On the numerical investigation of cardiovascular balloon-expandable stent using finite element method

被引:30
作者
Azaouzi, M. [1 ]
Makradi, A. [1 ]
Petit, J. [2 ]
Belouettar, S. [1 ]
Polit, O. [2 ]
机构
[1] Ctr Rech Publ Henri Tudor, Adv Mat & Struct Dept, L-4002 Esch Sur Alzette, Luxembourg
[2] Univ Paris Ouest, LEME, F-92410 Ville Davray, France
关键词
Cardiovascular stent; Finite element analysis; Numerical simulation; Fatigue life prediction; Goodman diagram; FATIGUE; SIZE; COMPONENTS; BEHAVIOR; FRACTURE; DESIGN; STRUTS;
D O I
10.1016/j.commatsci.2013.05.043
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The present paper shows an effective methodology which is based on Finite Element Analysis (FEA) for the fatigue life prediction (FLP) of a Balloon-Expandable (BE) stent made of stainless steel material (AISI316L). BE stent is a tubular, often mesh-like, structure which is expanded inside a diseased (stenosed) artery segment to restore blood flow and keep the vessel open following angioplasty. With design lives of 10-15 years, fatigue life prediction of these devices is critical for the designer, as fatigue failure may occur after deployment inside the blood vessel due to the heart beating (typically 4 x 10(7) cycles/year). Therefore, it is crucial to study the effect of stenting conditions and stent design on the fatigue life of BE stents. The objective of this paper is to provide quantitative measures of the stent's stress amplitude and mean stress which are generated by the cardiac pulse pressure. This allows prediction of the device's life and optimization of stent designs. In this paper, numerical investigations were conducted using an effective approach based on the well-known stress-based Goodman Diagram (GD) for the FLP of a given stent design. It has been demonstrated that the percentage of artery's expansion as well as the dimensions of the struts (the fundamental structural unit that forms the stent), have both an impact on the fatigue behavior after stent deployment. (C) 2013 Elsevier B. V. All rights reserved.
引用
收藏
页码:326 / 335
页数:10
相关论文
共 50 条
  • [21] In Vitro Evaluation of the Gore Viabahn Balloon-Expandable Stent-Graft for Fenestrated Endovascular Aortic Repair
    Torsello, Giovanni Federico
    Herten, Monika
    Mueller, Markus
    Andre-Frank
    Torsello, Giovanni B.
    Austermann, Martin
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2019, 26 (03) : 361 - 368
  • [22] Numerical investigation on dynamic characteristics of BGA structural integrity by using finite element method
    Chen, Zixia
    Yang, Ping
    [J]. International Journal of Materials and Structural Integrity, 2008, 2 (03) : 280 - 290
  • [23] Study of Coronary Stent Deformation using Finite Element Method
    Kalra, Shifali
    Mashuq-un-Nabi
    [J]. 2016 11TH INTERNATIONAL CONFERENCE ON INDUSTRIAL AND INFORMATION SYSTEMS (ICIIS), 2016, : 404 - 407
  • [24] Observation of the expansion behavior and quantitative evaluation of elastic recoil of a balloon-expandable stent in three dimensions using a Micro-CT system
    Mori, Futoshi
    Nakayama, Toshio
    Matsuzawa, Teruo
    Ohta, Makoto
    [J]. TECHNOLOGY AND HEALTH CARE, 2012, 20 (04) : 305 - 315
  • [25] Effect of Balloon-Expandable Transcatheter Aortic Valve Replacement Positioning: A Patient-Specific Numerical Model
    Bianchi, Matteo
    Marom, Gil
    Ghosh, Ram P.
    Fernandez, Harold A.
    Taylor, James R.
    Slepian, Marvin J.
    Bluestein, Danny
    [J]. ARTIFICIAL ORGANS, 2016, 40 (12) : E292 - E304
  • [26] Comparison of Balloon-Expandable Valve and Self-Expandable Valve in Transcatheter Aortic Valve Replacement: A Patient-Specific Numerical Study
    Li, Jianming
    Yan, Wentao
    Wang, Wenshuo
    Wang, Shengzhang
    Wei, Lai
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2022, 144 (10):
  • [27] Analysis of stress on inflation of balloon catheter using finite element method
    Shuib, Solehuddin
    Ridzwan, M. I. Z.
    Ibrahim, M. N. M.
    Ariffin, M. A. F. M.
    Abdullah, A. B.
    [J]. 3RD KUALA LUMPUR INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING 2006, 2007, 15 : 178 - +
  • [28] Numerical investigations of the mechanical properties of a braided non-vascular stent design using finite element method
    Ni, Xiao-Yu
    Pan, Chang-Wang
    Prusty, B. Gangadhara
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2015, 18 (10) : 1117 - 1125
  • [29] STENT VS STENT STUDY BY MEANS OF FINITE ELEMENT METHOD
    Imani, Misagh
    Goudarzi, A. M.
    Ganji, D. D.
    Gerdroodbary, M. Barzegar
    [J]. PROCEEDINGS OF THE ASME 12TH BIENNIAL CONFERENCE ON ENGINEERING SYSTEMS DESIGN AND ANALYSIS - 2014, VOL 1, 2014,
  • [30] Determination of the influence of stent strut thickness using the finite element method: implications for vascular injury and in-stent restenosis
    Zahedmanesh, Houman
    Lally, Caitriona