Biomechanical comparison of femoral neck system and cannulated screws coupled with medial plate for treating Pauwels III femoral neck fractures

被引:4
|
作者
Nan, Chong [1 ]
Li, Yanjun [1 ]
Liu, Yuxiu [2 ]
Ma, Liang [1 ]
Ma, Zhanbei [1 ]
机构
[1] Baoding 1 Cent Hosp, Dept Orthoped, Baoding, Hebei, Peoples R China
[2] Baoding 1 Cent Hosp, Dept Neurosurg, Baoding, Hebei, Peoples R China
关键词
Femoral neck system; femoral neck fractures; finite element analysis; biomechanics; FIXATION; RISK;
D O I
10.3233/THC-220267
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
BACKGROUND: The femoral neck system (FNS) has been considered as a novel strategy for femoral neck fracture. The diversity of internal fixation creates difficulties in choosing an effective option for Pauwels III type femoral neck fractures. Therefore, it is significant to investigate the biomechanical effects of FNS versus conventional approaches on bones. OBJECTIVE: To evaluate the biomechanical characteristics of FNS versus cannulated screws coupled with medial plate (CSS+MP) for the treatment of Pauwels III type femoral neck fractures. METHODS: Through three-dimensional computer software (Minics, Geomagic - Warp), the proximal femur model was rebuilt. Based on the present clinical characteristics, models of internal fixation were reconstructed in SolidWorks, including cannulated screws (CSS), medial plate (MP) and FNS. After parameter setting and meshing, boundary conditions and loads were set up for the final mechanical calculation in Ansys Software. Under identical experimental conditions, such as the same Pauwels angle and force loading, the peak values of displacement, shear stress and equivalent (von Mises) stress were recorded. RESULTS: This study showed that the displacement of the models was CSS, CSS+MP, and FNS in descending order of magnitude. The shear stress and equivalent stress of the models was CSS +MP, FNS, and CSS in descending order. The principal shear stress of CSS+MP was concentrated on the medial plate. The equivalent stress of FNS was more dispersed and distributed from the proximal main nail to the distal locking screw. CONCLUSION: CSS+MP and FNS exhibited better initial stability compared to CSS. However, the MP was subjected to more shear stress, which could increase the risk of internal fixation failure. Due to its unique design, FNS may be a good choice for the treatment of Pauwels III type femoral neck fractures.
引用
收藏
页码:1161 / 1170
页数:10
相关论文
共 50 条
  • [1] Biomechanical Evaluation of the Femoral Neck System in Unstable Pauwels III Femoral Neck Fractures: A Comparison with the Dynamic Hip Screw and Cannulated Screws
    Stoffel, Karl
    Zderic, Ivan
    Gras, Florian
    Sommer, Christoph
    Eberli, Ursula
    Mueller, David
    Oswald, Martin
    Gueorguiev, Boyko
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (03) : 131 - 137
  • [2] Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture
    Zheng Wang
    Yong Yang
    Gangning Feng
    Haohui Guo
    Zhirong Chen
    Yaogeng Chen
    Qunhua Jin
    BioMedical Engineering OnLine, 21
  • [3] Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture
    Wang, Zheng
    Yang, Yong
    Feng, Gangning
    Guo, Haohui
    Chen, Zhirong
    Chen, Yaogeng
    Jin, Qunhua
    BIOMEDICAL ENGINEERING ONLINE, 2022, 21 (01)
  • [4] Biomechanical evaluation of percutaneous compression plate and femoral neck system in Pauwels type III femoral neck fractures
    Xiaoping Xie
    Songqi Bi
    Qingxu Song
    Qiong Zhang
    Zhixing Yan
    Xiaoyang Zhou
    Tiecheng Yu
    Journal of Orthopaedics and Traumatology, 25 (1)
  • [5] Femoral Neck System vs. four cannulated screws in the treatment of Pauwels III femoral neck fracture
    Lin, Hongkuan
    Lai, Caosheng
    Zhou, Zhiping
    Wang, Chaoqiang
    Yu, Xinlin
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2023, 28 (06) : 1373 - 1378
  • [6] Biomechanical and clinical evaluation of interlocking hip screw in Pauwels III femoral neck fractures: A comparison with inverted triangle cannulated screws
    Zhang, Jian
    Jiang, Haozheng
    Dai, Wei
    Hersi, Salad Abdirahman
    Cheong, John Chun Tien Chui Wan
    Chu, Zhenchen
    Lou, Zhiyuan
    Zhang, Deqiang
    Liu, Changjian
    Tian, Kang
    Tang, Xin
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2022, 10
  • [7] Biomechanical comparison of femoral neck anti-rotation and support system versus femoral neck system for unstable pauwels III femoral neck fractures
    Wang, Taiyou
    Wang, Guangjian
    Zhu, Fukang
    Qiao, Bo
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2024, 19 (01):
  • [8] Bio-mechanical effects of femoral neck system versus cannulated screws on treating young patients with Pauwels type III femoral neck fractures: a finite element analysis
    Fan, Xiao
    Zhou, Yimin
    Dai, Shiyou
    Lao, Kecheng
    Zhang, Qiliang
    Yu, Tengbo
    BMC MUSCULOSKELETAL DISORDERS, 2024, 25 (01)
  • [9] Bio-mechanical effects of femoral neck system versus cannulated screws on treating young patients with Pauwels type III femoral neck fractures: a finite element analysis
    Xiao Fan
    Yimin Zhou
    Shiyou Dai
    Kecheng Lao
    Qiliang Zhang
    Tengbo Yu
    BMC Musculoskeletal Disorders, 25
  • [10] Medial femoral plate with cannulated screw for Pauwels type III femoral neck fracture: A meta-analysis
    Su, Zhengbing
    Liang, Linlin
    Hao, Yong
    JOURNAL OF BACK AND MUSCULOSKELETAL REHABILITATION, 2021, 34 (02) : 169 - 177