Measurement of joint kinematics using a conventional clinical single-perspective flat-panel radiography system

被引:2
作者
Seslija, Petar [1 ,2 ]
Teeter, Matthew G. [2 ,3 ]
Yuan, Xunhua [2 ,3 ]
Naudie, Douglas D. R. [2 ,4 ,5 ]
Bourne, Robert B. [4 ,5 ]
MacDonald, Steven J. [4 ,5 ]
Peters, Terry M. [1 ,2 ,3 ]
Holdsworth, David W. [1 ,2 ,3 ]
机构
[1] Univ Western Ontario, Biomed Engn Grad Program, London, ON N6A 5B9, Canada
[2] John P Robarts Res Inst, Imaging Res Labs, London, ON N6A 5K8, Canada
[3] Univ Western Ontario, Dept Med Biophys, London, ON N6A 5C1, Canada
[4] Univ Western Ontario, Div Orthopaed Surg, London, ON N6A 4L6, Canada
[5] London Hlth Sci Ctr, London, ON N6A 4L6, Canada
基金
加拿大健康研究院;
关键词
tracking; kinematics; registration; accuracy; flat-panel detector; KNEE; MODEL; FLUOROSCOPY; ACCURACY; VALIDATION; MOTION; RSA; TRACKING; GAIT; RECONSTRUCTION;
D O I
10.1118/1.4752205
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The ability to accurately measure joint kinematics is an important tool in studying both normal joint function and pathologies associated with injury and disease. The purpose of this study is to evaluate the efficacy, accuracy, precision, and clinical safety of measuring 3D joint motion using a conventional flat-panel radiography system prior to its application in an in vivo study. Methods: An automated, image-based tracking algorithm was implemented to measure the three-dimensional pose of a sparse object from a two-dimensional radiographic projection. The algorithm was tested to determine its efficiency and failure rate, defined as the number of image frames where automated tracking failed, or required user intervention. The accuracy and precision of measuring three-dimensional motion were assessed using a robotic controlled, tibiofemoral knee phantom programmed to mimic a subject with a total knee replacement performing a stair ascent activity. Accuracy was assessed by comparing the measurements of the single-plane radiographic tracking technique to those of an optical tracking system, and quantified by the measurement discrepancy between the two systems using the Bland-Altman technique. Precision was assessed through a series of repeated measurements of the tibiofemoral kinematics, and was quantified using the across-trial deviations of the repeated kinematic measurements. The safety of the imaging procedure was assessed by measuring the effective dose of ionizing radiation associated with the x-ray exposures, and analyzing its relative risk to a human subject. Results: The automated tracking algorithm displayed a failure rate of 2% and achieved an average computational throughput of 8 image frames/s. Mean differences between the radiographic and optical measurements for translations and rotations were less than 0.08 mm and 0.07 degrees in-plane, and 0.24 mm and 0.6 degrees out-of-plane. The repeatability of kinematics measurements performed using the radiographic tracking technique was better than +/- 0.09 mm and 0.12 degrees in-plane, and +/- 0.70 mm and +/- 0.07 degrees out-of-plane. The effective dose associated with the imaging protocol used was 15 mu Sv for 10 s of radiographic cine acquisition. Conclusions: This study demonstrates the ability to accurately measure knee-joint kinematics using a single-plane radiographic measurement technique. The measurement technique can be easily implemented at most clinical centers equipped with a modern-day radiographic x-ray system. The dose of ionizing radiation associated with the image acquisition represents a minimal risk to any subjects undergoing the examination. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4752205]
引用
收藏
页码:6090 / 6103
页数:14
相关论文
共 51 条
  • [31] Skin injuries from fluoroscopically guided procedures: part I, Characteristics of radiation injury
    Koenig, TR
    Wolff, D
    Mettler, FA
    Wagner, LK
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (01) : 3 - 11
  • [32] In vivo determination of total knee arthroplasty kinematics
    Komistek, Richard D.
    Mahfouz, Mohamed R.
    Bertin, Kim C.
    Rosenberg, Aaron
    Kennedy, William
    [J]. JOURNAL OF ARTHROPLASTY, 2008, 23 (01) : 41 - 50
  • [33] Implementation and validation of an implant-based coordinate system for RSA migration calculation
    Laende, Elise K.
    Deluzio, Kevin J.
    Hennigar, Allan W.
    Dunbar, Michael J.
    [J]. JOURNAL OF BIOMECHANICS, 2009, 42 (14) : 2387 - 2393
  • [34] Validation of a non-invasive fluoroscopic imaging technique for the measurement of dynamic knee joint motion
    Li, Guoan
    de Velde, Samuel K. Van
    Bingham, Jeffrey T.
    [J]. JOURNAL OF BIOMECHANICS, 2008, 41 (07) : 1616 - 1622
  • [35] A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images
    Mahfouz, MR
    Hoff, WA
    Komistek, RD
    Dennis, DA
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (12) : 1561 - 1574
  • [36] Effective doses in radiology and diagnostic nuclear medicine: A catalog
    Mettler, Fred A., Jr.
    Huda, Walter
    Yoshizumi, Terry T.
    Mahesh, Mahadevappa
    [J]. RADIOLOGY, 2008, 248 (01) : 254 - 263
  • [37] National Research Council of the National Academies, 2006, HLTH RISKS EXPOSURE
  • [38] PATIENT EFFECTIVE DOSE RESULTING FROM RADIOGRAPHIC EXAMINATIONS
    OKKALIDES, D
    FOTAKIS, M
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (798) : 564 - 572
  • [39] Ranstam J, 2000, ACTA ORTHOP SCAND, V71, P106
  • [40] AN ALGORITHM FOR ASSOCIATING THE FEATURES OF 2 IMAGES
    SCOTT, GL
    LONGUETHIGGINS, HC
    [J]. PROCEEDINGS OF THE ROYAL SOCIETY B-BIOLOGICAL SCIENCES, 1991, 244 (1309) : 21 - 26