Prolonged use of heat and moisture exchangers does not affect device efficiency or frequency rate of nosocomial pneumonia

被引:52
作者
Davis, K [1 ]
Evans, SL [1 ]
Campbell, RS [1 ]
Johannigman, JA [1 ]
Luchette, FA [1 ]
Porembka, DT [1 ]
Branson, RD [1 ]
机构
[1] Univ Cincinnati, Coll Med, Dept Surg, Div Trauma & Crit Care, Cincinnati, OH 45267 USA
关键词
humidity; heat and moisture exchanger; hygrometer; nosocomial infection; pneumonia; hygroscopic condenser humidifier; mechanical ventilatory support; airway resistance; cost; absolute humidity; humidification of inspired gases; tracheal intubation;
D O I
10.1097/00003246-200005000-00026
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To determine whether use of a single heat and moisture exchanger (HME) for less than or equal to 120 hrs affects efficiency, resistance, level of bacterial colonization, frequency rate of nosocomial pneumonia, and cost compared with changing the HME every 24 hrs. Design: Prospective, controlled, randomized, unblinded study. Setting: Surgical intensive care unit at a university teaching hospital. Patients: A total of 220 consecutive patients requiring mechanical ventilation for >48 hrs. Interventions: Patients were randomized to one of three groups: a) hygroscopic HME (Aqua+) changed every 24 hrs (HHME-24); b) hydrophobic HME (Duration HME) changed every 120 hrs (HME-120); and c) hygroscopic HME (Aqua+) changed every 120 hrs (HHME-120). Devices in all groups could be changed at the discretion of the staff when signs of occlusion or increased resistance were identified. Measurements and Main Results: Daily measurements of inspired gas temperature, inspired relative humidity, and device resistance were made. Additionally, daily cultures of the patient side of the device were accomplished. The frequency rate of nosocomial pneumonia was made by using clinical criteria. Ventilatory support variables, airway care, device costs, and clinical indicators of humidification efficiency (sputum volume, sputum efficiency) were also recorded. Prolonged use of both hygroscopic and hydrophobic devices did not diminish efficiency or increase resistance; There was no difference in the number of colony-forming units from device cultures over the 5-day period and no difference between colony-forming units in devices changed every 24 hrs compared with devices changed after 120 hrs. The average duration of use was 23 +/- 4 hrs in the HHME-24 group, 73 +/- 13 hrs in the HME-120 group, and 74 +/- 9 hrs in the HHME-120 group. Mean absolute humidity was greater for the hygroscopic devices (30.4 +/- 1.1 mg of H2O/L) compared with the hydrophobic devices (27.8 +/- 1.3 mg of H2O/L). The frequency rate of nosocomial pneumonia was 8% (8:100) in the HHME-24 group, 8.3% (5:60) in the HME-120 group, and 6.6% (4:60) in the HHME-120 group. Pneumonia rates per 1000 ventilatory support days were 20:1000 in the HHME-24 group, 20.8:1000 in the HME-120 group, and 16.6:1000 in the HHME-120 group. Costs per day were $3.24 for the HHME-24 group, $2.98 for the HME-120 group, and $1.65 for the HHME-120 group. Conclusions: Changing the hydrophobic or hygroscopic HME after 3 days does not diminish efficiency, increase resistance, or alter bacterial colonization. The frequency rate of nosocomial pneumonia was also unchanged. Use of HMEs for >24 hrs, up to 72 hrs, is safe and cost effective.
引用
收藏
页码:1412 / 1418
页数:7
相关论文
共 22 条
  • [1] [Anonymous], 1996, RESP CARE
  • [2] BRANSON R, 1996, RESP CARE, V4, P736
  • [3] PREVENTING NOSOCOMIAL PNEUMONIA - STATE-OF-THE-ART AND PERSPECTIVES FOR THE 1990S
    CRAVEN, DE
    STEGER, KA
    BARBER, TW
    [J]. AMERICAN JOURNAL OF MEDICINE, 1991, 91 : S44 - S53
  • [4] CRAVEN DE, 1984, AM REV RESPIR DIS, V129, P625
  • [5] CHANGING HEAT AND MOISTURE EXCHANGERS EVERY 48 HOURS RATHER THAN 24 HOURS DOES NOT AFFECT THEIR EFFICACY AND THE INCIDENCE OF NOSOCOMIAL PNEUMONIA
    DJEDAINI, K
    BILLIARD, M
    MIER, L
    LEBOURDELLES, G
    BRUN, P
    MARKOWICZ, P
    ESTAGNASIE, P
    COSTE, F
    BOUSSOUGANT, Y
    DREYFUSS, D
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1995, 152 (05) : 1562 - 1569
  • [6] DREYFUSS D, 1995, AM J RESP CRIT CARE, V151, P986
  • [7] PROSPECTIVE-STUDY OF NOSOCOMIAL PNEUMONIA AND OF PATIENT AND CIRCUIT COLONIZATION DURING MECHANICAL VENTILATION WITH CIRCUIT CHANGES EVERY 48 HOURS VERSUS NO CHANGE
    DREYFUSS, D
    DJEDAINI, K
    WEBER, P
    BRUN, P
    LANORE, JJ
    RAHMANI, J
    BOUSSOUGANT, Y
    COSTE, F
    [J]. AMERICAN REVIEW OF RESPIRATORY DISEASE, 1991, 143 (04): : 738 - 743
  • [8] CDC DEFINITIONS FOR NOSOCOMIAL INFECTIONS, 1988
    GARNER, JS
    JARVIS, WR
    EMORI, TG
    HORAN, TC
    HUGHES, JM
    [J]. AMERICAN JOURNAL OF INFECTION CONTROL, 1988, 16 (03) : 128 - 140
  • [9] GEORGE DL, 1993, INFECT CONT HOSP EP, V14, P163, DOI 10.1086/646705
  • [10] RESULTS OF A SURVEY OF INFECTION CONTROL PRACTICES FOR RESPIRATORY THERAPY EQUIPMENT
    GOULARTE, TA
    CRAVEN, DE
    [J]. INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 1986, 7 (06) : 327 - 330