On the numerical investigation of cardiovascular balloon-expandable stent using finite element method

被引:29
作者
Azaouzi, M. [1 ]
Makradi, A. [1 ]
Petit, J. [2 ]
Belouettar, S. [1 ]
Polit, O. [2 ]
机构
[1] Ctr Rech Publ Henri Tudor, Adv Mat & Struct Dept, L-4002 Esch Sur Alzette, Luxembourg
[2] Univ Paris Ouest, LEME, F-92410 Ville Davray, France
关键词
Cardiovascular stent; Finite element analysis; Numerical simulation; Fatigue life prediction; Goodman diagram; FATIGUE; SIZE; COMPONENTS; BEHAVIOR; FRACTURE; DESIGN; STRUTS;
D O I
10.1016/j.commatsci.2013.05.043
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The present paper shows an effective methodology which is based on Finite Element Analysis (FEA) for the fatigue life prediction (FLP) of a Balloon-Expandable (BE) stent made of stainless steel material (AISI316L). BE stent is a tubular, often mesh-like, structure which is expanded inside a diseased (stenosed) artery segment to restore blood flow and keep the vessel open following angioplasty. With design lives of 10-15 years, fatigue life prediction of these devices is critical for the designer, as fatigue failure may occur after deployment inside the blood vessel due to the heart beating (typically 4 x 10(7) cycles/year). Therefore, it is crucial to study the effect of stenting conditions and stent design on the fatigue life of BE stents. The objective of this paper is to provide quantitative measures of the stent's stress amplitude and mean stress which are generated by the cardiac pulse pressure. This allows prediction of the device's life and optimization of stent designs. In this paper, numerical investigations were conducted using an effective approach based on the well-known stress-based Goodman Diagram (GD) for the FLP of a given stent design. It has been demonstrated that the percentage of artery's expansion as well as the dimensions of the struts (the fundamental structural unit that forms the stent), have both an impact on the fatigue behavior after stent deployment. (C) 2013 Elsevier B. V. All rights reserved.
引用
收藏
页码:326 / 335
页数:10
相关论文
共 25 条
[1]  
[Anonymous], 2011, ABAQUS MANUAL VERSIO
[2]   Optimization based simulation of self-expanding Nitinol stent [J].
Azaouzi, M. ;
Lebaal, N. ;
Makradi, A. ;
Belouettar, S. .
MATERIALS & DESIGN, 2013, 50 :917-928
[3]  
Barrera O., 2012, COMPUTER METHODS BIO, V15, P93
[4]  
Bathias C., 2008, FATIGUE MAT STRUCTUR
[5]   Piezoelectric fatigue testing machines and devices [J].
Bathias, Claude .
INTERNATIONAL JOURNAL OF FATIGUE, 2006, 28 (11) :1438-1445
[6]   A METHOD FOR OBTAINING AND ANALYZING SENSITIVITY DATA [J].
DIXON, WJ ;
MOOD, AM .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1948, 43 (241) :109-126
[7]   On the effects of different strategies in modelling balloon-expandable stenting by means of finite element method [J].
Gervaso, Francesca ;
Capelli, Claudio ;
Petrini, Lorenza ;
Lattanzio, Simone ;
Di Virgilio, Luca ;
Migliavacca, Francesco .
JOURNAL OF BIOMECHANICS, 2008, 41 (06) :1206-1212
[8]   Anisotropic mechanical properties of tissue components in human atherosclerotic plaques [J].
Holzapfel, GA ;
Sommer, G ;
Regitnig, P .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2004, 126 (05) :657-665
[9]   Computational modeling of effects of intravascular stent design on key mechanical and hemodynamic behavior [J].
Hsiao, Hao-Ming ;
Chiu, Yi-Hsiang ;
Lee, Kuang-Huei ;
Lin, Chien-Han .
COMPUTER-AIDED DESIGN, 2012, 44 (08) :757-765
[10]   Fatigue-life assessment and validation techniques for metallic vascular implants [J].
James, Brad A. ;
Sire, Robert A. .
BIOMATERIALS, 2010, 31 (02) :181-186