A comparative analysis of four activated clotting time measurement devices in cardiac surgery with cardiopulmonary bypass

被引:18
作者
Li, Han [1 ]
Serrick, Cyril [1 ]
Rao, Vivek [2 ,3 ]
Yip, Paul M. [3 ,4 ]
机构
[1] Univ Hlth Network, Perfus Serv, Toronto, ON, Canada
[2] Univ Hlth Network, Cardiovasc Surg, Toronto, ON, Canada
[3] Univ Toronto, Toronto, ON, Canada
[4] Univ Hlth Network, Lab Med Program, Toronto, ON, Canada
来源
PERFUSION-UK | 2021年 / 36卷 / 06期
关键词
activated clotting time; cardiac surgery; heparin; point-of-care testing device; protamine; DIFFERENT CLINICAL SETTINGS; COAGULATION TIME; I-STAT; HEPARIN; HEMOCHRON; ACT;
D O I
10.1177/0267659120949351
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: In cardiac surgery on cardiopulmonary bypass (CPB), heparin anticoagulation is monitored by point-of-care measurement of activated clotting time (ACT). The objective of this study was to compare four ACT systems in cardiac surgery in terms of their reproducibility, agreement and potential clinical impact at relevant medical decision points. Methods: The study included 40 cardiac surgery patients. Samples were taken at five time points before (T1), after heparinization for CPB (T2, T3, T4), and after heparin reversal (T5). The reproducibility, correlation, and differences in ACT values were assessed with two devices from each of the four ACT systems: Instrumentation Laboratory Hemochron Elite (Hmch), Medtronic HMS Plus (HMS), Abbott i-STAT, and Helena Abrazo. Subrange analyses were performed for low ACT values (results from T1, T5) and high ACT values (results from T2, T3, T4). Results: Within-system analysis showed strong linear correlation between paired measurements (R = 0.968-0.993). However, Hmch showed poorer reproducibility with highest proportion of values that exceed a difference of 10% and highest overall standard error of 74 seconds across the measurement range compared to that of the others (range 39-47 seconds, respectively). For inter-system comparison, using Hmch as reference, ACTs were strongly correlated as follows: HMS (R = 0.938), i-STAT (R = 0.911), and Abrazo (R = 0.911). Agreement analysis in the high ACT range showed HMS tended to have higher ACT values with +11% bias over Hmch, whereas i-STAT (-8% bias) and Abrazo (-13% bias) tended to underestimate. Post-protamine ACT results were dependent on device type where Hmch yielded highest post-protamine ACT (+13% higher than baseline) compared to -16% for HMS, -10% for iSTAT and 0% for Abrazo. Conclusions: Each device had individual reproducibility and biases, which may impact peri-operative heparin management. Careful validation must be undertaken when adopting a different method as decision limits would be affected. Clinicians should also be cautious using ACT as the only indicator for full heparin reversal.
引用
收藏
页码:610 / 619
页数:10
相关论文
共 19 条
[1]   In vitro effects of the platelet glycoprotein IIb/IIIa receptor antagonist c7E3 Fab on the activated clotting time [J].
Ammar, T ;
Scudder, LE ;
Coller, BS .
CIRCULATION, 1997, 95 (03) :614-617
[2]   COMPARISON OF HEMOCHRON AND HEMOTEC ACTIVATED COAGULATION TIME TARGET VALUES DURING PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY [J].
AVENDANO, A ;
FERGUSON, JJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (04) :907-910
[3]   EVALUATION OF TESTS USED TO MONITOR HEPARIN THERAPY DURING EXTRACORPOREAL-CIRCULATION [J].
BULL, MH ;
HUSE, WM ;
BULL, BS .
ANESTHESIOLOGY, 1975, 43 (03) :346-353
[4]   LACK OF CORRELATION BETWEEN ACTIVATED CLOTTING TIME AND PLASMA HEPARIN DURING CARDIOPULMONARY BYPASS [J].
CULLIFORD, AT ;
GITEL, SN ;
STARR, N ;
THOMAS, ST ;
BAUMANN, FG ;
WESSLER, S ;
SPENCER, FC .
ANNALS OF SURGERY, 1981, 193 (01) :105-111
[5]  
Falter Florian, 2018, J Extra Corpor Technol, V50, P38
[6]  
Hansley FA Jr, 2012, PRACTICAL APPROACH C
[7]   ACTIVATED COAGULATION TIME OF WHOLE BLOOD [J].
HATTERSLEY, PG .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1966, 196 (05) :436-+
[8]   A two site comparison of two point-of-care activated clotting time systems [J].
Kemna, Evelien W. M. ;
Kuipers, Charline ;
Luttikhuis-Spanjer, Agnes M. A. Oude ;
Majoor, Stefan ;
Boudrie, Renate ;
Speekenbrink, Ron G. H. ;
Hoffman, Reinhard ;
Krabbe, Johannes G. .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (01) :E13-E16
[9]   Comparison of activated clotting times measured using the Hemochron Jr. Signature and Medtronic ACT Plus during cardiopulmonary bypass with acute normovolemic haemodilution [J].
Lee, Jung Min ;
Park, Eun Young ;
Kim, Kyung Mi ;
Won, Jong Chan ;
Jung, Tack Koon ;
Lee, Soo Kyung .
JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2018, 46 (02) :873-882
[10]   Clinical Evaluation of the i-STAT Kaolin Activated Clotting Time (ACT) Test in Different Clinical Settings in a Large Academic Urban Medical Center Comparison With the Medtronic ACT Plus [J].
Lewandrowski, Elizabeth Lee ;
Van Cott, Elizabeth M. ;
Gregory, Kimberly ;
Jang, Ik-Kyung ;
Lewandrowski, Kent B. .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2011, 135 (05) :741-748