Artifact robustness, inter- and intraindividual baseline stability, and rational EEG parameter selection

被引:26
作者
Bruhn, J
Bouillon, TW
Hoeft, A
Shafer, SL
机构
[1] Univ Bonn, Dept Anesthesia & Intens Care Med, D-53105 Bonn, Germany
[2] Palo Alto VA Hlth Care Syst, Palo Alto, CA USA
[3] Stanford Univ, Sch Med, Dept Anesthesia, Stanford, CA 94305 USA
关键词
D O I
10.1097/00000542-200201000-00015
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Artifact robustness (i.e., size of deviation of an electroencephalographic parameter value from baseline caused by artifacts) and baseline stability (i.e., consistency of median baseline values) of electroencephalographic parameters profoundly influence electroencephalography-based pharmacodynamic parameter estimation and the usefulness of the processed electroencephalogram as measure of the arousal state of the central nervous system (depth of anesthesia). In this study, the authors compared the artifact robustness and the interindividual and intraindividual baseline stability of several univariate descriptors of the electroencephalogram (Shannon entropy, approximate entropy, spectral edge frequency 95, delta ratio, and canonical univariate parameter). Methods: Electroencephalographic data of 16 healthy volunteers before and after administration of an intravenous bolus of propofol (2 mg/kg body weight) were analyzed. Each volunteer was studied twice. The baseline electroencephalogram was recorded for a median of 18 min before drug administration. For each electroencephalographic descriptor, the authors calculated the following: (1) baseline variability (= (median baseline - median effect) [i.e., signal]/SD baseline [i.e., noise]) without artifact rejection; (2) baseline variability with artifact rejection; and (3) baseline stability within and between individuals (= (median baseline - median effect) averaged over all volunteers/SD of all median baselines). Results: Without artifact rejection, Shannon entropy and canonical univariate parameter displayed the highest signal-to-noise ratio. After artifact rejection, approximate entropy, Shannon entropy, and the canonical univariate parameter displayed the highest signal-to-noise ratio. Baseline stability within and between individuals was highest for approximate entropy. Conclusions: With regard to robustness against artifacts, the electroencephalographic entropy parameters and the canonical univariate parameter were superior to spectral edge frequency 95 and delta ratio. Electroencephalographic approximate entropy displayed the best interindividual and intraindividual baseline stability.
引用
收藏
页码:54 / 59
页数:6
相关论文
共 18 条
  • [1] BARLOW JS, 1986, CLIN APPL COMPUTER A, P15
  • [2] A comparison of spectral edge, delta power, and bispectral index as EEG measures of alfentanil, propofol, and midazolam drug effect
    Billard, V
    Gambus, PL
    Chamoun, N
    Stanski, DR
    Shafer, SL
    [J]. CLINICAL PHARMACOLOGY & THERAPEUTICS, 1997, 61 (01) : 45 - 58
  • [3] Shannon entropy applied to the measurement of the electroencephalographic effects of desflurane
    Bruhn, J
    Lehmann, LE
    Röpcke, H
    Bouillon, TW
    Hoeft, A
    [J]. ANESTHESIOLOGY, 2001, 95 (01) : 30 - 35
  • [4] Approximate entropy as an electroencephalographic measure of anesthetic drug effect during desflurane anesthesia
    Bruhn, J
    Röpcke, H
    Hoeft, A
    [J]. ANESTHESIOLOGY, 2000, 92 (03) : 715 - 726
  • [5] Automatic detection of artefacts during intraoperative EEG monitoring
    Eckert, O
    [J]. BIOMEDIZINISCHE TECHNIK, 1998, 43 (09): : 236 - 242
  • [6] VALIDATION OF THE ALFENTANIL CANONICAL UNIVARIATE PARAMETERS AS A MEASURE OF OPIOID EFFECT ON THE ELECTROENCEPHALOGRAM
    GAMBUS, PL
    GREGG, KM
    SHAFER, SL
    [J]. ANESTHESIOLOGY, 1995, 83 (04) : 747 - 756
  • [7] APPLICATION OF SEMILINEAR CANONICAL CORRELATION TO THE MEASUREMENT OF OPIOID DRUG EFFECT
    GREGG, KM
    VARVEL, JR
    SHAFER, SL
    [J]. JOURNAL OF PHARMACOKINETICS AND BIOPHARMACEUTICS, 1992, 20 (06): : 611 - 635
  • [8] Electroencephalographic derivatives as a tool for predicting the depth of sedation and anesthesia induced by sevoflurane
    Katoh, T
    Suzuki, A
    Ikeda, K
    [J]. ANESTHESIOLOGY, 1998, 88 (03) : 642 - 650
  • [9] PHYSIOLOGICAL TIME-SERIES ANALYSIS - WHAT DOES REGULARITY QUANTIFY
    PINCUS, SM
    GOLDBERGER, AL
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1994, 266 (04): : H1643 - H1656
  • [10] A REGULARITY STATISTIC FOR MEDICAL DATA-ANALYSIS
    PINCUS, SM
    GLADSTONE, IM
    EHRENKRANZ, RA
    [J]. JOURNAL OF CLINICAL MONITORING, 1991, 7 (04): : 335 - 345