Noninvasive body contouring by focused ultrasound: Safety and efficacy of the contour I device in a multicenter, controlled, clinical study

被引:84
作者
Teitelbaum, Steven A.
Burns, John L.
Kubota, Junichiro
Matsuda, Hidenori
Otto, Morkel J.
Shirakabe, Yukio
Suzuki, Yoshiro
Brown, Spencer A.
机构
[1] Univ Texas, SW Med Ctr, Dept Plast Surg, Dallas, TX 75390 USA
[2] Univ Calif Los Angeles, Med Ctr, Los Angeles, CA 90024 USA
[3] Tokyo Univ Med, Tokyo, Japan
[4] Univ Texas, SW Med Ctr, Nancy L & Perry Bass Adv Wound Healing Lab, London, England
[5] Dallas Plast Surg Inst, Dallas, TX USA
关键词
D O I
10.1097/01.prs.0000270840.98133.c8
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: The removal of unwanted body fat using a noninvasive technique is desirable to patients and physicians. The authors describe a controlled, multicenter, clinical trial assessing the safety and efficacy of a focused therapeutic ultrasound device for noninvasive body contouring. Methods: Eligible healthy adult subjects were enrolled to the experimental group or the control group at five sites. The experimental group received one treatment with the Contour I device (UltraShape Ltd., Tel Aviv, Israel) in the abdomen, thighs, or flanks and were evaluated over a 12-week period. Efficacy outcomes were reduction of circumference and fat thickness. Circumference reduction was compared with the untreated group and with an untreated area (thigh) within the treated group. Safety monitoring included laboratory testing (including serum lipids), pulse oximetry, and liver ultrasound. Results: One hundred sixty-four subjects participated in the study (137 subjects in the experimental group and 27 in the control, untreated group). A single Contour I treatment was safe and well tolerated and produced a mean reduction of approximately 2 cm in treatment area circumference and approximately 2.9 mm in skin fat thickness. The majority of the effect was achieved within 2 weeks and was sustained at 12 weeks. No clinically significant changes in the measured safety parameters were recorded. Seven adverse events were reported, all of which were anticipated, mild, and resolved within the study period. Conclusion: The Contour I device provides a safe and effective noninvasive technology for body contouring.
引用
收藏
页码:779 / 789
页数:11
相关论文
共 21 条
  • [1] Analysis of the effects of deep mechanical massage in the porcine model
    Adcock, D
    Paulsen, S
    Jabour, K
    Davis, S
    Nanney, LB
    Shack, RB
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 2001, 108 (01) : 233 - 240
  • [2] *AM SOC AESTH PLAS, 2005, COSM SURG NATL DAT B
  • [3] Avram Mathew M, 2004, J Cosmet Laser Ther, V6, P181
  • [4] Perioperative management of cosmetic liposuction
    de Jong, RH
    Grazer, FM
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 2001, 107 (04) : 1039 - 1044
  • [5] Ultrasound-assisted liposuction:: An analysis of 348 cases
    Graf, R
    Auersvald, A
    Damasio, RCC
    Rippel, R
    de Araújo, LRR
    Bigarelli, LHC
    Franck, CL
    [J]. AESTHETIC PLASTIC SURGERY, 2003, 27 (02) : 146 - 153
  • [6] The safety of liposuction: Results of a national survey
    Housman, TS
    Lawrence, N
    Mellen, BG
    George, MN
    San Filippo, J
    Cerveny, KA
    DeMarco, M
    Feldman, SR
    Fleischer, AB
    [J]. DERMATOLOGIC SURGERY, 2002, 28 (11) : 971 - 978
  • [7] Practice advisory on liposuction
    Iverson, RE
    Lynch, DJ
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 2004, 113 (05) : 1478 - 1490
  • [8] The benefits of powered liposuction versus traditional liposuction: A paired comparison analysis
    Katz, BE
    Bruck, MC
    Coleman, WP
    [J]. DERMATOLOGIC SURGERY, 2001, 27 (10) : 863 - 867
  • [9] Thermodynamics, liposuction, and metbolism
    Kelley, DE
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2004, 350 (25) : 2542 - 2544
  • [10] Pharmacokinetics and safety of lidocaine and monoethylglycinexylidide in liposuction: A microdialysis study
    Kenkel, JM
    Lipschitz, AH
    Shepherd, G
    Armstrong, VW
    Streit, F
    Oellerich, M
    Luby, M
    Rohrich, RJ
    Brown, SA