The BOLD sensitivity of rapid steady-state sequences

被引:10
作者
Scheffler, Klaus [1 ,2 ]
Heule, Rahel [1 ]
Baez-Yanez, Mario G. [1 ,3 ]
Kardatzki, Bernd [2 ]
Lohmann, Gabriele [1 ,2 ]
机构
[1] Max Planck Inst Biol Cybernet, High Field MR Ctr, Tubingen, Germany
[2] Univ Tubingen, Dept Biomed Magnet Resonance, Tubingen, Germany
[3] Univ Tubingen, Grad Training Ctr Neurosci, Tubingen, Germany
关键词
blood oxygenation; diffusion; steady-state free precession; vessel size; SPIN-ECHO FMRI; GRADIENT-ECHO; 7; T; INTRAVASCULAR CONTRIBUTION; VISUAL-CORTEX; BALANCED SSFP; HUMAN BRAIN; BLOOD; DIFFUSION; MODEL;
D O I
10.1002/mrm.27585
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Relaxation and dephasing of water protons embedded in a vascular network is driven by local magnetic field inhomogeneities around deoxygenated blood vessels. These effects strongly depend on the relation between mean diffusion length and diameter of blood vessels, as well as on the chosen imaging sequence. In this work, the BOLD sensitivity of steady-state sequences as a function of vessel size, field strength, and sequence parameters are analyzed. Methods: Steady-state magnetization within a network of artificial cylinders is simulated with Monte Carlo methods for different coherence pathways. In addition, measurements on microspheres were performed to confirm theoretical results. Results: Simulations and phantom results demonstrate a vessel size-dependent signal attenuation effect of all coherence pathways. Both the FID and ECHO pathways show a signal profile similar to spin echo sequences where in the static dephasing regime the effect of larger vessels is suppressed. Conclusion: The BOLD effect measured in steady-state sequences is most sensitive to microvessels and might therefore be closer to the underlying neuronal event compared to gradient echo sequence
引用
收藏
页码:2526 / 2535
页数:10
相关论文
共 50 条
  • [1] THE DIFFUSION SENSITIVITY OF FAST STEADY-STATE FREE PRECESSION IMAGING
    BUXTON, RB
    MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (02) : 235 - 243
  • [2] Analytical Solution to the Transient Phase of Steady-State Free Precession Sequences
    Ganter, Carl
    MAGNETIC RESONANCE IN MEDICINE, 2009, 62 (01) : 149 - 164
  • [3] Intrinsic diffusion sensitivity of the balanced steady-state free precession (bSSFP) imaging sequence
    Baer, Sebastien
    Weigel, Matthias
    von Elverfeldt, Dominik
    Hennig, Juergen
    Leupold, Jochen
    NMR IN BIOMEDICINE, 2015, 28 (11) : 1383 - 1392
  • [5] Discovery, development and exploitation of steady-state biofilms
    Greenman, John
    Hewett, Keith
    Saad, Saliha
    JOURNAL OF BREATH RESEARCH, 2020, 14 (04)
  • [6] Stabilization of alternating TR steady-state free precession sequences
    Lee, Hsu-Lei
    Nayak, Krishna S.
    JOURNAL OF MAGNETIC RESONANCE, 2008, 195 (02) : 211 - 218
  • [7] SIMULTANEOUS CALCULATION OF FLOW AND DIFFUSION SENSITIVITY IN STEADY-STATE FREE PRECESSION IMAGING
    GUDBJARTSSON, H
    PATZ, S
    MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (04) : 567 - 579
  • [8] Modeling and suppression of respiration induced B0-fluctuations in non-balanced steady-state free precession sequences at 7 Tesla
    Goa, Pal Erik
    Poser, Benedikt Andreas
    Barth, Markus
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2013, 26 (04) : 377 - 387
  • [9] Steady-state and non-steady state operation of counter-current chromatography devices
    Kostanyan, Artak E.
    Ignatova, Svetlana N.
    Sutherland, Ian A.
    Hewitson, Peter
    Zakhodjaeva, Yulya A.
    Erastov, Andrey A.
    JOURNAL OF CHROMATOGRAPHY A, 2013, 1314 : 94 - 105
  • [10] Steady-state models of photosynthesis
    von Caemmerer, Susanne
    PLANT CELL AND ENVIRONMENT, 2013, 36 (09) : 1617 - 1630