Reducing ExacTrac intrafraction imaging uncertainty for prostate stereotactic body radiotherapy using a pre-treatment CBCT

被引:2
作者
Jordan, Barry [1 ]
Munoz, Luis [1 ]
Colyer, Christopher [1 ]
机构
[1] St Andrews Hosp Oncol, GenesisCare, Adelaide, SA, Australia
关键词
ExacTrac; CBCT; SBRT; Intrafraction; Prostate; IGRT; BEAM COMPUTED-TOMOGRAPHY; GUIDED RADIOTHERAPY; RADIATION-THERAPY; ORGAN MOTION; MULTIINSTITUTIONAL CONSORTIUM; GEOMETRIC ACCURACY; RADIOPAQUE MARKERS; IMPLANTED MARKERS; FIDUCIAL MARKERS; CANCER;
D O I
10.1007/s13246-022-01121-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study evaluated the intrafractional auto-matching uncertainties of prostate-implanted fiducial markers when using the ExacTrac v6.5 (Brainlab, Feldkirchen, Germany) X-ray stereoscopic system. A customised phantom with 3 implanted gold seeds was initially positioned at the isocentre using a cone beam CT (CBCT) prior to intrafractional imaging. Progressive offsets were applied to the phantom in all six directions (3 translational, 3 rotational) of 0 mm, 1 mm, 2 mm, 0 degrees, 1 degrees and 2 degrees. Subsequently, the ability of the ExacTrac image-matching functions to detect and correct these offsets was tested. For comparison, this procedure was repeated, but without a CBCT for pre-treatment positioning. The auto-matching uncertainties when a CBCT was introduced into the workflow were significantly reduced, and overall, the auto-matching statistics using the implanted marker (seeds) matching function was found to be more precise than the bony anatomy function in-phantom. The total standard deviations for the translational shifts using the implanted marker and bony anatomy functions respectively were 0.1 mm and 0.3 mm vertically, 0.1 mm and 0.3 mm longitudinally, and 0.1 mm and 0.4 mm laterally. The standard deviations for the rotational shifts using the implanted marker and bony anatomy matching functions respectively were 0.2 degrees and 1.2 degrees for the yaw (angle vert), 0.3 degrees and 1.1 degrees for the pitch (angle long), and 0.2 degrees and 1.2 degrees for the roll (angle lat) directions. The reduced uncertainties from introducing a CBCT for initial localisation resulted in decreased probability of inhibits due to false positives during treatment.
引用
收藏
页码:547 / 558
页数:12
相关论文
共 46 条
  • [1] Prostate planning treatment volume margin calculation based on the exactrac X-ray 6D image-guided system:: Margins for various clinical implementations
    Alonso-Arrizabalaga, Sara
    Gonzalez, Luis Brualla
    Ferrando, Juan V. Rosello
    Peidro, Jorge Pastor
    Torrecilla, Jose Lopez
    Meseguer, Domingo Planes
    Hernandez, Trinidad Garcia
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : 936 - 943
  • [2] Badakhshi H, 2013, ANTICANCER RES, V33, P4117
  • [3] Automated localization of the prostate at the time of treatment using implanted radiopaque markers: Technical feasibility
    Balter, JM
    Lam, KL
    Sandler, HM
    Littles, JF
    Bree, RL
    TenHaken, RK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05): : 1281 - 1286
  • [4] IMAGE-GUIDED RADIOTHERAPY (IGRT) FOR PROSTATE CANCER COMPARING kV IMAGING OF FIDUCIAL MARKERS WITH CONE BEAM COMPUTED TOMOGRAPHY (CBCT)
    Barney, Brandon M.
    Lee, R. Jeffrey
    Handrahan, Diana
    Welsh, Keith T.
    Cook, J. Taylor
    Sause, William T.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (01): : 301 - 305
  • [5] Stereotactic body radiation therapy: The report of AAPM Task Group 101
    Benedict, Stanley H.
    Yenice, Kamil M.
    Followill, David
    Galvin, James M.
    Hinson, William
    Kavanagh, Brian
    Keall, Paul
    Lovelock, Michael
    Meeks, Sanford
    Papiez, Lech
    Purdie, Thomas
    Sadagopan, Ramaswamy
    Schell, Michael C.
    Salter, Bill
    Schlesinger, David J.
    Shiu, Almon S.
    Solberg, Timothy
    Song, Danny Y.
    Stieber, Volker
    Timmerman, Robert
    Tome, Wolfgang A.
    Verellen, Dirk
    Wang, Lu
    Yin, Fang-Fang
    [J]. MEDICAL PHYSICS, 2010, 37 (08) : 4078 - 4101
  • [6] Fractionation and protraction for radiotherapy of prostate carcinoma
    Brenner, DJ
    Hall, EJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (05): : 1095 - 1101
  • [7] Is ExacTrac x-ray system an alternative to CBCT for positioning patients with head and neck cancers?
    Clemente, Stefania
    Chiumento, Costanza
    Fiorentino, Alba
    Simeon, Vittorio
    Cozzolino, Mariella
    Oliviero, Caterina
    Califano, Giorgia
    Caivano, Rocchina
    Fusco, Vincenzo
    [J]. MEDICAL PHYSICS, 2013, 40 (11)
  • [8] Clinical feasibility study for the use of implanted gold seeds in the prostate as reliable positioning markers during megavoltage irradiation
    Dehnad, H
    Nederveen, AJ
    van der Heide, UA
    van Moorselaar, RJA
    Hofman, P
    Lagendijk, JJW
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 67 (03) : 295 - 302
  • [9] Accurate patient dosimetry of kilovoltage cone-beam CT in radiation therapy
    Ding, George X.
    Duggan, Dennis M.
    Coffey, Charles W.
    [J]. MEDICAL PHYSICS, 2008, 35 (03) : 1135 - 1144
  • [10] IGRT of prostate cancer;: is the margin reduction gained from daily IG time-dependent?
    Enmark, Marika
    Korreman, Stine
    Nystrom, Hakan
    [J]. ACTA ONCOLOGICA, 2006, 45 (07) : 907 - 914