User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria and Calculation of Insulin Dosing Errors

被引:10
作者
Freckmann, Guido [1 ]
Jendrike, Nina [1 ]
Baumstark, Annette [1 ]
Pleus, Stefan [1 ]
Liebing, Christina [1 ]
Haug, Cornelia [1 ]
机构
[1] Univ Ulm, Inst Diabet Technol, Forsch & Entwicklungsgesell mbH, Ulm, Germany
关键词
Insulin dosing errors; ISO; 15197; Lay-user; Self-monitoring of blood glucose; Surveillance error grid; System accuracy; User performance evaluation; HANDS; INSTRUMENTS;
D O I
10.1007/s13300-018-0392-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction:The international standard ISO 15197:2013 requires a user performance evaluation to assess if intended users are able to obtain accurate blood glucose measurement results with a self-monitoring of blood glucose (SMBG) system. In this study, user performance was evaluated for four SMBG systems on the basis of ISO 15197:2013, and possibly related insulin dosing errors were calculated. Additionally, accuracy was assessed in the hands of study personnel. Methods:Accu-Chek (R) Performa Connect (A), Contour (R) plus ONE (B), FreeStyle Optium Neo (C), and OneTouch Select (R) Plus (D) were evaluated with one test strip lot. After familiarization with the systems, subjects collected a capillary blood sample and performed an SMBG measurement. Study personnel observed the subjects' measurement technique. Then, study personnel performed SMBG measurements and comparison measurements. Number and percentage of SMBG measurements within +/- 15 mg/dl and +/- 15% of the comparison measurements at glucose concentrations < 100 and >= 100 mg/dl, respectively, were calculated. In addition, insulin dosing errors were modelled. Results:In the hands of lay-users three systems fulfilled ISO 15197:2013 accuracy criteria with the investigated test strip lot showing 96% (A), 100% (B), and 98% (C) of results within the defined limits. All systems fulfilled minimum accuracy criteria in the hands of study personnel [99% (A), 100% (B), 99.5% (C), 96% (D)]. Measurements with all four systems were within zones of the consensus error grid and surveillance error grid associated with no or minimal risk. Regarding calculated insulin dosing errors, all 99% ranges were between dosing errors of -2.7 and +1.4 units for measurements in the hands of lay-users and between -2.5 and +1.4 units for study personnel. Frequent lay-user errors were not checking the test strips' expiry date and applying blood incorrectly. Conclusions:Data obtained in this study show that not all available SMBG systems complied with ISO 15197:2013 accuracy criteria when measurements were performed by lay-users.
引用
收藏
页码:683 / 697
页数:15
相关论文
共 39 条
[1]  
American Diabetes Association, 2017, Diabetes Care, V40, pS88
[2]  
[Anonymous], 2014, DTSCH RZTEBLATT
[3]   Accuracy and user performance evaluation of the Contour® Next Link 2.4 blood glucose monitoring system [J].
Bailey, Timothy ;
Wallace, Jane F. ;
Greene, Carmine ;
Pardo, Scott ;
Brown, Daniel ;
Pflug, Brian ;
Klaff, Leslie J. .
CLINICA CHIMICA ACTA, 2015, 448 :139-145
[4]  
Bailey Timothy S, 2017, J Diabetes Sci Technol, V11, P736, DOI 10.1177/1932296816680829
[5]  
Bailey Timothy S, 2016, J Diabetes Sci Technol, V10, P872, DOI 10.1177/1932296816634356
[6]   Performance of a New Blood Glucose Monitoring System in the Hands of Intended Users [J].
Bailey, Timothy S. ;
Wallace, Jane F. ;
Parkes, Joan Lee ;
Pardo, Scott ;
Yu, Ji ;
Schachner, Holly C. ;
Simmons, David A. ;
Chu, Amy .
DIABETES TECHNOLOGY & THERAPEUTICS, 2012, 14 (09) :783-789
[7]  
Bamberg Richard, 2005, Clin Lab Sci, V18, P203
[8]   Evaluation of Accuracy of Six Blood Glucose Monitoring Systems and Modeling of Possibly Related Insulin Dosing Errors [J].
Baumstark, Annette ;
Jendrike, Nina ;
Pleus, Stefan ;
Haug, Cornelia ;
Freckmann, Guido .
DIABETES TECHNOLOGY & THERAPEUTICS, 2017, 19 (10) :580-588
[9]  
Baumstark Annette, 2012, J Diabetes Sci Technol, V6, P1076
[10]  
Bedini Jose Luis, 2015, J Diabetes Sci Technol, V10, P85, DOI 10.1177/1932296815609368