Coronary Stent Off-Wire Dislodgement: Case Report of a Rare Complication

被引:2
作者
Lu, Meng-Ying [1 ]
Wang, Kuang-Te [1 ]
机构
[1] Taitung MacKay Mem Hosp, Dept Internal Med, Div Cardiol, 1 Ln 303,Changsha St, Taitung 95054, Taitung County, Taiwan
关键词
Aorta; complication; intraoperative; endovascular procedure; percutaneous coronary intervention; drug eluting stent; RETRIEVAL;
D O I
10.14503/THIJ-20-7505
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Although several techniques have been reported for managing an on-wire dislodged stent in the coronary artery, very few reports have focused on the much rarer complication of an off-wire dislodged stent. In a 73-year-old man who experienced an off-wire dislodged coronary stent, the proximal elongated segment was lodged in the left main coronary artery, and the distal segment was floating in the aorta like a wind sock. After a failed attempt at retrieval using a gooseneck microsnare, the dislodged stent was successfully removed using a 3-loop vascular snare via the left radial artery. There was no obvious vascular injury. This novel technique for removing a partially floating dislodged stent was successful after conventional retrieval techniques failed.
引用
收藏
页数:5
相关论文
共 8 条
  • [1] Alomar ME, 2013, J INVASIVE CARDIOL, V25, P637
  • [2] Incidence, retrieval methods, and outcomes of stent loss during percutaneous coronary intervention: A large single-center experience
    Brilakis, ES
    Best, PJM
    Elesber, AA
    Barsness, GW
    Lennon, RJ
    Holmes, DR
    Rihal, CS
    Garratt, KN
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 66 (03) : 333 - 340
  • [3] A Practical Approach to the Management of Complications During Percutaneous Coronary Intervention
    Giannini, Francesco
    Candilio, Luciano
    Mitomo, Satoru
    Ruparelia, Neil
    Chieffo, Alaide
    Baldetti, Luca
    Ponticelli, Francesco
    Latib, Azeem
    Colombo, Antonio
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2018, 11 (18) : 1797 - 1810
  • [4] Hernandez-Antolin Rosa Ana, 2008, J Invasive Cardiol, V20, pE67
  • [5] Jeong Min-Woong, 2016, Yeungnam University Journal of Medicine, V33, P138, DOI 10.12701/yujm.2016.33.2.138
  • [6] Dislodged Coronary Artery Stent Retrieved With an Endovascular Snare
    Senior, Jonathan
    Guillamo, Marina Roelas
    Ghattas, Angie
    Tapp, Luke
    [J]. TEXAS HEART INSTITUTE JOURNAL, 2020, 47 (03) : 213 - 215
  • [7] Udupa A, 2018, INTERV CARDIOL, V10, P13
  • [8] Dislodged and mechanically distorted stent stuck within a previously implanted drug-eluting stent
    Watabe, Hiroaki
    Sato, Akira
    Hoshi, Tomoya
    Obara, Kenichi
    Kawamura, Ryo
    Aonuma, Kazutaka
    [J]. HEART AND VESSELS, 2013, 28 (04) : 541 - 545