Biomechanical Evaluation of a Modified Dorsal Double-plating Fixation with Adjustable Joint and Microthread Designs for Comminuted Extra-articular Distal Radius Fractures

被引:0
作者
Lin, Yu-Hao [1 ]
Chen, Alvin Chao-Yu [2 ]
Kuo, Hsien-Nan [3 ]
Yu, Tsung-Chih [3 ]
Sun, Ming-Tsung [1 ]
Lin, Chun-Li [4 ]
机构
[1] Chang Gung Univ, Dept Mech Engn, Tao Yuan 333, Taiwan
[2] Chang Gung Mem Hosp, Dept Orthopaed Surg, Tao Yuan 333, Taiwan
[3] Met Ind Res & Dev Ctr, Med Devices Dev Sect, Kaohsiung 811, Taiwan
[4] Natl Yang Ming Univ, Dept Biomed Engn, Taipei 112, Taiwan
关键词
Biomechanics; Distal radius fractures; Dorsal double-plating; Fatigue test; Stiffness; INTERNAL-FIXATION; EXTERNAL FIXATION; FINITE-ELEMENT; T-PLATES; VOLAR; LOCKING; MODEL; ANGLE; STABILITY; SCREWS;
D O I
10.5405/jmbe.990
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study evaluates the biomechanical strength of a modified dorsal double-plating (MDDP) internal fixation on artificial radius bones using the dynamic fatigue test. The plates are 1.6 mm thick and the upper L plate has an adjustable joint to allow the angular disposition of the bone plate to be changed to secure various bone fracture types. A dual-thread locking screw is used to enhance mechanical retention at the screw/bone interface. A finite element analysis is performed to examine the mechanical bone and fixation system response using a dual-microthread locking screw. An MDDP fixation is made with stainless steel by a manufacturer with ISO13485 quality management systems. Eighteen radius Sawbones were randomly placed into three groups and cut to a standard length of 2.5 cm from the articular surface to form 0 degrees, 30 degrees, and 60 degrees fracture configurations according to the A2 classification in the AO surgery reference. The adjustable joint in the L plates is adjusted and fixed using a fastener connection to have a 0 degrees, 30 degrees, or 60 degrees configuration. Plates and screws are then positioned on the dorso-ulnar side for L plates and the dorso-radial side for I plates to create an angle of 90 degrees. The specimens were subjected to oscillating loads of 10 to 150 N at 5 Hz for 20000 cycles. The average stiffness values after 20000 cycles were 425.7, 461.1, and 532.1 N/mm for the 0 degrees, 30 degrees, and 60 degrees constructs, respectively. No difference in stiffness was found for constructs with a given angle throughout the 20000 cycles of testing (p > 0.05). However, significant differences (p < 0.05) in stiffness were found between constructs with different angles at each 500-cycle interval. The lack of gross construct failure during cyclic testing and the reasonable stiffness prove that MDDP internal fixation is sufficiently stable to support restricted postoperative loads.
引用
收藏
页码:29 / 34
页数:6
相关论文
共 33 条
  • [1] Intramedullary nails with two lag screws
    Brown, CJ
    Wang, CJ
    Yettram, AL
    Procter, P
    [J]. CLINICAL BIOMECHANICS, 2004, 19 (05) : 519 - 525
  • [2] Displaced intra-articular fractures of the distal aspect of the radius - Long-term results in young adults after open reduction and internal fixation
    Catalano, LW
    Cole, RJ
    Gelberman, RH
    Evanoff, BA
    Gilula, LA
    Borrelli, J
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (09) : 1290 - 1302
  • [3] Assessment of stiffness and strength of 4 different implants available for equine fracture treatment:: A study on a 20° oblique long-bone fracture model using a bone substitute
    Florin, M
    Arzdorf, M
    Linke, B
    Auer, JA
    [J]. VETERINARY SURGERY, 2005, 34 (03) : 231 - 238
  • [4] Biomechanical Comparison of Locking versus Nonlocking Volar and Dorsal T-Plates for Fixation of Dorsally Comminuted Distal Radius Fractures
    Gondusky, Joseph S.
    Carney, Joseph
    Erpenbach, Jonathan
    Robertson, Claire
    Mahar, Andrew
    Oka, Richard
    Thompson, Michael
    Mazurek, Michael
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 (01) : 44 - 50
  • [5] FRACTURE OF THE DISTAL RADIUS - CLASSIFICATION OF TREATMENT AND INDICATIONS FOR EXTERNAL FIXATION
    GRAFF, S
    JUPITER, J
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1995, 25 : SD14 - SD25
  • [6] GREATTING MD, 1993, ORTHOP CLIN N AM, V24, P301
  • [7] Fractures of the distal radius treated by internal fixation and early function - A prospective study of 73 consecutive patients
    Jakob, M
    Rikli, DA
    Regazzoni, P
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (03): : 340 - 344
  • [8] Jupiter, 1997, J Am Acad Orthop Surg, V5, P119
  • [9] FRACTURES OF THE DISTAL END OF THE RADIUS
    JUPITER, JB
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (03) : 461 - 469
  • [10] Does a volar locking plate provide equivalent stability as a dorsal nonlocking plate in a dorsally comminuted distal radius fracture? a biomechanical study
    Kandemir, Utku
    Matityahu, Amir
    Desai, Rohan
    Puttlitz, Christian
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (09) : 605 - 610