Dynamic Modeling of Carbon Dioxide Transport through the Skin Using a Capnometry Wristband

被引:3
作者
Grangeat, Pierre [1 ]
Duval Comsa, Maria-Paula [1 ]
Koenig, Anne [1 ]
Phlypo, Ronald [2 ]
机构
[1] Univ Grenoble Alpes, CEA, Leti, MINATEC Campus, F-38000 Grenoble, France
[2] Univ Grenoble Alpes, GIPSA Lab, CNRS, Grenoble INP, F-38000 Grenoble, France
基金
欧盟地平线“2020”;
关键词
capnometry; wearable health device; continuous monitoring; blood carbon dioxide concentration; carbon dioxide transcutaneous pressure; convection diffusion equation; compartmental model; state-space Markovian model; augmented dynamical system; Kalman filter; GAS; CO2; DIFFUSION; SENSOR; GILLS; LUNGS; PCO2;
D O I
10.3390/s23136096
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The development of a capnometry wristband is of great interest for monitoring patients at home. We consider a new architecture in which a non-dispersive infrared (NDIR) optical measurement is located close to the skin surface and is combined with an open chamber principle with a continuous circulation of air flow in the collection cell. We propose a model for the temporal dynamics of the carbon dioxide exchange between the blood and the gas channel inside the device. The transport of carbon dioxide is modeled by convection-diffusion equations. We consider four compartments: blood, skin, the measurement cell and the collection cell. We introduce the state-space equations and the associated transition matrix associated with a Markovian model. We define an augmented system by combining a first-order autoregressive model describing the supply of carbon dioxide concentration in the blood compartment and its inertial resistance to change. We propose to use a Kalman filter to estimate the carbon dioxide concentration in the blood vessels recursively over time and thus monitor arterial carbon dioxide blood pressure in real time. Four performance factors with respect to the dynamic quantification of the CO2 blood concentration are considered, and a simulation is carried out based on data from a previous clinical study. These demonstrate the feasibility of such a technological concept.
引用
收藏
页数:34
相关论文
共 90 条
[1]  
Agache P., 2017, AGACHES MEASURING SK, P1623
[2]  
Agache P., 2016, AGACHES MEASURING SK, P587
[3]  
Agache P., 2017, AGACHES MEASURING SK, P1607
[4]  
Akinwande D, 2019, NATURE, V576, P220, DOI 10.1038/d41586-019-03483-7
[5]   Identification of Coronary Artery Diseases Using Photoplethysmography Signals and Practical Feature Selection Process [J].
Al Fahoum, Amjed S. ;
Al-Haija, Ansam Omar Abu ;
Alshraideh, Hussam A. .
BIOENGINEERING-BASEL, 2023, 10 (02)
[6]   Modeling of chemical absorption of CO2 in membrane contactors [J].
Al-Marzouqi, M. ;
El-Naas, M. ;
Marzouk, S. ;
Abdullatiff, N. .
SEPARATION AND PURIFICATION TECHNOLOGY, 2008, 62 (03) :499-506
[7]   Compartmental modeling of skin transport [J].
Amarah, A. A. ;
Petlin, D. G. ;
Grice, J. E. ;
Hadgraft, J. ;
Roberts, M. S. ;
Anissimov, Y. G. .
EUROPEAN JOURNAL OF PHARMACEUTICS AND BIOPHARMACEUTICS, 2018, 130 :336-344
[8]  
Anderson B., 2004, Optimal Filtering
[9]   Mathematical and pharmacokinetic modelling of epidermal and dermal transport processes [J].
Anissimov, Yuri G. ;
Jepps, Owen G. ;
Dancik, Yuri ;
Roberts, Michael S. .
ADVANCED DRUG DELIVERY REVIEWS, 2013, 65 (02) :169-190
[10]  
[Anonymous], 1987, Kalman Filtering With Real-Time Applications