摘要
Background: We compared the AdvanSure hepatitis B virus real-time polymerase chain reaction (AdvanSure HBV) kit with three other HBV DNA quantification assays and evaluated its performance. Methods:The AdvanSure HBV real-time PCR assay was compared with the Abbott RealTime HBV Quantification Kit, the COBAS TaqMan HBV Test, and the VERSANT HBV branched DNA 3.0 assay. The precision, linearity, accuracy, limit of detection (LOD), cross reactivity, and genotype inclusivity of the assays were compared, and any influence of the sampling tube type was evaluated. Results: The AdvanSure HBV PCR showed good correlations with the three other HBV DNA assays. The R-2 coefficients were 0.944, 0.939, and 0.921 with the Abbott RealTime HBV Quantification Kit, the COBAS TaqMan HBV Test, and the VERSANT bDNA 3.0 assay, respectively. Linearity was good in the tested range of 1.15-8.45 log(10) IU/ml. The lower LOD result was consistent with the 18 IU/ml claimed by the manufacturer. HBV genotypes A-F were all correctly amplified, and no cross reactivity was found in samples with-high HCV RNA levels or high protein concentrations. The results were not influenced by the sample preparation tube (i.e. plain tube, SST, and EDTA containing tubes). Conclusion: The AdvanSure HBV real-time PCR assay is a reliable method for quantifying HBV DNA levels in routine laboratory testing.
引用
收藏
页码:230 / 237
页数:8
相关论文
共 50 条
  • [21] A real-time PCR method for the quantitative detection of HBV viral DNA.
    Rhoads, JA
    Riesing, KS
    Pabich, EK
    Robinson, JM
    Mullen, CR
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2005, 7 (05): : 669 - 670
  • [22] Development and assessment of a novel real-time PCR assay for quantitation of HBV DNA
    Paraskevis, D
    Haida, C
    Tassopoulos, N
    Raptopoulou, M
    Tsantoulas, D
    Papachristou, H
    Sypsa, V
    Hatzakis, A
    JOURNAL OF VIROLOGICAL METHODS, 2002, 103 (02) : 201 - 212
  • [23] Comparison of three real-time PCR for the quantification of polyomavirus BK
    Stellrecht, A.
    Espino, A. A.
    Nattanmai, S. M.
    Jackson, W. F.
    Conti, D. J.
    JOURNAL OF CLINICAL VIROLOGY, 2013, 56 (04) : 354 - 359
  • [24] Development of a new ultra sensitive real-time PCR assay (ultra sensitive RTQ-PCR) for the quantification of HBV-DNA
    Paraskevis, Dimitrios
    Beloukas, Apostolos
    Haida, Catherine
    Katsoulidou, Antigoni
    Moschidis, Zisis
    Hatzitheodorou, Helen
    Varaklioti, Agoritsa
    Sypsa, Vana
    Hatzakis, Angelos
    VIROLOGY JOURNAL, 2010, 7
  • [25] Development of a new ultra sensitive real-time PCR assay (ultra sensitive RTQ-PCR) for the quantification of HBV-DNA
    Dimitrios Paraskevis
    Apostolos Beloukas
    Catherine Haida
    Antigoni Katsoulidou
    Zisis Moschidis
    Helen Hatzitheodorou
    Agoritsa Varaklioti
    Vana Sypsa
    Angelos Hatzakis
    Virology Journal, 7
  • [26] Performance evaluation of cobas HBV real-time PCR assay on Roche cobas 4800 System in comparison with COBAS AmpliPrep/COBAS TaqMan HBV Test
    Kim, Hanah
    Hur, Mina
    Bae, Eunsin
    Lee, Kyung-A
    Lee, Woo-In
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (07) : 1133 - 1139
  • [27] A real-time PCR method for the quantitative analysis of HBV DNA in serum and liver tissue
    Cariani, E
    Zanella, I
    Rossini, A
    Domenighini, D
    Finazzi, D
    Albertini, A
    JOURNAL OF HEPATOLOGY, 2001, 34 : 141 - 141
  • [28] Use in Routine Clinical Practice of Two Commercial Real-Time PCR Assays for Detection of Mycobacterium tuberculosis Complex: Comparison of COBAS TaqMan MTB Test and AdvanSure TB/NTM Real-Time PCR
    Suh, S.
    Lee, O.
    Choi, H.
    Kee, S.
    Shin, J.
    Ryang, D.
    Park, B.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2013, 15 (06): : 888 - 888
  • [29] Usefulness of in-house real time PCR for HBV DNA quantification in serum and oral fluid samples
    Portilho, Moyra Machado
    da Fonseca Mendonca, Ana Carolina
    Bezerra, Cristianne Sousa
    do Espirito-Santos, Marcia Paschoal
    de Paula, Vanessa Salete
    Nabuco, Leticia Cancella
    Villela-Nogueira, Cristiane Alves
    Lewis-Ximenez, Lia Laura
    Lampe, Elisabeth
    Villar, Livia Melo
    JOURNAL OF VIROLOGICAL METHODS, 2018, 256 : 100 - 106
  • [30] Comparison of the BioSewoom Real-Q™ hepatitis B virus quantification test, the VERSANT® HBV 3.0 assay, and the COBAS amplicor HBV Monitor™ test for measurement of HBV DNA in serum
    Kee, S. J.
    Suh, S. P.
    Cho, D.
    Shin, M. G.
    Shin, J. H.
    Ryang, D. W.
    CLINICAL CHEMISTRY, 2006, 52 (06) : A49 - A49