Flow rate estimation of a centrifugal blood pump using the passively stabilized eccentric position of a magnetically levitated impeller

被引:8
作者
Shidai, Shuya [1 ]
Masuzawa, Toru [2 ]
Osa, Masahiro [2 ]
机构
[1] Ibaraki Univ, Grad Sch Sci & Engn, 4-12-1 Nakanarusawa, Hitachi, Ibaraki 3168511, Japan
[2] Ibaraki Univ, Dept Mech Engn, 4-12-1 Nakanarusawa, Hitachi, Ibaraki 3168511, Japan
关键词
Flow rate estimation; magnetically levitated impeller; centrifugal blood pump; passive stabilization; eccentric impeller position; PULSATILE FLOW; VISCOSITY; DEVICE;
D O I
10.1177/0391398819833372
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Flow rate estimation for ventricular assist devices without additional flow sensors can improve the quality of life of patients. In this article, a novel flow estimation method using the passively stabilized displacement of a magnetically levitated impeller is developed to achieve sufficient accuracy and precision of flow estimation for ventricular assist devices in a simple manner. The magnetically levitated impeller used is axially suspended by a magnetic bearing in a centrifugal blood pump that has been developed by our group. The radial displacement of the impeller, which is restricted by passive stability, can be correlated with the flow rate because the radial hydraulic force on the impeller varies according to the flow rate. To obtain the correlation with various blood viscosities, the relationships between the radial displacements of the magnetically levitated impeller and the pressure head-flow rate characteristics of the pump were determined simultaneously using aqueous solutions of glycerol with a potential blood viscosity range. The measurement results showed that accurate steady flow rates could be estimated with a coefficient of determination of approximately 0.97 and mean absolute error of approximately 0.22 L/min without fluid viscosity measurements by using the relationships between the impeller displacement and the flow rate. Moreover, a precision of approximately 0.01 (L/min)/mu m was obtained owing to a strong estimation indicator signal provided by the large displacement of the passively stabilized impeller; thus, the proposed estimation method can help ensure sufficient accuracy and precision for ventricular assist devices in a simple manner, even if the blood viscosity is unknown.
引用
收藏
页码:291 / 298
页数:8
相关论文
共 26 条
[1]   Non-invasive estimation of pulsatile flow and differential pressure in an implantable rotary blood pump for heart failure patients [J].
AlOmari, A. H. ;
Savkin, A. V. ;
Karantonis, D. M. ;
Lim, E. ;
Lovell, N. H. .
PHYSIOLOGICAL MEASUREMENT, 2009, 30 (04) :371-386
[2]   Alternation of left ventricular load by a continuous-flow left ventricular assist device with a native heart load control system in a chronic heart failure model [J].
Arakawa, Mamoru ;
Nishimura, Takashi ;
Takewa, Yoshiaki ;
Umeki, Akihide ;
Ando, Masahiko ;
Adachi, Hideo ;
Tatsumi, Eisuke .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 148 (02) :698-704
[3]   Dynamic characteristics of a magnetically levitated impeller in a centrifugal blood pump [J].
Asama, Junichi ;
Shinshi, Tadahiko ;
Hoshi, Hideo ;
Takatani, Setsuo ;
Shimokohbe, Akira .
ARTIFICIAL ORGANS, 2007, 31 (04) :301-311
[4]   Measurement for implantable rotary blood pumps [J].
Bertram, CD .
PHYSIOLOGICAL MEASUREMENT, 2005, 26 (04) :R99-R117
[5]   Control of the Fluid Viscosity in a Mock Circulation [J].
Boes, Stefan ;
Ochsner, Gregor ;
Amacher, Raffael ;
Petrou, Anastasios ;
Meboldt, Mirko ;
Daners, Marianne Schmid .
ARTIFICIAL ORGANS, 2018, 42 (01) :68-77
[6]   HeartMate III: Pump design for a centrifugal LVAD with a magnetically levitated rotor [J].
Bourque, K ;
Gernes, DB ;
Loree, HM ;
Richardson, JS ;
Poirier, VL ;
Barletta, N ;
Fleischli, A ;
Foiera, G ;
Gempp, TM ;
Schoeb, R ;
Litwak, KN ;
Akimoto, T ;
Watach, MJ ;
Litwak, P .
ASAIO JOURNAL, 2001, 47 (04) :401-405
[7]   Development of a Pump Flow Estimator for Rotary Blood Pumps to Enhance Monitoring of Ventricular Function [J].
Granegger, Marcus ;
Moscato, Francesco ;
Casas, Fernando ;
Wieselthaler, Georg ;
Schima, Heinrich .
ARTIFICIAL ORGANS, 2012, 36 (08) :691-699
[8]  
Gülich JF, 2010, CENTRIFUGAL PUMPS, SECOND EDITION, P1, DOI 10.1007/978-3-642-12824-0
[9]   Haemolysis in patients with ventricular assist devices: major differences between systems [J].
Heilmann, Claudia ;
Geisen, Ulrich ;
Benk, Christoph ;
Berchtold-Herz, Michael ;
Trummer, Georg ;
Schlensak, Christian ;
Zieger, Barbara ;
Beyersdorf, Friedhelm .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2009, 36 (03) :580-584
[10]   Sensorless Viscosity Measurement in a Magnetically-Levitated Rotary Blood Pump [J].
Hijikata, Wataru ;
Rao, Jun ;
Abe, Shodai ;
Takatani, Setsuo ;
Shinshi, Tadahiko .
ARTIFICIAL ORGANS, 2015, 39 (07) :559-568