High-precision measurements of δ2H, δ18O and δ17O in water with the aid of cavity ring-down laser spectroscopy

被引:28
作者
Pierchala, Anna [1 ]
Rozanski, Kazimierz [1 ]
Dulinski, Marek [1 ]
Gorczyca, Zbigniew [1 ]
Marzec, Michal [1 ]
Czub, Robert [1 ]
机构
[1] AGH Univ Sci & Technol, Fac Phys & Appl Comp Sci, Krakow, Poland
关键词
Cavity ring-down spectroscopy; hydrogen-2; isotope measurements; methods and equipment; oxygen-17; excess; oxygen-18; QA; QC; uncertainty; ISOTOPE-RATIO; SPECTROMETER; CALIBRATION; O-17-EXCESS; SURFACES; OXYGEN;
D O I
10.1080/10256016.2019.1609959
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A thorough evaluation of measurement uncertainty together with control of short-term and long-term precision of measurements should be a basis of any successful quality assurance/quality control (QA/QC) strategy aimed at maintaining a high quality of the analytical process. Here we present the results of a comprehensive assessment of the analytical performance of a Picarro L2140-i CRDS laser spectrometer analysing delta H-2, delta O-18 and delta O-17 in water. The assessment is based on results obtained during 15 months of continuous operation of this instrument (February 2017 to May 2018). The short-term precision of measured and derived quantities was 0.11, 0.036, 0.028, 0.23 parts per thousand and 11 per meg, for delta H-2, delta O-18, delta O-17, d-excess and Delta O-17, respectively, and is comparable to the precision reported by the manufacturer. The long-term precision of the L2140-i, defined as standard uncertainty of the time series of 153 analyses of a laboratory standard conducted throughout 15 months, was roughly two times lower (0.24, 0.053, 0.038, 0.37 parts per thousand and 21 per meg, for delta H-2, delta O-18, delta O-17, d-excess and Delta O-17). In-depth assessment of the measurement uncertainty of a single analysis revealed that assigned uncertainty of the calibration standards is an important component of the uncertainty budget, especially in case of delta H-2 analysis.
引用
收藏
页码:290 / 307
页数:18
相关论文
共 31 条
[1]  
[Anonymous], 2009, TRACERS HYDROLOGY, DOI DOI 10.1002/9780470747148
[2]  
[Anonymous], 1998, ISOTOPE TRACERS CATC, DOI DOI 10.1016/C2009-0-10239-8
[3]   Inter-comparison of salt effect correction for δ18O and δ2H measurements in seawater by CRDS and IRMS using the gas-H2O equilibration method [J].
Benetti, M. ;
Sveinbjornsdottir, A. E. ;
Olafsdottir, R. ;
Leng, M. J. ;
Arrowsmith, C. ;
Debondt, K. ;
Fripiat, F. ;
Aloisi, G. .
MARINE CHEMISTRY, 2017, 194 :114-123
[4]   Cavity ring-down spectroscopy: Experimental schemes and applications [J].
Berden, G ;
Peeters, R ;
Meijer, G .
INTERNATIONAL REVIEWS IN PHYSICAL CHEMISTRY, 2000, 19 (04) :565-607
[5]   Measurement of δ18O, δ17O, and 17O-excess in Water by Off-Axis Integrated Cavity Output Spectroscopy and Isotope Ratio Mass Spectrometry [J].
Berman, Elena S. F. ;
Levin, Naomi E. ;
Landais, Amaelle ;
Li, Shuning ;
Owano, Thomas .
ANALYTICAL CHEMISTRY, 2013, 85 (21) :10392-10398
[6]  
Cook PG, 2000, ENV TRACERS SUBSURFA
[7]   Residence times of water molecules on stainless steel and aluminum surfaces in vacuum and atmosphere [J].
Dobrozemsky, R. ;
Menhart, S. ;
Buchtela, K. .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A, 2007, 25 (03) :551-556
[8]  
Ellison SLR., 2012, Quantifying uncertainty in analytical measurement
[9]   A continuous stream flash evaporator for the calibration of an IR cavity ring-down spectrometer for the isotopic analysis of water [J].
Gkinis, Vasileios ;
Popp, Trevor J. ;
Johnsen, Sigfus J. ;
Blunier, Thomas .
ISOTOPES IN ENVIRONMENTAL AND HEALTH STUDIES, 2010, 46 (04) :463-475
[10]   Improved water δ2H and δ18O calibration and calculation of measurement uncertainty using a simple software tool [J].
Groening, Manfred .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2011, 25 (19) :2711-2720