Experimental Evaluation of High Intensity Ultrasound Source System using Acoustic Waveguide and Concave Transducer with 100 mm Diameter for Calibration of Hydrophone

被引:0
作者
Igarashi, Shigeru [1 ]
Morishita, Takeshi [2 ]
Takeuchi, Shinichi [2 ]
机构
[1] Polytech Univ, Elect Circuit Engn Unit, Kodaira, Tokyo, Japan
[2] Toin Univ Yokohama, Fac Biomed Engn, Dept Clin Engn, Yokohama, Kanagawa, Japan
来源
2018 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS) | 2018年
关键词
Ultrasound source; Acoustic waveguide; Concave Transducer; Calibration; Hydrophone;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Recently, equipments using high intensity ultrasound have been increasing in medical and industrial fields. A high intensity ultrasound source for calibrating a hydrophone to evaluate a high intensity ultrasound field is required. Especially, to calibrate the receiving sensitivity of the membrane type hydrophone as the primary standard, it is necessary to vibrate the required area of acoustic receiving surface on the primary standard hydrophone uniformly. We proposed and developed a high intensity ultrasound source system using a cylindrical acoustic waveguide and a concave transducer. In this system, the output ultrasound wave forms main beam similar to that formed by a flat type transducer and is expected to increase the ultrasound power acoustically by using of the concave transducer with larger diameter. We constructed an experimental system using an 1 MHz concave transducer with an effective diameter of 40 mm, and confirmed the peak acoustic pressure of 670 kPa at the position of 10 mm from the transmitting aperture. Furthermore, we constructed an experimental system using a concave transducer with an effective diameter of 100 mm, and measured the acoustic pressure distribution on the lateral axis at distances of 10, 20, and 30 mm from the transmitting aperture. As the result, it was confirmed that the -6 dB beam width become almost equal to the conventional source system with the effective diameter of 40 mm and the transmitting sensitivity increased about 2-fold with that.
引用
收藏
页数:4
相关论文
共 21 条
  • [1] [Anonymous], 2007, 621272 IEC
  • [2] Dimitescu A., 2018, MATER SCI ENG, V400
  • [3] Synthesis of graphene oxide/polyaniline nanocomposite for measuring cadmium(II) by solid phase extraction combined with dispersive liquid-liquid microextraction
    Farajvand, Mohammad
    Farajzadeh, Kamran
    Faghani, Gholamreza
    [J]. MATERIALS RESEARCH EXPRESS, 2018, 5 (07)
  • [4] Effect of controlled offset of focal position in cavitation-enhanced high-intensity focused ultrasound treatment
    Goto, Kota
    Takagi, Ryo
    Miyashita, Takuya
    Jimbo, Hayato
    Yoshizawa, Shin
    Umemura, Shin-ichiro
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS, 2015, 54 (07)
  • [5] Bubble size distribution analysis and control in high frequency ultrasonic cleaning processes
    Hauptmann, M.
    Struyf, H.
    Mertens, P.
    Heyns, M.
    De Gendt, S.
    Glorieux, C.
    Brems, S.
    [J]. INTERNATIONAL SYMPOSIUM ON ULTRASOUND IN THE CONTROL OF INDUSTRIAL PROCESSES (UCIP 2012), 2012, 42
  • [6] Igarashi S., 2015, P ULTR EL S USE, V36, p1P5
  • [7] Igarashi S., 2016, J ACOUST SOC AM, V140, p5pEA1
  • [8] Igarashi S., 2016, P ULTR EL S USE, V37
  • [9] Experimental evaluation of high-intensity ultrasound source system using acoustic waveguide for calibration of hydrophone
    Igarashi, Shigeru
    Morishita, Takeshi
    Uchida, Takeyoshi
    Takeuchi, Shinichi
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS, 2017, 56 (07)
  • [10] Prediction of thermal coagulation from the instantaneous strain distribution induced by high-intensity focused ultrasound
    Iwasaki, Ryosuke
    Takagi, Ryo
    Tomiyasu, Kentaro
    Yoshizawa, Shin
    Umemura, Shin-ichiro
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS, 2017, 56 (07)