Design of linear anti-scatter grid geometry with optimum performance for screen-film and digital mammography systems

被引:12
作者
Khodajou-Chokami, H. [1 ]
Sohrabpour, M. [1 ]
机构
[1] Sharif Univ Technol, Dept Energy Engn, Tehran, Iran
关键词
mammography; Monte Carlo; anti-scatter grid; simulation; ANTISCATTER GRIDS; SIMULATION; RADIOLOGY; REMOVAL;
D O I
10.1088/0031-9155/60/15/5753
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A detailed 3D Monte Carlo simulation of the grid geometrical parameters in screen-film mammography (SFM) and digital mammography (DM) systems has been performed. A combination of IEC 60627:2013 international standard conditions and other more clinically relevant parameters were used for this simulation. Accuracy of our results has been benchmarked with previously published data and good agreement has been obtained. Calculations in a wide range of linear anti-scatter grid geometries have been carried out. The evaluated parameters for the SFM system were the Bucky factor (BF) and contrast improvement factor (CIF) and for the DM system it was signal difference-to-noise ratio improvement factor (SIF). The CIF parameter was chosen to be nearly the same as the commercial grade, the BF and SIF were significantly improved compared to commercial grids in use today. Our optimized grid parameters for the SFM system were lead strip thickness d = 12 mu m, grid ratio r = 5 and strip density N = 65 lines/cm. And for the DM system these parameters were d = 5 mu m, r = 3 and N = 100 lines/cm. Both optimized grid sets have thinner d and higher N compared to the commercial grids.
引用
收藏
页码:5753 / 5765
页数:13
相关论文
共 32 条
[1]   MONTE-CARLO TECHNIQUES IN MEDICAL RADIATION PHYSICS [J].
ANDREO, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) :861-920
[2]   X-ray spectrum optimization of full-field digital mammography: Simulation and phantom study [J].
Bernhardt, Philipp ;
Mertelmeier, Thomas ;
Hoheisel, Martin .
MEDICAL PHYSICS, 2006, 33 (11) :4337-4349
[3]   QUALITY AND CHOICE OF POTTER BUCKY GRIDS .3. THE CHOICE OF A BUCKY GRID [J].
BONENKAMP, JG ;
BOLDINGH, WH .
ACTA RADIOLOGICA, 1959, 52 (03) :241-253
[4]   Scatter/primary in mammography: Comprehensive results [J].
Boone, JM ;
Lindfors, KK ;
Cooper, VN ;
Seibert, JA .
MEDICAL PHYSICS, 2000, 27 (10) :2408-2416
[5]   Monte Carlo validation in diagnostic radiological imaging [J].
Boone, JM ;
Buonocore, MH ;
Cooper, VN .
MEDICAL PHYSICS, 2000, 27 (06) :1294-1304
[6]   Grid and slot scan scatter reduction in mammography: Comparison by using Monte Carlo techniques [J].
Boone, JM ;
Seibert, JA ;
Tang, CM ;
Lane, SM .
RADIOLOGY, 2002, 222 (02) :519-527
[7]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[8]   Development and Monte Carlo Analysis of Antiscatter Grids for Mammography [J].
Boone, John M. ;
Makarova, Olga V. ;
Zyryanov, Vladislav N. ;
Tang, Cha-Mei ;
Mancini, Derrick C. ;
Moldovan, Nikolaie ;
Divan, Ralu .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2002, 1 (06) :441-447
[9]  
Bucky G., 1915, US Patent, Patent No. [1,164,987, 1164987]
[10]  
Bushberg J.T., 2012, ESSENTIAL PHYS MED I, V3