Development and validation of a quantitative determination method of blood ethanol by gas chromatography with headspace (GC-HS)

被引:8
作者
Dorubet, Diana [1 ,2 ]
Moldoveanu, Sorin [2 ]
Mircea, Cornelia [3 ]
Butnaru, Elena [2 ]
Astarastoae, Vasile [4 ]
机构
[1] Vaslui Cty Dept Forens Med, Toxicol Lab, Vaslui, Romania
[2] Gr T Popa Univ Med & Pharm Iasi, Sch Pharm, Dept Toxicol, Iasi, Romania
[3] Gr T Popa Univ Med & Pharm Iasi, Sch Pharm, Dept Gen & Appl Biochem, Iasi, Romania
[4] Gr T Popa Univ Med & Pharm Iasi, Sch Med, Dept Forens Med, Iasi, Romania
来源
ROMANIAN JOURNAL OF LEGAL MEDICINE | 2009年 / 17卷 / 04期
关键词
blood; ethanol; GC-HS; validation; ALCOHOL DETERMINATION;
D O I
10.4323/rjlm.2009.303
中图分类号
DF [法律]; D9 [法律]; R [医药、卫生];
学科分类号
0301 ; 10 ;
摘要
Quantitative determination of ethanol from blood is used for scientific research, for evaluation of ethanol metabolism in the human body or in the forensic laboratory. The aim of our paper is to develop and validate a method in order to determine the quantity of blood ethanol using gas chromatography with headspace. We used a GC 6890 N Agilent equipment, with headspace G 1888, two columns for alcohol analysis (DB-ALC1 and DB-ALC2) and FID detectors. The retention tune has been higher when using column DB-ALC1 (1.362 min) rather than for DB-ALC2 column (1.313 min). Internal standard for this method was isopropanol. The developed method for ethanol is specific because the retention time is different from the internal standard. Our method is linear in the range of 0.5-4 g/L. The correlation coefficients were 0.99958 for column DB-ALC1, respectively 0.99953 for column DB-ALC2. Using standard solutions like samples, we determined method precision and the recovery in percentage from theoretical concentrations. The recovery percents were between 98.89% (theoretical concentration 3 g ethanol/L) and 107.70% (theoretical concentration 0.5 g ethanol/L). In order to establish the influence of blood matrix, blood samples with known concentration of alcohol and internal standards, have been evaluated. In this case, the values for RSD were between 0.5 and 3.5% in direct correlation with sample homogeneity. The detection limits were 0.055 g/L (DB-ALC2) and 0.053 g/L (DB-ALC2). In the same conditions, the quantification limits were 0.186 g/L (DB-ALC2) and 0.177 g/L (DB-ALC2). The developed and validated method, in the mentioned conditions, is linear, precise and accurate and could be used for ethanol determination.
引用
收藏
页码:303 / 308
页数:6
相关论文
共 50 条
  • [31] Rapid Simultaneous Determination of Cyanide, Azide, and Ethanol in Whole Blood Using Headspace Gas Chromatography-Mass Spectrometry
    Murakami, Takaya
    Iwamuro, Yoshiaki
    Sakamoto, Yuki
    Minami, Eriko
    Ishimaru, Reiko
    Tsuchihashi, Hitoshi
    Chinaka, Satoshi
    CHROMATOGRAPHIA, 2023, 86 (11-12) : 701 - 706
  • [32] Development and validation of a GC/MS method for the simultaneous determination of levetiracetam and lamotrigine in whole blood
    Nikolaou, Panagiota
    Papoutsis, Loannis
    Dona, Artemisia
    Spiliopoulou, Chara
    Athanaselis, Sotiris
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2015, 102 : 25 - 32
  • [33] Development and validation of determination of genotoxic impurity Bromoethane in Vigabatrin drug substance using head space gas chromatographic method [HS-GC]
    Narapereddy, Krishna Prasad
    Alladi, Devi Sravanthi
    PHARMACIA, 2023, 70 (01) : 203 - 207
  • [34] VALIDATION OF A HEADSPACE GAS CHROMATOGRAPHY METHOD FOR VOLATILE RESIDUAL SOLVENTS IDENTIFICATION IN RADIOPHARMACEUTICALS
    Tudoroiu-cornoiu, Maria-Roxana
    Chilug, Elena Livia
    Ion, Ion
    Niculae, Dana
    UNIVERSITY POLITEHNICA OF BUCHAREST SCIENTIFIC BULLETIN SERIES B-CHEMISTRY AND MATERIALS SCIENCE, 2024, 86 (04): : 119 - 130
  • [35] Rapid and sensitive static headspace gas chromatography-mass spectrometry method for the analysis of ethanol and abused inhalants in blood
    Wasfi, IA
    Al-Awadhi, AH
    Al-Hatali, ZN
    Al-Rayami, FJ
    Al Katheeri, NA
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 799 (02): : 331 - 336
  • [36] DETERMINATION OF KETONE-BODIES IN POSTMORTEM BLOOD BY HEADSPACE GAS-CHROMATOGRAPHY
    FELBY, S
    NIELSEN, E
    FORENSIC SCIENCE INTERNATIONAL, 1994, 64 (2-3) : 83 - 88
  • [37] In-house validation of a simple headspace gas chromatography-mass spectrometry method for determination of furan levels in food
    Vranova, Janka
    Bednarikova, Alena
    Ciesarova, Zuzana
    JOURNAL OF FOOD AND NUTRITION RESEARCH, 2007, 46 (03): : 123 - 127
  • [38] Determination of ethanol in ionic liquids using headspace solid-phase microextraction–gas chromatography
    Fei Zhao
    Thehazhnan K. Ponnaiyan
    Christa M. Graham
    Constance A. Schall
    Sasidhar Varanasi
    Jared L. Anderson
    Analytical and Bioanalytical Chemistry, 2008, 392 : 1271 - 1275
  • [39] Analytical method development and validation of residual solvents in ethosuximide by GC-headspace technique
    Patan, Afroz
    Poojah, A.
    ANNALS OF PHYTOMEDICINE-AN INTERNATIONAL JOURNAL, 2024, 13 (01): : 979 - 987
  • [40] Development and validation of a gas chromatography-mass spectrometric method for the determination of sildenafil and desmethyl-sildenafil in whole blood
    Papoutsis, Ioannis
    Nikolaou, Panagiota
    Athanaselis, Sotirios
    Alevizopoulos, Giorgos
    Pistos, Constantinos
    Paraskevopoulou, Christina
    Spiliopoulou, Chara
    JOURNAL OF SEPARATION SCIENCE, 2011, 34 (21) : 3037 - 3042