Is the new Occlutech duct occluder an appropriate device for transcatheter closure of patent ductus arteriosus?

被引:7
作者
Godart, Francois [1 ]
Houeijeh, Ali [1 ]
Domanski, Olivia [1 ]
Guillaume, Marie-Paule [1 ]
Brard, Melanie [2 ]
Lucron, Hugues [2 ]
机构
[1] Univ Lille Nord France, Dept Pediat & Congenital Heart Dis, Lille, France
[2] Univ Hosp Martinique, Antilles Guyane Pediat Cardiol Ctr M3C, Fort De France, Martinique, France
关键词
Patent ductus arteriosus; Transcatheter closure; Occlutech duct occluder; CATHETER CLOSURE; IMMEDIATE; PDA; EXPERIENCE; OCCLUSION;
D O I
10.1016/j.ijcard.2018.03.059
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim: To describe our initial experience with the Occlutech Duct Occluder (ODO) for percutaneous closure of patent ductus arteriosus (PDA). Methods: Retrospective review of patients undergoing transcatheter PDA closure with the ODO in 2 academic centers. Results: From April 2013 to September 2017, 42 patients underwent PDA closure. Median age at implantation was 34 months (range 4 months-68 years) and median weight was 12 kg (range 4.1-57 kg). Ducts were Krichenko type A duct (n = 34), type E (n = 6), and type C (n = 2). The mean duct diameter was 3.76 mm(range 1.69 to 9.95 mm, median 3.1 mm). Implantation succeeded in all. There was neither device embolization nor hemolysis. At device release, immediate angiogram showed a small residual shunt in 54.7%. During follow-up, Doppler echocardiography demonstrated 71% of full occlusion at day one, rising to 95% at one month and 100% at one year and half after implantation. The mean maximal systolic pressure gradient in left pulmonary artery was 4.2 +/- 4.3 mm and across the distal aortic arch 5.4 +/- 4.7 mm Hg. No patient had any significant stenosis with clinical relevance. Conclusions: ODO is safe and effective in transcatheter closure of PDA including relatively large sized ducts. The results are satisfactory with a high level of full occlusion and a low rate of complications. Further evaluation with larger studies and longer follow-up will be required to confirm these preliminary good results. (c) 2018 Published by Elsevier B.V.
引用
收藏
页码:54 / 57
页数:4
相关论文
共 20 条
  • [1] The new Occlutech® PDA occluder: Initial human experience
    Abdelbasit, M. A. Elbashier
    Alwi, Mazeni
    Kandavello, Geetha
    Mood, Marhisham Che
    Samion, Hasri
    Hijazi, Ziyad M.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2015, 86 (01) : 94 - 99
  • [2] Transcatheter Closure of Persistent Ductus Arteriosus in Adults
    Bentham, James R.
    Thomson, John D. R.
    Gibbs, John L.
    [J]. JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2012, 25 (05) : 501 - 504
  • [3] Transcatheter Closure of Patent Ductus Arteriosus in Children with the Occlutech Duct Occluder
    Bilici, Meki
    Demir, Fikri
    Akin, Alper
    Ture, Mehmet
    Balik, Hasan
    Kuyumcu, Mahir
    [J]. PEDIATRIC CARDIOLOGY, 2017, 38 (08) : 1598 - 1605
  • [4] The Amplatzer duct occluder: Experience in 209 patients
    Bilkis, AA
    Alwi, M
    Hasri, S
    Haifa, AL
    Geetha, K
    Rehman, MA
    Hasanah, I
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (01) : 258 - 261
  • [5] The new Occlutech® patent ductus arteriosus occluder: Single centre experience
    Boudjemline, Younes
    [J]. ARCHIVES OF CARDIOVASCULAR DISEASES, 2016, 109 (6-7) : 384 - 389
  • [6] Percutaneous Closure of Patent Ductus Arteriosus: A Multiinstitutional Registry Comparing Multiple Devices
    Brunetti, Marissa A.
    Ringel, Richard
    Owada, Carl
    Coulson, John
    Jennings, Jacky M.
    Hoyer, Mark H.
    Everett, Allen D.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 76 (05) : 696 - 702
  • [7] Short-Term Outcomes of Patent Ductus Arteriosus Closure With New Occlutech® Duct Occluder: A Multicenter Study
    Dedeoglu, Reyhan
    Bilici, Meki
    Demir, Fikri
    Demir, Fadli
    Acar, Onur Caglar
    Hallioglu, Olgu
    Pac, Aysenur
    Kiziltas, Alev
    Karabel, Duran
    Kula, Serdar
    Cimen, Derya
    Baspinar, Osman
    Atik, Sezen Ugan
    Saltik, Irfan Levent
    [J]. JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2016, 29 (03) : 325 - 331
  • [8] Faella HJ, 2000, CATHETER CARDIO INTE, V51, P50, DOI 10.1002/1522-726X(200009)51:1<50::AID-CCD11>3.0.CO
  • [9] 2-6
  • [10] Trans-Catheter Closure of Patent Ductus Arteriosus-What Is the Best Device?
    Ghasemi, Abdolrahim
    Pandya, Swati
    Reddy, Surendranath Veeram
    Turner, Daniel R.
    Du, Wei
    Navabi, Mohamad Ali
    Mirzaaghayan, Mohammad Reza
    Kiani, Abdolrazagh
    Sloan, Katherine
    Forbes, Thomas J.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 76 (05) : 687 - 695