Evaluation of the mechanical stability of intraocular lenses using digital image correlation

被引:1
作者
Cabeza-Gil, Iulen [1 ]
Frechilla, Javier [1 ]
Calvo, Begona [1 ,2 ]
机构
[1] Univ Zaragoza, Aragon Inst Engn Res i3A, Zaragoza, Spain
[2] Biomat & Nanomed Networking Biomed Res Ctr CIBER B, Zaragoza, Spain
基金
欧盟地平线“2020”;
关键词
HAPTIC DESIGN; DECENTRATION;
D O I
10.1038/s41598-023-36694-0
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
This study aimed to evaluate the mechanical stability of seven different intraocular lens (IOL) haptic designs by using digital image correlation to measure their mechanical biomarkers (axial displacement, tilt, and rotation) under quasi-static compression. The IOLs were compressed between two clamps from 11.00 up to 9.50 mm whilst a 3D deformation dataset was acquired every 0.04 mm. Results revealed that flexible and mixed IOL designs exhibited better mechanical response for smaller compression diameters compared to stiff designs. Conversely, stiff designs performed better for larger compression diameters. These findings may aid in the selection and development of more mechanically stable IOL designs.
引用
收藏
页数:10
相关论文
共 31 条
[1]  
Ale JB, 2011, NEPAL J OPHTHALMOL, V3, P68, DOI 10.3126/nepjoph.v3i1.4281
[2]   Clinical and Optical Intraocular Performance of Rotationally Asymmetric Multifocal IOL Plate-Haptic Design Versus C-Loop Haptic Design [J].
Alio, Jorge L. ;
Plaza-Puche, Ana B. ;
Javaloy, Jaime ;
Jose Ayala, Maria ;
Vega-Estrada, Alfredo .
JOURNAL OF REFRACTIVE SURGERY, 2013, 29 (04) :252-259
[3]  
[Anonymous], 2012, 119793 BS EN ISO
[4]  
Bluestein EC, 1996, J PEDIATR OPHTHALMOL, V33, P18
[5]   Geometry of Acrylic, Hydrophobic IOLs and Changes in Haptic-Capsular Bag Relationship According to Compression and Different Well Diameters: A Bench Study Using Computed Tomography [J].
Borkenstein, Andreas F. ;
Borkenstein, Eva-Maria .
OPHTHALMOLOGY AND THERAPY, 2022, 11 (02) :711-727
[6]   Double-C loop platform in combination with hydrophobic and hydrophilic acrylic intraocular lens materials [J].
Bozukova, Dimitriya ;
Werner, Liliana ;
Mamalis, Nick ;
Gobin, Laure ;
Pagnoulle, Christophe ;
Floyd, Anne ;
Liu, Erica ;
Stallings, Shannon ;
Morris, Caleb .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (07) :1490-1502
[7]   Biomechanical and optical properties of 2 new hydrophobic platforms for intraocular lenses [J].
Bozukova, Dimitriya ;
Pagnoulle, Christophe ;
Jerome, Christine .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (09) :1404-1414
[8]   Mechanical characterisation of hydrophobic and hydrophilic acrylates used in intraocular lenses through depth sensing indentation [J].
Cabeza-Gil, I ;
Calvo, B. ;
Rico, A. ;
Reinhards-Hervas, C. ;
Rodriguez, J. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2022, 126
[9]   Effect of haptic geometry in C-loop intraocular lenses on optical quality [J].
Cabeza-Gil, I. ;
Perez-Gracia, J. ;
Remon, L. ;
Calvo, B. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2021, 114
[10]   Systematic Study on the Biomechanical Stability of C-Loop Intraocular Lenses: Approach to an Optimal Design of the Haptics [J].
Cabeza-Gil, I. ;
Ariza-Gracia, M. A. ;
Remon, L. ;
Calvo, B. .
ANNALS OF BIOMEDICAL ENGINEERING, 2020, 48 (04) :1127-1136