Towards a hygroscopic growth calibration for low-cost PM2.5 sensors

被引:3
|
作者
Patel, Milan Y. [1 ]
Vannucci, Pietro F. [1 ]
Kim, Jinsol [2 ]
Berelson, William M. [2 ]
Cohen, Ronald C. [1 ,3 ]
机构
[1] Univ Calif Berkeley, Dept Chem, Berkeley, CA 94720 USA
[2] Univ Southern Calif, Dept Earth Sci, Los Angeles, CA 90089 USA
[3] Univ Calif Berkeley, Dept Earth & Planetary Sci, Berkeley, CA 94720 USA
关键词
AEROSOL; AMBIENT; TRENDS; LINK; PART;
D O I
10.5194/amt-17-1051-2024
中图分类号
P4 [大气科学(气象学)];
学科分类号
0706 ; 070601 ;
摘要
Low-cost particulate matter (PM) sensors continue to grow in popularity, but issues such as aerosol-size-dependent sensitivity drive the need for effective calibration schemes. Here we devise a time-evolving calibration method for the Plantower PMS5003 PM2.5 mass concentration measurements. We use 2 years of measurements from the Berkeley Environmental Air-quality and CO2 Network sensors deployed in San Francisco and Los Angeles in our analysis. The calibration uses a hygroscopic growth correction factor derived from kappa-K & ouml;hler theory, where the calibration parameters are determined empirically using US Environmental Protection Agency Air Quality System (EPA AQS) reference data at co-location sites during the period from 2021-2022. The parameters are found to vary cyclically through the seasons, and the seasonal cycles match changes in sulfate and elemental carbon PM composition fractions throughout the year. In both regions, the seasonal RH dependence calibration performs better than the uncalibrated data and data calibrated with the EPA's national Plantower calibration algorithm. In the San Francisco Bay Area, the seasonal RH dependence calibration reduces the root mean square error (RMSE) by similar to 40 % from the uncalibrated data and maintains a mean bias much smaller than the EPA national calibration scheme (-0.90 vs -2.73 mu g m(-3)). We also find that calibration parameters forecasted beyond those fit with the EPA reference data continue to outperform the uncalibrated data and EPA calibration data, enabling real-time application of the calibration scheme even in the absence of reference data. While the correction greatly improves the data accuracy, non-Gaussian distribution of the residuals suggests that other processes besides hygroscopic growth can be parameterized for future improvement of this calibration.
引用
收藏
页码:1051 / 1060
页数:10
相关论文
共 50 条
  • [1] Concentration-Temporal Multilevel Calibration of Low-Cost PM2.5 Sensors
    Day, Rong-Fuh
    Yin, Peng-Yeng
    Huang, Yuh-Chin T.
    Wang, Cheng-Yi
    Tsai, Chih-Chun
    Yu, Cheng-Hsien
    SUSTAINABILITY, 2022, 14 (16)
  • [2] Evaluation of Low-Cost Sensors for Ambient PM2.5 Monitoring
    Badura, Marek
    Batog, Piotr
    Drzeniecka-Osiadacz, Anetta
    Modzel, Piotr
    JOURNAL OF SENSORS, 2018, 2018
  • [3] Ensemble learning of model hyperparameters and spatiotemporal data for calibration of low-cost PM2.5 sensors
    Yin, Peng-Yeng
    Tsai, Chih-Chun
    Day, Rong-Fuh
    Tung, Ching-Ying
    Bhanu, Bir
    MATHEMATICAL BIOSCIENCES AND ENGINEERING, 2019, 16 (06) : 6858 - 6873
  • [4] Statistical field calibration of a low-cost PM2.5 monitoring network in Baltimore
    Datta, Abhirup
    Saha, Arkajyoti
    Zamora, Misti Levy
    Buehler, Colby
    Hao, Lei
    Xiong, Fulizi
    Gentner, Drew R.
    Koehler, Kirsten
    ATMOSPHERIC ENVIRONMENT, 2020, 242
  • [5] Inter- versus Intracity Variations in the Performance and Calibration of Low-Cost PM2.5 Sensors: A Multicity Assessment in India
    Sreekanth, V
    Bhargav, R. Ajay
    Kulkarni, Padmavati
    Puttaswamy, Naveen
    Prabhu, Vignesh
    Agrawal, Pratyush
    Upadhya, Adithi R.
    Rao, Sofiya
    Sutaria, Ronak
    Mor, Suman
    Dey, Sagnik
    Khaiwal, Ravindra
    Balakrishnan, Kalpana
    Tripathi, Sachchida Nand
    Singh, Pratima
    ACS EARTH AND SPACE CHEMISTRY, 2022, 6 (12): : 3007 - 3016
  • [6] On the distribution of low-cost PM2.5 sensors in the US: demographic and air quality associations
    deSouza, Priyanka
    Kinney, Patrick L.
    JOURNAL OF EXPOSURE SCIENCE AND ENVIRONMENTAL EPIDEMIOLOGY, 2021, 31 (03) : 514 - 524
  • [7] Development of Low-Cost Air Quality Stations for Next Generation Monitoring Networks: Calibration and Validation of PM2.5 and PM10 Sensors
    Cavaliere, Alice
    Carotenuto, Federico
    Di Gennaro, Filippo
    Gioli, Beniamino
    Gualtieri, Giovanni
    Martelli, Francesca
    Matese, Alessandro
    Toscano, Piero
    Vagnoli, Carolina
    Zaldei, Alessandro
    SENSORS, 2018, 18 (09)
  • [8] Long-term evaluation and calibration of three types of low-cost PM2.5 sensors at different air quality monitoring stations
    Hong, Gung-Hwa
    Le, Thi-Cuc
    Tu, Jing-Wei
    Wang, Chieh
    Chang, Shuenn-Chin
    Yu, Jhih-Yuan
    Lin, Guan-Yu
    Aggarwal, Shankar G.
    Tsai, Chuen-Jinn
    JOURNAL OF AEROSOL SCIENCE, 2021, 157
  • [9] Performance calibration of low-cost and portable particular matter (PM) sensors
    Liu, Di
    Zhang, Qiang
    Jiang, Jingkun
    Chen, Da-Ren
    JOURNAL OF AEROSOL SCIENCE, 2017, 112 : 1 - 10
  • [10] Evaluations of Low-cost Air Quality Sensors for Particulate Matter (PM2.5) under Indoor and Outdoor Conditions
    Rabuan, Utbah
    Nadzir, Mohd Shahrul Mohd
    Sham, Siti Zahanah Abdullah
    Bahri, Sharifah Batrisyia Izzati Wan Shaiful
    Borah, Jintu
    Majumdar, Shubhankar
    Lei, Thomas M. T.
    Ali, Sawal Hamid Md
    Wahab, Muhammad Ikram A.
    Yunus, Noor Hidayah Mohd
    SENSORS AND MATERIALS, 2023, 35 (08) : 2881 - 2895