (03) : 548 - 556
  • [3] Accuracy of Intraocular Lens Power Formulas Involving 148 Eyes with Long Axial Lengths: A Retrospective Chart-Review Study
    Chen, Chong
    Xu, Xian
    Miao, Yuyu
    Zheng, Gaoxin
    Sun, Yong
    Xu, Xun
    [J]. JOURNAL OF OPHTHALMOLOGY, 2015, 2015
  • [4] Comparison of the Kane formula with existing formulas for intraocular lens power selection
    Connell, Benjamin J.
    Kane, Jack X.
    [J]. BMJ OPEN OPHTHALMOLOGY, 2019, 4 (01):
  • [5] AXIAL LENGTH MEASUREMENTS AND FUNDUS CHANGES OF MYOPIC EYE
    CURTIN, BJ
    KARLIN, DB
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1971, 71 (01) : 42 - &
  • [6] Improved prediction of intraocular lens power using partial coherence interferometry
    Findl, O
    Drexler, W
    Menapace, R
    Heinzl, H
    Hitzenberger, CK
    Fercher, AF
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (06) : 861 - 867
  • [7] Intraocular lens calculation in extreme myopia
    Haigis, Wolfgang
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (05) : 906 - 911
  • [8] Protocols for Studies of Intraocular Lens Formula Accuracy
    Hoffer, Kenneth J.
    Aramberri, Jaime
    Haigis, Wolfgang
    Olsen, Thomas
    Savini, Giacomo
    Shammas, H. John
    Bentow, Stanley
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2015, 160 (03) : 403 - 405
  • [9] First results of cataract surgery and implantation of negative power intraocular lenses in highly myopic eyes
    Kohnen, S
    Brauweiler, P
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 (04) : 416 - 420
  • [10] Complications of foldable intraocular lenses requiring explantation or secondary intervention - 1998 survey
    Mamalis, N
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (05) : 766 - 772