The validity of different display monitors in the assessment of dental implant site dimensions in cone beam computed tomography images

被引:6
作者
Al-Ekrish, Asma'a A. [1 ]
Ekram, Mohamed I. H. [1 ]
Al Faleh, Wafa [1 ]
Alkhader, Mustafa [1 ]
Al-Sadhan, Ra'ed [1 ]
机构
[1] King Saud Univ, Coll Dent, Dept Oral Med & Diagnost Sci, Riyadh, Saudi Arabia
关键词
CBCT; computer terminals; radiography; dental; digital; reliability; DIGITAL RADIOGRAPHS; APPROXIMAL CARIES; IN-VITRO; ACCURACY; RESOLUTION; SETTINGS; SYSTEMS; MAXILLA;
D O I
10.3109/00016357.2012.741709
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. To determine if the differing contrast resolution of various LCD monitors affects the reliability or accuracy of measurements of proposed implant sites. Materials and methods. Edentulous areas of human dry skulls were marked with radiopaque markers in order to standardize the plane of the transverse cross-sections of the ridges and path of measurements. The skulls were imaged by a CBCT device and the images stored in proprietary format on the workstation. The data sets were then transferred, with the proprietary reformatting software, to two different laptops using CDs. Transverse cross-sectional images of the ridges were reformatted on all three computers and ridge dimensions were recorded using the linear measurement tool of the proprietary software. Ridge dimensions were recorded directly from the three different monitors by two observers and compared to measurements recorded directly from the bone. The measurement errors and intra-and inter-examiner reliability were calculated for each monitor and compared with each other. Results. Intra-and inter-examiner reliability scores for the measurements recorded from all three devices were very high and ranged between 0.993-0.999. The mean of the absolute errors was 0.55 mm for the workstation, 0.61 mm for laptop 1 and 0.68 mm for laptop 2. The absolute errors were statistically significant for all three monitors (p-value < 0.001), but there was no statistically significant difference between the absolute errors obtained from the three monitors. Conclusions. No differences in the reliability or accuracy of measurements of implant site dimensions were obtained using color LCD monitors with different contrast resolution capabilities.
引用
收藏
页码:1085 / 1091
页数:7
相关论文
共 19 条
[1]   A comparative study of the accuracy and reliability of multidetector computed tomography and cone beam computed tomography in the assessment of dental implant site dimensions [J].
Al-Ekrish, A. A. ;
Ekram, M. .
DENTOMAXILLOFACIAL RADIOLOGY, 2011, 40 (02) :67-75
[2]  
[Anonymous], 2002, The Essential Physics of Medical Imaging
[3]   Image quality performance of liquid crystal display systems: Influence of display resolution, magnification and window settings on contrast-detail detection [J].
Bacher, Klaus ;
Smeets, Peter ;
De Hauwere, An ;
Voet, Tony ;
Duyck, Philippe ;
Verstraete, Koenraad ;
Thierens, Hubert .
EUROPEAN JOURNAL OF RADIOLOGY, 2006, 58 (03) :471-479
[4]  
Fatemitabar SA, 2010, INT J ORAL MAX IMPL, V25, P499
[5]  
Fuchs T, 2003, PHYS MEDICA, V19, P153
[6]   Influence of room lighting on grey-scale perception with a CRT- and a TFT monitor display [J].
Haak, R ;
Wicht, MJ ;
Hellmich, M ;
Nowak, G ;
Noack, MJ .
DENTOMAXILLOFACIAL RADIOLOGY, 2002, 31 (03) :193-197
[7]   Comparison of viewing angle and observer performances in different types of liquid-crystal display monitors [J].
Hatanaka S. ;
Morishita J. ;
Hiwasa T. ;
Dogomori K. ;
Toyofuku F. ;
Ohki M. ;
Higashida Y. .
Radiological Physics and Technology, 2009, 2 (2) :166-174
[8]   Effect of ambient light and monitor brightness and contrast settings on the detection of approximal caries in digital radiographs: an in vitro study [J].
Hellen-Halme, K. ;
Petersson, A. ;
Warfvinge, G. ;
Nilsson, M. .
DENTOMAXILLOFACIAL RADIOLOGY, 2008, 37 (07) :380-384
[9]   Detection of Artificial Occlusal Caries in a Phosphor Imaging Plate System with Two Types of LCD Monitors Versus Three Different Films [J].
Ilguy, Mehmet ;
Dincer, Semanur ;
Ilguy, Dilhan ;
Bayirli, Zuenduez .
JOURNAL OF DIGITAL IMAGING, 2009, 22 (03) :242-249
[10]   Effect of monitor display on detection of approximal caries lesions in digital radiographs [J].
Isidor, S. ;
Faaborg-Andersen, M. ;
Hintze, H. ;
Kirkevang, L-L ;
Frydenberg, M. ;
Haiter-Neto, F. ;
Wenzel, A. .
DENTOMAXILLOFACIAL RADIOLOGY, 2009, 38 (08) :537-541