Prevention of serious air embolism during cryoballoon ablation; risk assessment of air intrusion into the sheath by catheter selection and change in intrathoracic pressure: An ex vivo study

被引:7
|
作者
Tsukahara, Keita [1 ]
Oginosawa, Yasushi [1 ]
Fujino, Yoshihisa [2 ]
Ohe, Hisaharu [1 ]
Yamagishi, Yasunobu [1 ]
Iwataki, Mai [1 ]
Sonoda, Shinjo [1 ]
Kohno, Ritsuko [3 ]
Otsuji, Yutaka [1 ]
Abe, Haruhiko [3 ]
机构
[1] Univ Occupat & Environm Hlth, Dept Internal Med 2, Kitakyushu, Japan
[2] Univ Occupat & Environm Hlth, Inst Ind Ecol Sci, Dept Environm Epidemiol, Kitakyushu, Japan
[3] Univ Occupat & Environm Hlth, Div Heart Rhythm Management, Kitakyushu, Japan
基金
日本学术振兴会;
关键词
air embolism; air intrusion; atrial fibrillation; cryoballoon ablation; obstructive sleep apnea syndrome; PAROXYSMAL ATRIAL-FIBRILLATION; RADIOFREQUENCY ABLATION;
D O I
10.1111/jce.14208
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction One cause of cerebral infarction during cryoballoon ablation is the entry of air into a sheath due to the use of inappropriate catheters. It is known that the left atrial pressure of patients with obstructive sleep apnea syndrome can be negative. However, the effects of catheter selection and negative pressure changes in the sheath on air intrusion are not yet well understood. The aim of this study was to evaluate how catheter selection and negative pressure changes affect air intrusion and to perform countermeasures for air intrusion. Methods and Results This experiment used siphon principle to create negative pressure in the sheath. Noncryoablation catheters (not designed exclusively for cryoballoon ablation) and cryoballoon catheters were investigated. Catheters were inserted into the sheath and then removed. Thereafter, the amount of air in the sheath was measured. For catheters producing significantly larger amounts of air intrusion, the catheters were inserted via a long sheath in the sheath (sheath-in-sheath technique) and the same procedures were repeated. We found that the amount of air intrusion through most of the noncryoablation catheters was significantly larger than that through cryoablation catheters. An increase in the magnitude of negative pressure in the sheath resulted in a proportional increase in air intrusion, but the sheath-in-sheath technique significantly reduced air intrusion. Conclusion The amount of air intrusion increased when using catheters with complicated tip shapes and thin outer diameters and when the magnitude of negative pressure in the sheath increased. The sheath-in-sheath technique may be an effective countermeasure.
引用
收藏
页码:2944 / 2949
页数:6
相关论文
共 1 条
  • [1] Techniques for reducing air bubble intrusion into the left atrium during radiofrequency catheter and cryoballoon ablation procedures: An ex vivo study with a high-resolution camera
    Takami, Mitsuru
    Fujiwara, Ryudo
    Kijima, Yoichi
    Nagoshi, Ryoji
    Kozuki, Amane
    Mochizuki, Yasuhide
    Shibata, Hiroyuki
    Nakano, Shinsuke
    Fukuyama, Yusuke
    Fujimoto, Daichi
    Kakizaki, Shunsuke
    Masuko, Eri
    Kurimoto, Hiroyuki
    Kyo, Shokan
    Miyata, Tomohiro
    Shite, Junya