Although regional myocardial perfusion can be currently quantified with myocardial contrast echocardiography (MCE) by using intermittent harmonic imaging (IHI), the method is tedious and time-consuming in the clinical setting. We hypothesized that regional myocardial perfusion can be quantified and the severity of coronary stenosis determined during hyperemia with MCE using real-time imaging (RTI) where microbubbles are not destroyed. Six open-chest dogs were studied during maximal hyperemia induced by adenosine in the absence or presence of coronary stenoses varying from mild to severe. Myocardial blood flow (MBF) was measured at each stage by using radiolabeled microspheres. MCE was performed using both IHI and RTL Data for the latter were acquired in both end-systole and end-diastole. No differences were found between myocardial flow velocity (MFV) derived from IHI and RTI when end-systolic frames were used for the latter. MFV was consistently higher for RTI (P < .01) when end-diastolic frames were used. A linear relation was noted between MFV and radiolabeled microsphere-derived MBF ratios from the stenosed and the normal beds when end-systolic frames were used for RTI (r = 0.78, P < .001), whereas no relation was found when end-diastolic frames were used (r = 0.08, P =.78). The scatter for assessing MBF (A-P) was minimal for HU and RTI (9%-10%) with end-systolic frames, whereas that for RTI with end-diastolic frames was large (30%). Furthermore the correlation with radiolabeled microsphere-derived MBF was significantly (P < .01) weaker with RTI when end-diastolic frames were used (,r = 0.53) than when end-systolic frames (r = 0.94) or IHI was used (r = 0.99). Data acquisition for IM was 10 minutes, whereas it was 8 seconds for RTL Thus, RTI can be used to quantify regional myocardial perfusion and stenosis severity during MCE. Only end-systolic frames, however, provide accurate data. RTI offers a rapid and easy means of assessing regional myocardial perfusion with MCE.
引用
收藏
页码:1173 / 1182
页数:10
相关论文
共 50 条
  • [31] CONTRAST ECHOCARDIOGRAPHY FOR THE DETERMINATION OF MYOCARDIAL PERFUSION
    ZWEHL, W
    KLINISCHE WOCHENSCHRIFT, 1986, 64 (21): : 1148 - 1149
  • [32] QUANTIFICATION OF MYOCARDIAL PERFUSION DEFECTS BY CONTRAST ECHOCARDIOGRAPHY
    SCHARTL, M
    FRITZSCH, T
    FRIEDMANN, W
    LANGE, L
    ZEITSCHRIFT FUR KARDIOLOGIE, 1984, 73 (09): : 560 - 567
  • [34] Global impaired myocardial perfusion in the setting of acute ischemia: A real-time myocardial contrast echocardiography assessment
    Khankirawatana, B
    Khankirawatana, S
    Lof, J
    Porter, TR
    CIRCULATION, 2001, 104 (17) : 469 - 469
  • [35] A new quantification software of myocardial perfusion using contrast echocardiography
    Giatrakos, N
    Gao, J
    Yang, G
    Nihoyannopoulos, P
    EUROPEAN HEART JOURNAL, 2004, 25 : 550 - 550
  • [36] Quantification of myocardial blood flow with power modulation real-time myocardial contrast echocardiography in dogs.
    Van, G
    Ay, T
    Pasquet, A
    London, V
    Bol, A
    Giselu, G
    Heyndrickx, GR
    Rafter, P
    Melin, JA
    Vanoverschelde, JL
    CIRCULATION, 2000, 102 (18) : 471 - 471
  • [37] Quantitative assessment of microcollateral recruitment during coronary occlusion using real-time intravenous myocardial contrast echocardiography
    Ishikura, Fuminobu
    Otani, Kentaro
    Kayano, Hiroyuki
    Toshida, Tsutomu
    Iwata, Akiko
    Asanuma, Toshihiko
    Kitakaze, Masaftimi
    Shinozaki, Yoshihiro
    Mori, Hidezo
    Beppu, Shintaro
    JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2008, 21 (02) : 139 - 145
  • [38] Quantitative assessment of the transmural distribution of myocardial perfusion by contrast echocardiography during graded coronary stenosis
    Masugata, H
    Cotter, B
    Peters, B
    Kwan, OL
    Strachan, GM
    DeMaria, AN
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (02) : 445A - 445A
  • [39] Detection and quantification of coronary stenosis severity with myocardial contrast echo cardiography
    Wei, K
    PROGRESS IN CARDIOVASCULAR DISEASES, 2001, 44 (02) : 81 - 100
  • [40] Visual assessment of reactive hyperemia during coronary angioplasty by myocardial contrast echocardiography
    Kurisu, S
    Sato, H
    Tateishi, H
    Kawagoe, T
    Ishihara, M
    Shimatani, Y
    Sakai, K
    Ueda, K
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 29 (02) : 7644 - 7644