Increasing iterative reconstruction strength at low tube voltage in coronary CT angiography protocols using 3D-printed and Catphan®500 phantoms

被引:6
作者
Abdullah, Kamarul A. [1 ]
McEntee, Mark F. [2 ]
Reed, Warren M. [2 ]
Kench, Peter L. [2 ]
机构
[1] Univ Sultan Zainal Abidin UniSZA, Fac Hlth Sci, Kuala Terengganu, Malaysia
[2] Univ Sydney, Med Imaging Optimisat & Percept Grp MIOPeG, Discipline Med Imaging Sci, Fac Med & Hlth,Sydney Sch Hlth Sci, Camperdown, NSW, Australia
关键词
3D printing; CT; image quality; iterative reconstruction; phantom; radiation dose; IMAGE QUALITY; RADIATION-EXPOSURE; CONTRAST AGENT; DOSE REDUCTION;
D O I
10.1002/acm2.12977
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The purpose of this study was to investigate the effect of increasing iterative reconstruction (IR) algorithm strength at different tube voltages in coronary computed tomography angiography (CCTA) protocols using a three-dimensional (3D)-printed and Catphan (R) 500 phantoms. Methods A 3D-printed cardiac insert and Catphan 500 phantoms were scanned using CCTA protocols at 120 and 100 kVp tube voltages. All CT acquisitions were reconstructed using filtered back projection (FBP) and Adaptive Statistical Iterative Reconstruction (ASIR) algorithm at 40% and 60% strengths. Image quality characteristics such as image noise, signal-noise ratio (SNR), contrast-noise ratio (CNR), high spatial resolution, and low contrast resolution were analyzed. Results There was no significant difference (P > 0.05) between 120 and 100 kVp measures for image noise for FBP vs ASIR 60% (16.6 +/- 3.8 vs 16.7 +/- 4.8), SNR of ASIR 40% vs ASIR 60% (27.3 +/- 5.4 vs 26.4 +/- 4.8), and CNR of FBP vs ASIR 40% (31.3 +/- 3.9 vs 30.1 +/- 4.3), respectively. Based on the Modulation Transfer Function (MTF) analysis, there was a minimal change of image quality for each tube voltage but increases when higher strengths of ASIR were used. The best measure of low contrast detectability was observed at ASIR 60% at 120 kVp. Conclusions Changing the IR strength has yielded different image quality noise characteristics. In this study, the use of 100 kVp and ASIR 60% yielded comparable image quality noise characteristics to the standard CCTA protocols using 120 kVp of ASIR 40%. A combination of 3D-printed and Catphan(R)500 phantoms could be used to perform CT dose optimization protocols.
引用
收藏
页码:209 / 214
页数:6
相关论文
共 22 条
[1]   Evaluation of an integrated 3D-printed phantom for coronary CT angiography using iterative reconstruction algorithm [J].
Abdullah, Kamarul A. ;
McEntee, Mark F. ;
Reed, Warren ;
Kench, Peter L. .
JOURNAL OF MEDICAL RADIATION SCIENCES, 2020, 67 (03) :170-176
[2]   3D printing in medical imaging and healthcare services [J].
Abdullah, Kamarul A. ;
Reed, Warren .
JOURNAL OF MEDICAL RADIATION SCIENCES, 2018, 65 (03) :237-239
[3]   Development of an organ-specific insert phantom generated using a 3D printer for investigations of cardiac computed tomography protocols [J].
Abdullah, Kamarul A. ;
McEntee, Mark F. ;
Reed, Warren ;
Kench, Peter L. .
JOURNAL OF MEDICAL RADIATION SCIENCES, 2018, 65 (03) :175-183
[4]   Radiation dose and diagnostic image quality associated with iterative reconstruction in coronary CT angiography: A systematic review [J].
Abdullah, Kamarul Amin ;
McEntee, Mark F. ;
Reed, Warren ;
Kench, Peter L. .
JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2016, 60 (04) :459-468
[5]   The Year in Coronary Artery Disease [J].
Achenbach, Stephan ;
Dilsizian, Vasken ;
Kramer, Christopher M. ;
Zoghbi, William A. .
JACC-CARDIOVASCULAR IMAGING, 2009, 2 (06) :774-786
[6]   Coronary Computed Tomography Angiography for Stable Angina: Past, Present, and Future [J].
Ajlan, Amr M. ;
Heilbron, Brett G. ;
Leipsic, Jonathon .
CANADIAN JOURNAL OF CARDIOLOGY, 2013, 29 (03) :266-274
[7]   Intravenous Contrast Medium Administration and Scan Timing at CT: Considerations and Approaches [J].
Bae, Kyongtae T. .
RADIOLOGY, 2010, 256 (01) :32-61
[8]   Image quality in coronary CT angiography: challenges and technical solutions [J].
Ghekiere, Olivier ;
Salgado, Rodrigo ;
Buls, Nico ;
Leiner, Tim ;
Mancini, Isabelle ;
Vanhoenacker, Piet ;
Dendale, Paul ;
Nchimi, Alain .
BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1072)
[9]   Optimizing radiation dose by using advanced modelled iterative reconstruction in high-pitch coronary CT angiography [J].
Gordic, Sonja ;
Desbiolles, Lotus ;
Sedlmair, Martin ;
Manka, Robert ;
Plass, Andre ;
Schmidt, Bernhard ;
Husarik, Daniela B. ;
Maisano, Francesco ;
Wildermuth, Simon ;
Alkadhi, Hatem ;
Leschka, Sebastian .
EUROPEAN RADIOLOGY, 2016, 26 (02) :459-468
[10]   What is the preferred strength setting of the sinogram-affirmed iterative reconstruction algorithm in abdominal CT imaging? [J].
Hardie A.D. ;
Nelson R.M. ;
Egbert R. ;
Rieter W.J. ;
Tipnis S.V. .
Radiological Physics and Technology, 2015, 8 (1) :60-63