Patient-Specific Quality Assurance Using Monte Carlo Dose Calculation and Elekta Log Files for Prostate Volumetric-Modulated Arc Therapy

被引:12
作者
Katsuta, Yoshiyuki [1 ,2 ]
Kadoya, Noriyuki [2 ]
Fujita, Yukio [3 ]
Shimizu, Eiji [1 ]
Matsunaga, Kenichi [1 ]
Sawada, Kinya [1 ]
Matsushita, Haruo [2 ]
Majima, Kazuhiro [1 ]
Jingu, Keiichi [2 ]
机构
[1] Takeda Gen Hosp, Dept Radiol, Aizu Wakamatsu, Fukushima, Japan
[2] Tohoku Univ, Grad Sch Med, Dept Radiat Oncol, Sendai, Miyagi, Japan
[3] Tokai Univ, Grad Sch Med, Dept Radiat Oncol, Isehara, Kanagawa, Japan
关键词
patient-specific QA; log file; Monte Carlo; volumetric-modulated arc therapy; radiotherapy; oncology; VMAT QA; ACCURACY; DELIVERY; RECONSTRUCTION; ALGORITHMS; IMRT; MLC;
D O I
10.1177/1533034617745250
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Log file-based methods are attracting increasing interest owing to their ability to validate volumetric-modulated arc therapy outputs with high resolution in the leaf and gantry positions and in delivered dose. Cross-validation of these methods for comparison with measurement-based methods using the ionization chamber/ArcCHECK-3DVH software (version 3.2.0) under the same conditions of treatment anatomy and plan enables an efficient evaluation of this method. In this study, with the purpose of cross-validation, we evaluate the accuracy of a log file-based method using Elekta log files and an X-ray voxel Monte Carlo dose calculation technique in the case of leaf misalignment during prostate volumetric-modulated arc therapy. In this study, 10 prostate volumetric-modulated arc therapy plans were used. Systematic multileaf collimator leaf positional errors (+/- 0.4 and +/- 0.8 mm for each single bank) were deliberately introduced into the optimized plans. Then, the delivered 3-dimensional doses to a phantom with a certain patient anatomy were estimated by our system. These doses were compared with the ionization chamber dose and the ArcCHECK-3DVH dose. For the given phantom and patient anatomy, the estimated dose strongly coincided with the ionization chamber/ArcCHECK-3DVH dose (P < .01). In addition, good agreement between the estimated dose and the ionization chamber/ArcCHECK-3DVH dose was observed. The dose estimation accuracy of our system, which combines Elekta log files and X-ray voxel Monte Carlo dose calculation, was evaluated.
引用
收藏
页码:1220 / 1225
页数:6
相关论文
共 23 条
[1]   Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries [J].
Agnew, A. ;
Agnew, C. E. ;
Grattan, M. W. D. ;
Hounsell, A. R. ;
McGarry, C. K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (09) :N49-N63
[2]   The impact of electron transport on the accuracy of computed dose [J].
Arnfield, MR ;
Siantar, CH ;
Siebers, J ;
Garmon, P ;
Cox, L ;
Mohan, R .
MEDICAL PHYSICS, 2000, 27 (06) :1266-1274
[3]   A Varian DynaLog file-based procedure for patient dose-volume histogram-based IMRT QA [J].
Calvo-Ortega, Juan F. ;
Teke, Tony ;
Moragues, Sandra ;
Pozo, Miquel ;
Casals, Joan .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02) :100-109
[4]   Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium [J].
Carrasco, P ;
Jornet, N ;
Duch, MA ;
Weber, L ;
Ginjaume, M ;
Eudaldo, T ;
Jurado, D ;
Ruiz, A ;
Ribas, M .
MEDICAL PHYSICS, 2004, 31 (10) :2899-2911
[5]   Parallel/Opposed: IMRT QA using treatment log files is superior to conventional measurement-based method [J].
Childress, Nathan ;
Chen, Quan ;
Rong, Yi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (01) :4-7
[6]   Sensitivity of volumetric modulated arc therapy patient specific QA results to multileaf collimator errors and correlation to dose volume histogram based metrics [J].
Coleman, Linda ;
Skourou, Christina .
MEDICAL PHYSICS, 2013, 40 (11)
[7]   Advanced kernel methods vs. Monte Carlo-based dose calculation for high energy photon beams [J].
Fotina, Irina ;
Winkler, Peter ;
Kuenzler, Thomas ;
Reiterer, Jochen ;
Simmat, Isabell ;
Georg, Dietmar .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (03) :645-653
[8]   Quality assurance of volumetric modulated arc therapy using Elekta Synergy [J].
Haga, Akihiro ;
Nakagawa, Keiichi ;
Shiraishi, Kenshiro ;
Itoh, Saori ;
Terahara, Atsuro ;
Yamashita, Hideomi ;
Ohtomo, Kuni ;
Saegusa, Shigeki ;
Imae, Toshikazu ;
Yoda, Kiyoshi ;
Pellegrini, Roberto .
ACTA ONCOLOGICA, 2009, 48 (08) :1193-1197
[9]   Tolerance levels of EPID-based quality control for volumetric modulated arc therapy [J].
Jorgensen, M. K. ;
Hoffmann, L. ;
Petersen, J. B. B. ;
Praestegaard, L. H. ;
Hansen, R. ;
Muren, L. P. .
MEDICAL PHYSICS, 2011, 38 (03) :1425-1434
[10]   Evaluation of patient DVH-based QA metrics for prostate VMAT: correlation between accuracy of estimated 3D patient dose and magnitude of MLC misalignment [J].
Kadoya, Noriyuki ;
Saito, Masahide ;
Ogasawara, Makoto ;
Fujita, Yukio ;
Ito, Kengo ;
Sato, Kiyokazu ;
Kishi, Kazuma ;
Dobashi, Suguru ;
Takeda, Ken ;
Jingu, Keiichi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03) :179-189