Feasibility and safety of high-power ablation of atrial fibrillation with contact force-sensing catheter: The lesion size index-guided ablation

被引:2
|
作者
Parwani, Abdul Shokor [1 ]
Hohendanner, Felix [1 ]
Kluck, Amelie [1 ]
Blaschke, Florian [1 ]
Pieske, Burkert [1 ]
Boldt, Leif-Hendrik [1 ]
机构
[1] Charite Univ Med Berlin, Campus Virchow Klinikum, Berlin, Germany
来源
ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE | 2022年 / 31卷 / 07期
关键词
atrial fibrillation; high-power ablation; lesion size index; contact force sensing; PULMONARY VEIN ISOLATION; SHORT-DURATION ABLATION;
D O I
10.17219/acem/146917
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background. Radiofrequency (RF) ablation is a commonly used method of atrial fibrillation (AF) treatment. High-power short-duration (HPSD) ablation has been suggested as a method to reduce procedure times whilst creating safe and lasting lesions. High-power ablation with contact force (CF)-sensing technology catheters might aid in a further improvement of safety whilst generating lasting transmural lesions. Objectives. We report our experience using lesion size index (LSI)-guided 50 W ablation with a CF-sensing catheter of AF. Materials and methods. We performed LSI-guided 50 W point-by-point ablation using a CF-sensing catheter (TactiCath). Target LSI at the anterior left atrium (LA) was 5.0 and at the posterior LA it was 4.5. Results. Altogether, 4641 RF lesions were created in 86 consecutive patients. To reach a mean LSI of 4.9 +/- 0.01, a mean RF ablation time of 14.3 +/- 0.1 s was applied with a mean CF of 13.4 +/- 0.1 g. The RF time per lesion at the anterior wall of LA was 15.9 +/- 0.2 s, while it was 13 +/- 0.2 s at the posterior wall of LA. We observed force time integral (FTI) values between 36 g and 310 g. Procedure duration was 107 +/- 4 min with a RF ablation time of 15.4 +/- 0.6 min. No audible steam pops occurred. No pericardial effusion was observed. After a 1-year follow-up, no adverse events were reported and 83% of patients had no symptomatic arrhythmia recurrence. Conclusions. We provide evidence for the safety and efficacy of LSI-guided 50 W ablation using the TactiCath CF-sensing ablation catheter. These data support the use of high-power ablation with CF sensing technology to improve both safety and efficacy.
引用
收藏
页码:715 / 721
页数:7
相关论文
共 50 条
  • [31] Treatment of Atrial Fibrillation Using Ablation Index-Guided Contact Force Ablation: A Matching-Adjusted Indirect Comparison to Cryoballoon Ablation
    Ahmed Hussein
    Dhiraj Gupta
    Tom De Potter
    Paul Spin
    Kiefer Eaton
    Laura Goldstein
    Maria Velleca
    Graça Costa
    Daniel Grima
    Leena Patel
    Giuseppe Stabile
    Advances in Therapy, 2020, 37 : 785 - 799
  • [32] Safety and Efficacy of High Power Shorter Duration Ablation Guided by Ablation Index or Lesion Size Index in Atrial Fibrillation Ablation: A Systematic Review and Meta-Analysis
    Liu, Xing
    Gui, Chun
    Wen, Weiming
    He, Yan
    Dai, Weiran
    Zhong, Guoqiang
    JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2021, 2021
  • [33] Real-world safety of catheter ablation for atrial fibrillation with contact force or cryoballoon ablation
    Natale, Andrea
    Mohanty, Sanghamitra
    Goldstein, Laura
    Gomez, Tara
    Hunter, Tina D.
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2021, 60 (03) : 445 - 452
  • [34] Use of a contact force-sensing ablation catheter with advanced catheter location significantly reduces fluoroscopy time and radiation dose in catheter ablation of atrial fibrillation
    Lee, Geoffrey
    Hunter, Ross J.
    Lovell, Matthew J.
    Finlay, Malcom
    Ullah, Waqas
    Baker, Victoria
    Dhinoja, Mehul B.
    Sporton, Simon
    Earley, Mark J.
    Schilling, Richard J.
    EUROPACE, 2016, 18 (02): : 211 - 218
  • [35] Assessment of High-Power Catheter Ablation in Patients With Atrial Fibrillation: A Meta-Analysis
    Mao, Zhi-Jie
    Pei, Yan
    Lin, Hui
    Xiang, Yin
    Huang, Zhou-Qing
    Xiao, Fang-Yi
    Chen, Yi-He
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2021, 8
  • [36] Reproducibility of Paroxysmal Atrial Fibrillation Ablation Clinical Outcomes Using Composite Ablation Index With Different Contact Force-Sensing Catheters
    Di Biase, Luigi
    Tabereaux, Paul
    Liu, Christopher F.
    Vasaiwala, Smit
    Chugh, Aman
    Saliba, Walid I.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2022, 15 (12)
  • [37] Safety of catheter ablation of atrial fibrillation using fiber optic-based contact force sensing
    Mansour, Moussa
    Lakkireddy, Dhanunjaya
    Packer, Douglas
    Day, John D.
    Mahapatra, Srijoy
    Brunner, Kyle
    Reddy, Vivek
    Natale, Andrea
    HEART RHYTHM, 2017, 14 (11) : 1631 - 1636
  • [38] Clinical outcomes in patients with persistent atrial fibrillation after technologic advances including contact force-guided and ablation index-guided ablation
    Won Kyeong Jeon
    So-Ryoung Lee
    Eue-Keun Choi
    Seil Oh
    International Journal of Arrhythmia, 23 (1)
  • [39] Catheter-related complications and mortality of atrial fibrillation ablation following introduction of contact force-sensing technology
    Akar, Joseph G.
    Hummel, James P.
    Yao, Xiaoxi
    Sangaralingham, Lindsey
    Dhruva, Sanket
    Dong, Jun
    Ward, Rebecca
    Shah, Nilay D.
    Ross, Joseph
    Noseworthy, Peter A.
    BMJ SURGERY INTERVENTIONS & HEALTH TECHNOLOGIES, 2020, 2 (01)
  • [40] Safety and efficacy of atrial fibrillation ablation guided by Ablation Index module
    Solimene, Francesco
    Schillaci, Vincenzo
    Shopova, Gergana
    Urraro, Francesco
    Arestia, Alberto
    Iuliano, Assunta
    Maresca, Fabio
    Agresta, Alessia
    La Rocca, Vincenzo
    De Simone, Antonio
    Stabile, Giuseppe
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2019, 54 (01) : 9 - 15