High-precision measurements of δ2H, δ18O and δ17O in water with the aid of cavity ring-down laser spectroscopy

被引:28
作者
Pierchala, Anna [1 ]
Rozanski, Kazimierz [1 ]
Dulinski, Marek [1 ]
Gorczyca, Zbigniew [1 ]
Marzec, Michal [1 ]
Czub, Robert [1 ]
机构
[1] AGH Univ Sci & Technol, Fac Phys & Appl Comp Sci, Krakow, Poland
关键词
Cavity ring-down spectroscopy; hydrogen-2; isotope measurements; methods and equipment; oxygen-17; excess; oxygen-18; QA; QC; uncertainty; ISOTOPE-RATIO; SPECTROMETER; CALIBRATION; O-17-EXCESS; SURFACES; OXYGEN;
D O I
10.1080/10256016.2019.1609959
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A thorough evaluation of measurement uncertainty together with control of short-term and long-term precision of measurements should be a basis of any successful quality assurance/quality control (QA/QC) strategy aimed at maintaining a high quality of the analytical process. Here we present the results of a comprehensive assessment of the analytical performance of a Picarro L2140-i CRDS laser spectrometer analysing delta H-2, delta O-18 and delta O-17 in water. The assessment is based on results obtained during 15 months of continuous operation of this instrument (February 2017 to May 2018). The short-term precision of measured and derived quantities was 0.11, 0.036, 0.028, 0.23 parts per thousand and 11 per meg, for delta H-2, delta O-18, delta O-17, d-excess and Delta O-17, respectively, and is comparable to the precision reported by the manufacturer. The long-term precision of the L2140-i, defined as standard uncertainty of the time series of 153 analyses of a laboratory standard conducted throughout 15 months, was roughly two times lower (0.24, 0.053, 0.038, 0.37 parts per thousand and 21 per meg, for delta H-2, delta O-18, delta O-17, d-excess and Delta O-17). In-depth assessment of the measurement uncertainty of a single analysis revealed that assigned uncertainty of the calibration standards is an important component of the uncertainty budget, especially in case of delta H-2 analysis.
引用
收藏
页码:290 / 307
页数:18
相关论文
共 31 条
[21]   THE INTERACTION OF WATER WITH SOLID-SURFACES - FUNDAMENTAL-ASPECTS [J].
THIEL, PA ;
MADEY, TE .
SURFACE SCIENCE REPORTS, 1987, 7 (6-8) :211-385
[22]   Optimization of instrument setup and post-run corrections for oxygen and hydrogen stable isotope measurements of water by isotope ratio infrared spectroscopy (IRIS) [J].
van Geldern, Robert ;
Barth, Johannes A. C. .
LIMNOLOGY AND OCEANOGRAPHY-METHODS, 2012, 10 :1024-1036
[23]   Seeking excellence: An evaluation of 235 international laboratories conducting water isotope analyses by isotope-ratio and laser-absorption spectrometry [J].
Wassenaar, L. I. ;
Terzer-Wassmuth, S. ;
Douence, C. ;
Araguas-Araguas, L. ;
Aggarwal, P. K. ;
Coplen, T. B. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2018, 32 (05) :393-406
[24]   Approaches for Achieving Long-Term Accuracy and Precision of δ18O and δ2H for Waters Analyzed using Laser Absorption Spectrometers [J].
Wassenaar, Leonard I. ;
Coplen, Tyler B. ;
Aggarwal, Pradeep K. .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2014, 48 (02) :1123-1131
[25]   Cavity ring-down spectroscopy [J].
Wheeler, MD ;
Newman, SM ;
Orr-Ewing, AJ ;
Ashfold, MNR .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS, 1998, 94 (03) :337-351
[26]  
[No title captured]
[27]  
[No title captured]
[28]  
[No title captured]
[29]  
[No title captured]
[30]  
[No title captured]