Full waveform inversion for ultrasound computed tomography with high-sensitivity scan method

被引:1
作者
Suzuki, Atsuro [1 ]
Tsubota, Yushi [1 ]
Wu, Wenjing [1 ]
Yamanaka, Kazuhiro [1 ]
Terada, Takahide [1 ]
Kawabata, Kenichi [1 ]
机构
[1] Hitachi Ltd, Res & Dev Grp, Tokyo, Japan
来源
MEDICAL IMAGING 2019: ULTRASONIC IMAGING AND TOMOGRAPHY | 2019年 / 10955卷
关键词
Ultrasound CT; breast imaging; full waveform inversion; FREQUENCY-DOMAIN;
D O I
10.1117/12.2512639
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For breast cancer imaging by ultrasound computed tomography (CT) without dependence on patient breast size, we previously developed a high-sensitivity scan method in which a virtual fan-beam (vfan-beam) is generated from ultrasound waves emitted from 128 sources with unique delay times. Full waveform inversion (FWI) with multiple sound sources has not been previously applied to ultrasound CT using a ring transducer array. We have now developed a FWI calculation process that enables a vfan-beam to generate accurate sound speed images. A vfan-beam is accurately modeled by positioning the 128 sources and considering the delay times. The performance of the FWI calculation process with a vfan-beam was evaluated using a prototype ultrasound CT. For a circular phantom, the spatial resolution of a FWI image obtained with a vfan-beam was better than that of a filtered back projection (FBP) image. The image contrast of the FWI calculation process with a vfan-beam was comparable to that of the process with a conventional fan-beam generated from a single source. For a high-attenuation ellipse phantom, the sound speed image obtained with a conventional fan-beam had severe artifacts due to the low signal to noise ratio (SNR). Using a vfan-beam reduced the number of artifacts in the images due to the higher SNR. The FWI calculation process with a vfan-beam visualized a 3 mm inclusion more clearly than the FBP process. A measurement study demonstrated that the FWI process with a vfan-beam with a high SNR reduced the number of artifacts in the sound speed images and improved the spatial resolution for a high-attenuation breast.
引用
收藏
页数:6
相关论文
共 7 条
[1]   COPOLYMER-IN-OIL PHANTOM MATERIALS FOR ELASTOGRAPHY [J].
Oudry, J. ;
Bastard, C. ;
Miette, V. ;
Willinger, R. ;
Sandrin, L. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2009, 35 (07) :1185-1197
[2]   Seismic waveform inversion in the frequency domain, Part 1: Theory and verification in a physical scale model [J].
Pratt, RG .
GEOPHYSICS, 1999, 64 (03) :888-901
[3]   Frequency domain ultrasound waveform tomography: breast imaging using a ring transducer [J].
Sandhu, G. Y. ;
Li, C. ;
Roy, O. ;
Schmidt, S. ;
Duric, N. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (14) :5381-5398
[4]   Oil-gel-based phantom for mimicking wave refraction of breast in ultrasound computed tomography [J].
Suzuki, Atsuro ;
Tsubota, Yushi ;
Wu, Wenjing ;
Yamanaka, Kazuhiro ;
Terada, Takahide ;
Kawabata, Kenichi .
MEDICAL IMAGING 2018: ULTRASONIC IMAGING AND TOMOGRAPHY, 2018, 10580
[5]   Highly precise acoustic calibration method of ring-shaped ultrasound transducer array for plane-wave-based ultrasound tomography [J].
Terada, Takahide ;
Yamanaka, Kazuhiro ;
Suzuki, Atsuro ;
Tsubota, Yushi ;
Wu, Wenjing ;
Kawabata, Ken-ichi .
JAPANESE JOURNAL OF APPLIED PHYSICS, 2017, 56 (07)
[6]   3-D Nonlinear Acoustic Inverse Scattering: Algorithm and Quantitative Results [J].
Wiskin, J. W. ;
Borup, D. T. ;
Iuanow, E. ;
Klock, J. ;
Lenox, Mark W. .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2017, 64 (08) :1161-1174
[7]  
Wu W., 2019, SPIE MED IMAGI UNPUB