共 40 条
Cathodically pretreated poly(1-aminoanthraquinone)-modified electrode for determination of ascorbic acid, dopamine, and uric acid
被引:20
作者:
Troiani, Estela de Pieri
[1
]
Faria, Ronaldo Censi
[1
]
机构:
[1] Univ Fed Sao Carlos, Dept Quim, BR-13565905 Sao Carlos, SP, Brazil
基金:
巴西圣保罗研究基金会;
关键词:
Dopamine;
Ascorbic acid;
Uric acid;
Cathodic pretreatment;
Conducting polymer;
Poly(1-aminoanthraquinone);
OVER-OXIDIZED POLYPYRROLE;
GLASSY-CARBON ELECTRODE;
ELECTROCHEMICAL DETECTION;
IN-VIVO;
ELECTROPOLYMERIZATION;
PHTHALOCYANINES;
COMPOSITE;
POLYMER;
FILMS;
D O I:
10.1007/s10800-013-0577-7
中图分类号:
O646 [电化学、电解、磁化学];
学科分类号:
081704 ;
摘要:
This paper describes a rapid and reliable method for dopamine (DA), ascorbic acid (AA), and uric acid (UA) determination in human urine using a cathodically pretreated poly(1-aminoanthraquinone) (PAAQ)-modified electrode. By applying a simple cathodic pretreatment to the PAAQ electrode well-defined voltammetric peaks for AA, DA, and UA were obtained. The pretreated PAAQ showed good selectivity, sensitivity, and repeatability for measuring AA, DA, and UA with detection limits of 2.50 x 10(-5), 3.05 x 10(-6), and 1.15 x 10(-5) M, respectively. The practical applicability of the modified electrode is illustrated by selective measurements of AA and UA in human urine without any preliminary treatment. Recovery values between 94.8 and 102 % for AA and between 77.8 and 100 % for UA were obtained with a relative standard deviation of 2.74 and 2.98 %, respectively.
引用
收藏
页码:919 / 926
页数:8
相关论文