Optical response of the FXG solution to different phantom materials

被引:6
作者
Cavinato, C. C. [1 ]
Sakuraba, R. K. [1 ,2 ]
Cruz, J. C. [2 ]
Campos, L. L. [1 ]
机构
[1] IPEN CNEN SP, BR-05508000 Sao Paulo, Brazil
[2] HIAE, Ctr Oncol, Unidade Radioterapia, BR-05652000 Sao Paulo, Brazil
关键词
Fricke xylenol gel dosimeter; Phantom materials; Clinical electron beams; Optical absorption spectrophotometry;
D O I
10.1016/j.radmeas.2011.05.050
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The purpose of this work is to evaluate the performance of the Fricke xylenol gel (FXG) solution developed at IPEN, prepared with 270 Bloom gelatine (made in Brazil), for clinical electron beams to the reference depth, using different phantom materials. The colour change, optical absorption spectra, intra and inter-batches reproducibility, dose-response, lower detection limit, energy and dose rate dependent response and response uniformity were studied. The excellent results obtained indicate the viability of employing this solution in 2D spectrophotometric dosimetry (could be extended to 3D MRI dosimetry) to be applied in quality assurance for clinical radiotherapy treatment planning of superficial tumours being treated with clinical electron beams. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1928 / 1931
页数:4
相关论文
共 5 条
[1]  
Andreo P, 2004, TECHNICAL REPORTS SE, V398
[2]   Radiochromic gel dosemeter for three-dimensional dosimetry [J].
Bero, MA ;
Gilboy, WB ;
Glover, PM .
RADIATION PHYSICS AND CHEMISTRY, 2001, 61 (3-6) :433-435
[3]   FERROUS SULFATE GELS FOR DETERMINATION OF ABSORBED DOSE DISTRIBUTIONS USING MRI TECHNIQUE - BASIC STUDIES [J].
OLSSON, LE ;
PETERSSON, S ;
AHLGREN, L ;
MATTSSON, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (01) :43-52
[4]  
Strydom W., 2005, Radiation Oncology Physics: A Handbook for Teachers and Students, P273
[5]  
Williams J.R., 1993, Radiotherapy Physics in Practice