A fuzzy system cardio pulmonary bypass rotary blood pump controller

被引:15
作者
Casas, F
Orozco, A
Smith, WA
De Abreu-García, JA
Durkin, J
机构
[1] Cleveland Clin Fdn, Lerner Res Inst, Dept Biomed Engn, Cleveland, OH 44195 USA
[2] Univ Akron, Dept Elect & Comp Engn, Akron, OH 44325 USA
关键词
rotary blood pump; fuzzy control; non-linear model; flow estimation;
D O I
10.1016/j.eswa.2003.09.006
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Fuzzy logic regulation of flow was investigated in simulation of a model of the Cleveland Clinic Foundation rotodynamic Cardio Pulmonary Bypass blood pump comprised of impeller model #4079 and volute model #4080. A non-linear model of the pump was derived from pump maps obtained from measurements using a custom made pump dynamometer. Pump speed and generated delta pressure were used as inputs to a fuzzy engine in charge of manipulating the pump's speed; the fuzzy controller was capable of maintaining a setpoint of 6 +/- 0.5 l/min when presented with pressure disturbances over a range of +/- 50 mmHg from the baseline of 100 mmHg. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:357 / 361
页数:5
相关论文
共 3 条
[1]   Non-invasive flow estimation in an implantable rotary blood pump: a study considering non-pulsatile and pulsatile flows [J].
Ayre, PJ ;
Lovell, NH ;
Woodard, JC .
PHYSIOLOGICAL MEASUREMENT, 2003, 24 (01) :179-189
[2]   Computer simulation of sensorless fuzzy control of a rotary blood pump to assure normal physiology [J].
Fu, MH ;
Xu, LY .
ASAIO JOURNAL, 2000, 46 (03) :273-278
[3]   In vivo test of pressure head and flow rate estimation in a continuous-flow artificial heart [J].
Tanaka, A ;
Yoshizawa, M ;
Abe, K ;
Takeda, H ;
Yambe, T ;
Nitta, S .
ARTIFICIAL ORGANS, 2003, 27 (01) :99-103