Intraoperative radiofrequency ablation using a loop internally cooled-perfusion electrode:: In vitro and in vivo experiments

被引:3
|
作者
Lee, JM
Han, JK
Eoh, H
Kim, SH
Lee, JY
Lee, MW
Choi, BI
机构
[1] Seoul Natl Univ Hosp, Dept Diagnost Radiol, Clin Res Inst, Seoul 110744, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Radiol, Seoul, South Korea
[3] Seoul Natl Univ, Coll Med, Inst Radiat Med, Seoul, South Korea
关键词
liver; interventional procedures; radio-frequency ablation; experimental study;
D O I
10.1016/j.jss.2005.11.584
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective. We sought to validate the efficiency of intraoperative radiofrequency ablation (RFA) using a prototype loop internally cooled-perfusion (LICP) electrode to induce coagulation in the subcapsular portion of the liver. Materials and methods. In in vitro experiments, 30 ablation regions were created using a 200-W generator and a LICP electrode featured simultaneous intraelectrode cooling and continuous flow of hypertonic saline along the shaft in explanted bovine liver. In the in vivo experiments, 26 ablation zones were created according to one of the five protocols in 10 dogs: group A, RFA using a cooled-tip electrode (n = 6); group B, RFA using a LICP electrode with 2 cm loop tip (n = 6); group C, RFA using a LICP electrode with a 3-cm loop tip (n = 6); group D, RFA using a cooled-tip electrode and Pringle maneuver (n = 4); and group E, RFA using a LICP electrode with 2 cm loop tip and Pringle maneuver (n = 4). The dimensions of the coagulation parameters were compared between the groups. Results. In the in vitro experiments, RFA using a 2- or 3-cm diameter LICP electrode (3.6 +/- 0.3 cm, 3.4 +/- 0.5 cm, respectively) created deeper dimensions of coagulation than did a 4-cm electrode (2.3 +/- 0.2 cm; P < 0.05). In the in vivo experiments, the RITA using the LICP electrodes in porcine liver with normal perfusion induced wider but superficial ablation regions when compared to standard RFA using an internally cooled electrode (P < 0.05). However, using a Pringle maneuver, RFA with a LICP electrode created a larger volume of ablation area when compared to RFA using an internally cooled electrode with a similar range of axial diameter along the electrode axis: 30.0 +/- 6.1 cm(3) (group E; P < 0.05) (group D) versus 68.5 +/- 14.0 cm(3) Conclusions. Intraoperative RFA using the LICP electrode induced a well-defined semicircular coagulation with a 3.5-cm axial diameter in the subcapsular region of the liver. This device appears to be promising for the treatment of superficial tumors during intraoperative RFA. (C) 2006 Elsevier Inc. All rights reserved.
引用
收藏
页码:215 / 224
页数:10
相关论文
共 50 条
  • [41] Optimization of wet radiofrequency ablation using a perfused-cooled electrode:: A comparative study in Ex vivo bovine livers
    Lee, JM
    Han, JK
    Kim, SH
    Lee, Y
    Shin, KS
    Han, CJ
    Lee, MW
    Choi, JI
    Choi, BI
    KOREAN JOURNAL OF RADIOLOGY, 2004, 5 (04) : 250 - 257
  • [42] Comparison of ablation performance between dual internally cooled wet tip and conventional dual internally cooled tip radiofrequency electrodes: an experimental study in ex vivo bovine liver
    Cha, Dong Ik
    Lee, Min Woo
    Jeong, Woo Kyoung
    Ha, Sang Yun
    Ahn, Soo Hyun
    Rhim, Hyunchul
    Lim, Hyo Keun
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2021, 38 (01) : 332 - 340
  • [43] Hepatic radiofrequency ablation: in vivo and ex vivo comparisons of 15-gauge (G) and 17-G internally cooled electrodes
    Song, K. D.
    Lee, M. W.
    Park, H. J.
    Cha, D. I.
    Kang, T. W.
    Lee, J.
    Moon, J. Y.
    Rhim, H.
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1050):
  • [44] The loop electrode: In vitro evaluation of a device for ultrasound-guided interstitial tissue ablation using radiofrequency electrosurgery
    Lorentzen, T
    ACADEMIC RADIOLOGY, 1996, 3 (03) : 219 - 224
  • [45] Increasing radiofrequency ablation volumes with the use of internally cooled electrodes and injected hydrochloric acid in ex vivo bovine livers
    Jiang, Xiong-Ying
    Zhang, Tian-Qi
    Li, Guo
    Gu, Yang-Kui
    Gao, Fei
    Yao, Wang
    Zhang, Yan-yang
    Huang, Jin-hua
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2019, 35 (01) : 37 - 43
  • [46] Utility of tip thermometry during radiofrequency ablation in humans using an internally perfused saline cooled catheter
    Wharton, JM
    Wilber, DJ
    Calkins, H
    Epstein, AE
    Ellenbogen, KA
    Carlson, MD
    Stevenson, WG
    Packer, DL
    Hummel, J
    Stark, S
    Kopelman, H
    Tohou, PJ
    CIRCULATION, 1997, 96 (08) : 1775 - 1775
  • [47] Multipolar radiofrequency ablation using internally cooled electrodes in ex vivo bovine liver: Correlation between volume of coagulation and amount of applied energy
    Clasen, Stephan
    Rempp, Hansjoerg
    Schmidt, Diethard
    Schraml, Christina
    Hoffmann, Ruediger
    Claussen, Claus D.
    Pereira, Philippe L.
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (01) : 111 - 113
  • [48] Radiofrequency ablation with a new perfused-cooled electrode using a single pump:: An experimental study in Ex vivo bovine liver
    Kim, SK
    Se, JW
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 28 (06) : 779 - 788
  • [49] Radiofrequency Ablation with a New Perfused-Cooled Electrode Using a Single Pump: An Experimental Study in Ex Vivo Bovine Liver
    Seung Kwon Kim
    Jung Wook Seo
    CardioVascular and Interventional Radiology, 2005, 28 : 779 - 788
  • [50] Percutaneous radiofrequency ablation of HCC: reduced ablation duration and increased ablation size using single, internally cooled electrodes with an optimized pulsing algorithm
    Solbiati, Luigi
    Ierace, Tiziana
    Gennaro, Nicolo
    Muglia, Riccardo
    Cosman, Eric R., Jr.
    Goldberg, S. Nahum
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2020, 37 (01) : 861 - 867