Dislocation force of scleral flange-fixated intraocular lens haptics

被引:2
作者
Pukl, Spela Stunf [1 ,2 ,3 ]
Kronschlaeger, Martin [1 ]
Ruiss, Manuel [1 ]
Blouin, Stephane [4 ,5 ]
Akcan, Emre Ruestue [1 ]
Findl, Oliver [1 ]
机构
[1] Hanusch Hosp, VIROS Vienna Inst Res Ocular Surg, Heinrich Collin Str 30, AT-1140 Vienna, Austria
[2] Univ Ljubljana, Med Fac, Ljubljana, Slovenia
[3] Univ Clin Ctr Ljubljana, Eye Hosp, Ljubljana, Slovenia
[4] Hanusch Hosp OEGK, Ludwig Boltzmann Inst Osteol, Vienna, Austria
[5] Med Dept Hanusch Hosp, AUVA Trauma Ctr Meidling, Vienna, Austria
关键词
Scleral fixation; Flange-fixated IOL; Yamane technique; Dislocation force; COLLAGEN FIBRILS; HUMAN CORNEA; ORGANIZATION;
D O I
10.1186/s12886-024-03369-x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PurposeTo measure the dislocation forces in relation to haptic material, flange size and needle used.SettingHanusch Hospital, Vienna, Austria.DesignLaboratory Investigation.Methods, main outcome measures30 G (gauge) thin wall and 27 G standard needles were used for a 2 mm tangential scleral tunnel in combination with different PVDF (polyvinylidene fluoride) and PMMA (polymethylmethacrylate haptics). Flanges were created by heating 1 mm of the haptic end, non-forceps assisted in PVDF and forceps assisted in PMMA haptics. The dislocation force was measured in non-preserved cadaver sclera using a tensiometer device.ResultsPVDF flanges achieved were of a mushroom-like shape and PMMA flanges were of a conic shape. For 30 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 1.58 +/- 0.68 N (n = 10) and 0.70 +/- 0.14 N (n = 9) (p = 0.003) respectively. For 27 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 0.31 +/- 0.35 N (n = 3) and 0.0 N (n = 4), respectively. The flange size correlated with the occurring dislocation force in experiments with 30 G needle tunnels (r = 0.92), when flanges were bigger than 384 micrometres.ConclusionsThe highest dislocation forces were found for PVDF haptic flanges and their characteristic mushroom-like shape for 30 G thin wall needle scleral tunnels. Forceps assisted flange creation in PMMA haptics did not compensate the disadvantage of PMMA haptics with their characteristic conic shape flange.
引用
收藏
页数:9
相关论文
共 14 条
[1]   Creating the flange in Yamane's technique [J].
Agrawal, Vishal ;
Raju, Biju .
INDIAN JOURNAL OF OPHTHALMOLOGY, 2021, 69 (02) :465-465
[2]  
Canabrava S., 2022, Arq Bras Oftalmol, V7
[3]   Regional variation in the biomechanical properties of the human sclera [J].
Elsheikh, Ahmed ;
Geraghty, Brendan ;
Alhasso, Daad ;
Knappett, Jonathan ;
Campanelli, Marino ;
Rama, Paolo .
EXPERIMENTAL EYE RESEARCH, 2010, 90 (05) :624-633
[4]  
Jacob S, 2017, ASIA-PAC J OPHTHALMO, V6, P381, DOI 10.22608/APO.2017158
[5]   Collagen and mature elastic fibre organisation as a function of depth in the human cornea and limbus [J].
Kamma-Lorger, Christina S. ;
Boote, Craig ;
Hayes, Sally ;
Moger, Julian ;
Burghammer, Manfred ;
Knupp, Carlo ;
Quantock, Andrew J. ;
Sorensen, Thomas ;
Di Cola, Emanuela ;
White, Nick ;
Young, Robert D. ;
Meek, Keith M. .
JOURNAL OF STRUCTURAL BIOLOGY, 2010, 169 (03) :424-430
[6]  
KOMAI Y, 1991, INVEST OPHTH VIS SCI, V32, P2244
[7]   Attaining the optimal flange for intrascleral intraocular lens fixation [J].
Kronschlaeger, Martin ;
Blouin, Stephane ;
Roschger, Paul ;
Varsits, Ralph ;
Findl, Oliver .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (11) :1303-1305
[8]   A Biomechanical Study of Flanged Intrascleral Haptic Fixation of Three-Piece Intraocular Lenses [J].
Ma, Kevin K. ;
Yuan, Amy ;
Sharifi, Sina ;
Pineda, Roberto .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2021, 227 :45-52
[9]  
Newton RH, 1998, INVEST OPHTH VIS SCI, V39, P1125
[10]   Size variability and fitting of 30-gauge thin-wall needles and 3-piece intraocular lens haptics for the flanged intrascleral fixation technique [J].
Pukl, Spela Stunf ;
Kronschlager, Martin ;
Ruiss, Manuel ;
Blouin, Stephane ;
Findl, Oliver .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2023, 49 (08) :874-878