Immunohistochemistry Validation Procedures and Practices A College of American Pathologists Survey of 727 Laboratories

被引:34
作者
Hardy, Lindsay B. [1 ]
Fitzgibbons, Patrick L. [2 ]
Goldsmith, Jeffery D. [1 ]
Eisen, Richard N. [3 ]
Beasley, Mary Beth [4 ]
Souers, Rhona J. [5 ]
Nakhleh, Raouf E. [6 ]
机构
[1] Beth Israel Deaconess Med Ctr, Dept Pathol, Boston, MA 02215 USA
[2] St Jude Med Ctr, Dept Pathol, Fullerton, CA USA
[3] Greenwich Hosp, Dept Pathol, Greenwich, CT USA
[4] Mt Sinai Med Ctr, Dept Pathol, New York, NY 10029 USA
[5] Coll Amer Pathologists, Northfield, IL USA
[6] Mayo Clin Florida, Dept Pathol, Jacksonville, FL 32224 USA
关键词
D O I
10.5858/arpa.2011-0676-CP
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Context.-The immunohistochemistry (IHC) laboratory represents a dynamic area of surgical pathology with limited practice guidelines. Studies have shown significant interlaboratory variability in results. Objective.-To establish baseline parameters for IHC validation procedures and practice, and to assess their feasibility of implementation. Design.-In September 2010, a questionnaire was distributed by the College of American Pathologists. It was composed of 32 questions relating to nonpredictive assays as well as non-US Food and Drug Administration (non-FDA)-approved, predictive IHC assays other than human epidermal growth factor 2 (HER2/neu). Results.-For non-FDA approved, nonpredictive IHC assays, 68% of laboratories had a written validation procedure. Eighty-six percent of laboratories validated the most recently introduced nonpredictive antibody. Seventy-five percent used 21 or fewer total cases for the validation and 40% used weakly or focally positive cases. Forty-six percent of respondents had a written procedure for validation procedures for non-FDA approved, predictive marker IHC assays other than HER2/neu. Seventy-five percent of laboratories validated the most recently introduced predictive antibody other than HER2/neu. Fewer than half used 25 or more cases for the validation, and 47% used weakly or focally positive cases. Conclusion.-Some laboratories have written validation procedures that appear to build upon HER2/neu testing guidelines. Some laboratories also manage to validate new antibodies according to those standards; however, many do not. There appears to be a need for further validation guideline development for nonpredictive and non-FDA approved predictive antibody IHC assays. (Arch Pathol Lab Med. 2013;137:19-25; doi: 10.5858/arpa.2011-0676-CP)
引用
收藏
页码:19 / 25
页数:7
相关论文
共 11 条
[1]  
[Anonymous], ILA28A2 CLSI
[2]  
Clinical and Laboratory Standards Institute, 2010, MM06A2 CLSI
[3]  
CLSI/NCCLS, 2002, EP19R CLSINCCLS
[4]  
Fitzgibbons PL, 2010, ARCH PATHOL LAB MED, V134, P930, DOI 10.1043/1543-2165-134.6.930
[5]  
Hammond MEH, 2010, ARCH PATHOL LAB MED, V134, P907, DOI [10.5858/134.6.907, 10.1200/JCO.2009.25.6529, 10.1200/JOP.777003]
[6]  
Nakhleh RE, 2010, ARCH PATHOL LAB MED, V134, P728, DOI 10.1043/1543-2165-134.5.728
[7]   Real-world performance of HER2 testing - National surgical adjuvant breast and bowel project experience [J].
Paik, S ;
Bryant, J ;
Tan-Chiu, E ;
Romond, E ;
Hiller, W ;
Park, K ;
Brown, A ;
Yothers, G ;
Anderson, S ;
Smith, R ;
Wickerham, DL ;
Wolmark, N .
JNCI-JOURNAL OF THE NATIONAL CANCER INSTITUTE, 2002, 94 (11) :852-854
[8]   Concordance between local and central laboratory HER2 testing in the breast intergroup trial N9831 [J].
Roche, PC ;
Suman, VJ ;
Jenkins, RB ;
Davidson, NE ;
Martino, S ;
Kaufman, PA ;
Addo, FK ;
Murphy, B ;
Ingle, JN ;
Perez, EA .
JNCI-JOURNAL OF THE NATIONAL CANCER INSTITUTE, 2002, 94 (11) :855-857
[9]  
Taylor CR, 1999, AM J CLIN PATHOL, V111, P443
[10]  
Wolff AC, 2007, ARCH PATHOL LAB MED, V131, P18, DOI 10.1043/1543-2165(2007)131[18:ASOCCO]2.0.CO