Is it possible to improve the accuracy of implants inserted with a stereolithographic surgical guide by reducing the tolerance between mechanical components?

被引:26
作者
Cassetta, M. [1 ]
Di Mambro, A. [1 ]
Giansanti, M. [1 ]
Stefanelli, L. V. [1 ]
Barbato, E. [1 ]
机构
[1] Univ Roma La Sapienza, Dept Oral & Maxillofacial Sci, Sch Dent, I-00144 Rome, Italy
关键词
COMPLICATIONS;
D O I
10.1016/j.ijom.2013.01.011
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
When using a stereolithographic surgical guide, a potentially clinically relevant error may be the mechanical error caused by the bur guide cylinder gap due to the presence of a rotational allowance of drills in the tubes. The aim of the present study was to determine if it is possible to reduce the total error by limiting the tolerance among the mechanical components and to evaluate its clinical incidence. Sixty implants were inserted in eight totally edentate subjects using the External Hex Safe((R)) system with mechanical components modified to minimize the tolerance. Pre- and postoperative computed tomography images were compared, and the angular deviation was calculated between the planned and the placed implants. The mean angular deviation was 2.02 degrees (range 0.81-3.48 degrees; standard deviation 0.87). The results of the present study show that by limiting the error that originates from mechanical components, the total error could be significantly reduced.
引用
收藏
页码:887 / 890
页数:4
相关论文
共 10 条
[1]   Accuracy of Two Stereolithographic Guide Systems for Computer-Aided Implant Placement: A Computed Tomography-Based Clinical Comparative Study [J].
Arisan, Volkan ;
Karabuda, Z. Cuneyt ;
Ozdemir, Tayfun .
JOURNAL OF PERIODONTOLOGY, 2010, 81 (01) :43-51
[2]   Depth deviation and occurrence of early surgical complications or unexpected events using a single stereolithographic surgi-guide [J].
Cassetta, M. ;
Stefanelli, L. V. ;
Giansanti, M. ;
Di Mambro, A. ;
Calasso, S. .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2011, 40 (12) :1377-1387
[3]  
Cassetta M, 2011, CLIN IMPLANT DENT R, DOI DOI 10.1111/J.1708-8208.2011.00369.X.JULY
[4]  
Cassetta M, 2012, INT J ORAL MAXILLOF, DOI DOI 10.1016/J.IJ0M.2012.06.010.JULY
[5]   Accuracy and Complications Using Computer-Designed Stereolithographic Surgical Guides for Oral Rehabilitation by Means of Dental Implants: A Review of the Literature [J].
D'haese, Jan ;
Van De Velde, Tommie ;
Komiyama, Ai ;
Hultin, Margaretha ;
De Bruyn, Hugo .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2012, 14 (03) :321-335
[6]   A Prospective Study on the Accuracy of Mucosally Supported Stereolithographic Surgical Guides in Fully Edentulous Maxillae [J].
D'haese, Jan ;
Van De Velde, Tommie ;
Elaut, Lucien ;
De Bruyn, Hugo .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2012, 14 (02) :293-303
[7]  
Mandelaris GA, 2010, INT J PERIODONT REST, V30, P275
[8]  
Pettersson A, 2010, CLIN IMPLANT DENT RE, DOI DOI 10.HTTP://DX.D0I.0RG/10.1L1L/J.1708-8208.2010.00285.X.MAY
[9]   A systematic review on the accuracy and the clinical outcome of computer-guided template-based implant dentistry [J].
Schneider, David ;
Marquardt, Pascal ;
Zwahlen, Marcel ;
Jung, Ronald E. .
CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 :73-86
[10]   Tolerance within a surgical guide [J].
Van Assche, N. ;
Quirynen, M. .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (04) :455-458