A Simple Strategy to Improve the Accuracy of the Injection Step in Batch Injection Analysis Systems with Amperometric Detection

被引:27
作者
Gimenes, Denise Tofanello [1 ]
Pereira, Polyana Fernandes [1 ]
Cunha, Rafael Rodrigues [1 ]
Bezerra da Silva, Rodrigo Amorim [1 ]
Abarza Munoz, Rodrigo Alejandro [1 ]
Richter, Eduardo Mathias [1 ]
机构
[1] Univ Fed Uberlandia, Inst Quim, Uberlandia, MG, Brazil
关键词
Reproducible manual injection; Internal standard method; Amperometric detection; BIA; BORON-DOPED DIAMOND; FLOW-INJECTION; ELECTROCHEMICAL DETECTION; ELECTRODES; PARACETAMOL; CYSTEINE; ETHANOL; TIME;
D O I
10.1002/elan.201200254
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this study, we describe for the first time the application of an internal standard method to compensate for random errors associated with the injection procedure in batch injection analysis (BIA) systems with multiple pulse amperometric detection. A sequence of potential pulses was selected in such a way that the internal standard (IS) compound was detected individually at one potential pulse and both the IS and analyte, were detected at another potential pulse. The current ratio (IIS+analyte/IIS) was used in the construction of the calibration curve and then to compensate for random errors. The use of disposable syringes or manual pipettes in BIA systems increases the robustness of the method and dispenses with skilled operators.
引用
收藏
页码:1805 / 1810
页数:6
相关论文
共 29 条
[1]   Batch injection analysis with amperometric detection: application for simultaneous analysis using a single working electrode [J].
Bezerra da Silva, Rodrigo Amorim ;
Gimenes, Denise Tofanello ;
Tormin, Thiago Faria ;
Abarza Munoz, Rodrigo Alejandro ;
Richter, Eduardo Mathias .
ANALYTICAL METHODS, 2011, 3 (12) :2804-2808
[2]   Development of a Simple and Fast Electrochemical Method to Evaluate Physical Stress in Athletes [J].
Bezerra da Silva, Rodrigo Amorim ;
Cunha, Rafael Rodrigues ;
Coelho da Silva, Abilio Tomaz ;
Bessa de Oliveira, Artur Luis ;
Espindola, Foued Salmen ;
Abarza Munoz, Rodrigo Alejandro ;
Richter, Eduardo Mathias .
ELECTROANALYSIS, 2011, 23 (11) :2601-2606
[3]   Thin-film gold electrodes produced by magnetron sputtering. Voltammetric characteristics and application in batch injection analysis with amperometric detection [J].
Brett, AMO ;
Matysik, FM ;
Vieira, MT .
ELECTROANALYSIS, 1997, 9 (03) :209-212
[4]  
Brett CMA, 1999, ELECTROANAL, V11, P1013, DOI 10.1002/(SICI)1521-4109(199910)11:14<1013::AID-ELAN1013>3.0.CO
[5]  
2-Y
[6]   AMPEROMETRIC BATCH INJECTION-ANALYSIS - THEORETICAL ASPECTS OF CURRENT TRANSIENTS AND COMPARISON WITH WALL-JET ELECTRODES IN CONTINUOUS-FLOW [J].
BRETT, CMA ;
BRETT, AMO ;
MITOSERIU, LC .
ELECTROANALYSIS, 1995, 7 (03) :225-229
[7]   SIMULTANEOUS DETERMINATION OF ASCORBIC ACID AND ACETYLSALICYLIC ACID USING FLOW INJECTION ANALYSIS WITH MULTIPLE PULSE AMPEROMETRIC DETECTION [J].
de Miranda, Joyce A. T. ;
Cunha, Rafael R. ;
Gimenes, Denise T. ;
Munoz, Rodrigo A. A. ;
Richter, Eduardo M. .
QUIMICA NOVA, 2012, 35 (07) :1459-1463
[8]   Simple Flow Injection Amperometric System for Simultaneous Determination of Dipyrone and Paracetamol in Pharmaceutical Formulations [J].
dos Santos, Wallans T. P. ;
Gimenes, Denise T. ;
de Almeida, Edimar G. N. ;
Eiras, Sebastiao de P. ;
Albuquerque, Yaico D. T. ;
Richter, Eduardo M. .
JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2009, 20 (07) :1249-1255
[9]   Flow-Injection Amperometric Method for Determination of Diclofenac in Pharmaceutical Formulations Using a Boron-Doped Diamond Electrode [J].
Gimenes, D. T. ;
de Freitas, J. M. ;
Munoz, R. A. A. ;
Richter, E. M. .
ELECTROANALYSIS, 2011, 23 (11) :2521-2525
[10]   Internal standard in flow injection analysis with amperometric detection [J].
Gimenes, Denise Tofanello ;
Pio dos Santos, Wallans Torres ;
Abarza Munoz, Rodrigo Alejandro ;
Richter, Eduardo Mathias .
ELECTROCHEMISTRY COMMUNICATIONS, 2010, 12 (02) :216-218