A high-density multichannel surface electromyography system for the characterization of single motor units

被引:93
作者
Blok, JH [1 ]
van Dijk, JP [1 ]
Drost, G [1 ]
Zwarts, MJ [1 ]
Stegeman, DF [1 ]
机构
[1] Univ Med Ctr Nijmegen, Inst Neurol, Dept Clin Neurophysiol, Nijmegen, Netherlands
关键词
D O I
10.1063/1.1455134
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An electromyography (EMG) system is presented that noninvasively records the electrical activity of a muscle with 126 densely spaced skin-surface electrodes. The electrodes are arranged in a two-dimensional array and integrated in a single container for ease of application. Signals are recorded "monopolarly", with a reference electrode placed at a distance from the array. With this recording configuration, the surface EMG (sEMG) potential distribution can be described not only as a function of time, but also topographically. The availability of topographical information opens up a range of applications. Some of these have been described previously. However, the system presented is unique in that it allows exploration of all clinical and scientific possibilities of topographical sEMG. In its design, special attention was paid to user-friendliness and flexibility. With high-density multichannel sEMG, both the properties of a whole muscle and those of single motor units, the functional units of a muscle, can be studied. The latter belong to a realm that was long considered accessible only with needle-EMG, a conventional, invasive diagnostic technique. It is demonstrated that the additional topographical information can be used to characterize motor units in a way that is partially superior to needle EMG. (C) 2002 American Institute of Physics.
引用
收藏
页码:1887 / 1897
页数:11
相关论文
共 46 条
[1]  
[Anonymous], HUMAN PHYSL
[2]  
BLOK JH, 1999, CLIN NEUROPHYSIOL, V110, P90
[3]  
BLOK JH, 1997, STATE ART SENSORS SE
[4]  
BLOK JH, UNPUB ANN BIOMED ENG
[5]  
D F., 2001, European Journal of Physical and Rehabilitation Medicine (EUROPA MEDICOPHYSICA), V37, P57
[6]   Non-invasive detection of the single motor unit action potential by averaging the spatial potential distribution triggered on a spatially filtered motor unit action potential [J].
Disselhorst-Klug, C ;
Rau, G ;
Schmeer, A ;
Silny, J .
JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 1999, 9 (01) :67-72
[7]   Propagation disturbance of motor unit action potentials during transient paresis in generalized myotonia - A high-density surface EMG study [J].
Drost, G ;
Blok, JH ;
Stegeman, DF ;
van Dijk, JP ;
van Engelen, BGM ;
Zwarts, MJ .
BRAIN, 2001, 124 :352-360
[8]  
Dumitru D, 2002, Electrodiagnostic medicine, V2nd
[9]  
Huppertz HJ, 1997, MUSCLE NERVE, V20, P1360, DOI 10.1002/(SICI)1097-4598(199711)20:11<1360::AID-MUS3>3.0.CO
[10]  
2-8