Multiple Brain Metastases Radiosurgery with CyberKnife Device: Dosimetric Comparison between Fixed/Iris and Multileaf Collimator Plans

被引:2
|
作者
Ianiro, Anna [1 ]
Infusino, Erminia [1 ]
D'Andrea, Marco [1 ]
Marucci, Laura [2 ]
Farneti, Alessia [2 ]
Sperati, Francesca [3 ]
Cassano, Bartolomeo [1 ]
Ungania, Sara [1 ]
Soriani, Antonella [1 ]
机构
[1] IRCCS Regina Elena Natl Canc Inst, Dept Med Phys, Rome, Italy
[2] IRCCS Regina Elena Natl Canc Inst IFO, Dept Radiat Oncol, Rome, Italy
[3] IRCCS San Gallicano Dermatol Inst IFO, Dept Biostat & Bioinformat, Clin Trial Ctr, Rome, Italy
关键词
Brain radiosurgery; CyberKnife (R); dosimetric comparison; BODY RADIATION-THERAPY; STEREOTACTIC RADIOSURGERY; QUALITY-ASSURANCE; RADIOTHERAPY; SURVIVAL; SYSTEM; IRRADIATION; NUMBER; CANCER;
D O I
10.4103/jmp.jmp_82_22
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In our institution, stereotactic radiosurgery of multiple brain metastases is performed with the CyberKnife (R) (CK) device, using fixed/Iris collimators. In this study, nineteen fixed/Iris plans were recalculated with the multileaf collimator (MLC), to assess if it is possible to produce plans with comparable dosimetric overall quality. Materials and Methods: For consistent comparisons, MLC plans were re-optimized and re-normalized in order to achieve the same minimum dose for the total planning target volume (PTVtot). Conformation number (CN), homogeneity index (HI) and dose gradient index (DGI) metrics were evaluated. The dose to the brain was evaluated as the volume receiving 12 Gy (V-12) and as the integral dose (ID). The normal tissue complication probability (NTCP) for brain radionecrosis was calculated as a function of V-12. Results: The reoptimized plans were reviewed by the radiation oncologist and were found clinically acceptable according to the The American Association of Physicists in Medicine (AAPM) Task Group-101 protocol. However, fixed/Iris plans provided significantly higher CN (+8.6%), HI (+2.2%), and DGI (+44.0%) values, and significantly lower ID values (-35.9%). For PTVtot less than the median value of 2.58cc, fixed/Iris plans provided significantly lower NTCP values. On the other side, MLC plans provided significantly lower treatment times (-18.4%), number of monitor units (-33.3%), beams (-46.0%) and nodes (-21.3%). Conclusions: CK-MLC plans for the stereotactic treatment of brain multi metastases could provide an important advantage in terms of treatment duration. However, to contain the increased risk for brain radionecrosis, it could be useful to calculate MLC plans only for patients with large PTVtot.
引用
收藏
页码:120 / 128
页数:9
相关论文
共 50 条
  • [21] Dosimetric comparison of Linac-based (BrainLAB®) and robotic radiosurgery (CyberKnife®) stereotactic system plans for acoustic schwannoma
    Debnarayan Dutta
    S. Balaji Subramanian
    V. Murli
    H. Sudahar
    P. G. Gopalakrishna Kurup
    Mahadev Potharaju
    Journal of Neuro-Oncology, 2012, 106 : 637 - 642
  • [22] Dosimetric comparison of Linac-based (BrainLAB®) and robotic radiosurgery (CyberKnife®) stereotactic system plans for acoustic schwannoma
    Dutta, Debnarayan
    Subramanian, S. Balaji
    Murli, V.
    Sudahar, H.
    Kurup, P. G. Gopalakrishna
    Potharaju, Mahadev
    JOURNAL OF NEURO-ONCOLOGY, 2012, 106 (03) : 637 - 642
  • [23] Dosimetric Comparison of Intensity-Modulated Radiosurgery and Helical Tomotherapy for the Treatment of Multiple Intracranial Metastases
    VanderSpek, Lauren
    Bauman, Glenn
    Wang, Jia-Zhu
    Yartsev, Slav
    Menard, Cynthia
    Cho, Young-Bin
    Mundt, Arno J.
    Lawson, Joshua D.
    Murphy, Kevin T.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2009, 8 (05) : 361 - 367
  • [24] Effect of collimator angle on HyperArc stereotactic radiosurgery planning for single and multiple brain metastases
    Ohira, Shingo
    Sagawa, Tomohiro
    Ueda, Yoshihiro
    Inui, Shoki
    Masaoka, Akira
    Akino, Yuichi
    Mizuno, Hirokazu
    Miyazaki, Masayoshi
    Koizumi, Masahiko
    Teshima, Teruki
    MEDICAL DOSIMETRY, 2020, 45 (01) : 85 - 91
  • [25] A Comparison of Helical Tomotherapy to Circular Collimator-Based Linear-Accelerator Radiosurgery for the Treatment of Brain Metastases
    Soisson, Emilie T.
    Mehta, Minesh P.
    Tome, Wolfgang A.
    AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2011, 34 (04): : 388 - 394
  • [26] Preliminary dosimetric comparison between fixed and rotating source stereotactic radiosurgery systems
    Draeger, Emily
    Chen, Zhe
    Hansen, James E.
    Chiang, Veronica
    Tien, Christopher J.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (05):
  • [27] Single isocenter stereotactic radiosurgery for patients with multiple brain metastases: dosimetric comparison of VMAT and a dedicated DCAT planning tool
    Hofmaier, Jan
    Bodensohn, Raphael
    Garny, Sylvia
    Hadi, Indrawati
    Fleischmann, Daniel F.
    Eder, Michael
    Dinc, Yavuz
    Reiner, Michael
    Corradini, Stefanie
    Parodi, Katia
    Belka, Claus
    Niyazi, Maximilian
    RADIATION ONCOLOGY, 2019, 14 (1)
  • [28] Single isocenter stereotactic radiosurgery for patients with multiple brain metastases: dosimetric comparison of VMAT and a dedicated DCAT planning tool
    Jan Hofmaier
    Raphael Bodensohn
    Sylvia Garny
    Indrawati Hadi
    Daniel F. Fleischmann
    Michael Eder
    Yavuz Dinc
    Michael Reiner
    Stefanie Corradini
    Katia Parodi
    Claus Belka
    Maximilian Niyazi
    Radiation Oncology, 14
  • [29] Dosimetric effect of rotational setup errors in stereotactic radiosurgery with HyperArc for single and multiple brain metastases
    Sagawa, Tomohiro
    Ohira, Shingo
    Ueda, Yoshihiro
    Akino, Yuichi
    Mizuno, Hirokazu
    Matsumoto, Masao
    Miyazaki, Masayoshi
    Teshima, Teruki
    Koizumi, Masahiko
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (10): : 84 - 91
  • [30] Dosimetric Study of Automatic Brain Metastases Planning in Comparison with Conventional Multi-Isocenter Dynamic Conformal Arc Therapy and Gamma Knife Radiosurgery for Multiple Brain Metastases
    Mori, Yoshimasa
    Kaneda, Naoki
    Hagiwara, Masahiro
    Ishiguchi, Tuneo
    CUREUS, 2016, 8 (11):