Performance of an implantable cardioverter-defibrillator lead family

被引:1
|
作者
Klampfleitner, Stefanie [1 ,2 ]
Mundel, Markus [1 ]
Schinke, Karin [1 ]
Neuberger, Hans-Ruprecht [1 ,2 ]
机构
[1] Klinikum Traunstein, Sekt Kardiol Rhythmol, Cuno Niggl Str 3, D-83278 Traunstein, Germany
[2] Univ Saarland, Klin Innere Med 3, Univ Klinikum Homburg, Saarbrucken, Saarland, Germany
关键词
implantable cardioverter-defibrillator; lead; lead failure; lead performance; Linox; Linox Smart; BIOTRONIK LINOX; FAILURE;
D O I
10.1111/jce.15804
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundLead failure is the major limitation in implantable cardioverter-defibrillator (ICD) therapy. Long-term follow-up data for Biotronik Linox ICD leads are limited. Therefore, we analyzed the performance of all these leads implanted at our institution. Materials and MethodsAll Linox and Linox Smart ICD leads implanted between 2006 and 2015 were identified. Lead failure was defined as electrical dysfunction (oversensing, abnormal impedance, exit block). Lead survival was described, according to Kaplan-Meier. Associations between lead failure and specific variables were examined. p < .05 was considered significant. ResultsWe included 417 ICD leads. The median follow-up time for Linox (n = 205) was 81 months and for Linox Smart (n = 212) 75 months. During that follow-up time, 30 Linox (14.6%) and 16 Linox Smart leads (7.6%) showed a malfunction. The 5-year lead survival probability was 97.4% for Linox and 95.2% for Linox Smart (log-rank test, p = .19). The 6- and 8-year lead survival probability for Linox was 93.6% and 84.6%, and for Linox Smart 93% and 91.9%. The only factor significantly associated with lead failure was younger patient age at implantation (hazard ratio/year: 0.97, 95% CI: 0.94-0.99, p = .002). ConclusionThis relatively large study with a long follow-up period highlights a relevant failure rate of Biotronik Linox leads. The performance of Linox versus Linox Smart ICD leads was comparable. Although we show an acceptable 5-year lead survival probability, we observed a marked drop after just 1 more year of follow-up. In an era of improving heart failure survival probability, a prolonged follow-up of ICD leads is increasingly clinically relevant.
引用
收藏
页码:700 / 709
页数:10
相关论文
共 50 条
  • [1] Implantable cardioverter-defibrillator lead performance
    Maisel, William H.
    Kramer, Daniel B.
    CIRCULATION, 2008, 117 (21) : 2721 - 2723
  • [2] Clinical performance of implantable cardioverter-defibrillator lead monitoring diagnostics
    Poole, Jeanne E.
    Swerdlow, Charles D.
    Tarakji, Khaldoun G.
    Mittal, Suneet
    Ellenbogen, Kenneth A.
    Greenspon, Arnold J.
    Kennergren, Charles
    Philbert, Berit T.
    Moore, JoEllyn
    Jones, R. Chris
    Schaller, Robert D.
    Hansalia, Riple
    Simmers, Timothy
    Mihalcz, Attila
    DeBus, Becky
    Lexcen, Daniel R.
    Gunderson, Bruce
    Wilkoff, Bruce L.
    HEART RHYTHM, 2022, 19 (03) : 363 - 371
  • [3] Pacemaker/implantable cardioverter-defibrillator problem: Right ventricular lead missensing in a biventricular implantable cardioverter-defibrillator
    Otten, Richard F.
    Foreman, Lynne D.
    Groh, William J.
    HEART RHYTHM, 2008, 5 (01) : 158 - 159
  • [4] The implantable cardioverter-defibrillator
    Raj, SR
    Sheldon, RS
    PROGRESS IN CARDIOVASCULAR DISEASES, 2001, 44 (03) : 169 - 194
  • [5] The implantable cardioverter-defibrillator
    Greene, HL
    CLINICAL CARDIOLOGY, 2000, 23 (05) : 315 - 326
  • [6] The implantable cardioverter-defibrillator
    Groh, WJ
    Foreman, LD
    Zipes, DP
    AMERICAN FAMILY PHYSICIAN, 1998, 57 (02) : 310 - 312
  • [7] INFECTIVE ENDOCARDITIS OF THE TRANSVENOUS LEAD OF AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    WERNER, GS
    GONSKA, BD
    HERSE, B
    KREUZER, H
    ZEITSCHRIFT FUR KARDIOLOGIE, 1995, 84 (01): : 51 - 54
  • [8] Management of the Patient With Implantable Cardioverter-Defibrillator Lead Failure
    Kalahasty, Gautham
    Ellenbogen, Kenneth A.
    CIRCULATION, 2011, 123 (12) : 1352 - 1354
  • [9] Lead failure in an entirely subcutaneous implantable cardioverter-defibrillator
    Kolb, Christof
    Weigand, Severin
    Schaarschmidt, Claudia
    Lennerz, Carsten
    EUROPACE, 2020, 22 (02): : 183 - 183
  • [10] The Implantable Cardioverter-Defibrillator Lead: Principles, Progress, and Promises
    Haqqani, Haris M.
    Mond, Harry G.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2009, 32 (10): : 1336 - 1353