Evaluation of Different Contrast Agents for Regional Lung Perfusion Measurement Using Electrical Impedance Tomography: An Experimental Pilot Study

被引:9
作者
Muders, Thomas [1 ]
Hentze, Benjamin [1 ,2 ]
Leonhardt, Steffen [2 ]
Putensen, Christian [1 ]
机构
[1] Univ Hosp Bonn, Dept Anesthesiol & Intens Care Med, D-53127 Bonn, Germany
[2] Rhein Westfal TH Aachen, Chair Med Informat Technol, D-52074 Aachen, Germany
关键词
individualized therapy; electrical impedance tomography; bedside monitoring; bedside visualization; pulmonary perfusion; contrast agent; functional imaging; IMPACT; INJURY; MEDIA; MODEL;
D O I
10.3390/jcm12082751
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Monitoring regional blood flow distribution in the lungs appears to be useful for individually optimizing ventilation therapy. Electrical impedance tomography (EIT) can be used at the bedside for indicator-based regional lung perfusion measurement. Hypertonic saline is widely used as a contrast agent but could be problematic for clinical use due to potential side effects. In five ventilated healthy pigs, we investigated the suitability of five different injectable and clinically approved solutions as contrast agents for EIT-based lung perfusion measurement. Signal extraction success rate, signal strength, and image quality were analyzed after repeated 10 mL bolus injections during temporary apnea. The best results were obtained using NaCl 5.85% and sodium-bicarbonate 8.4% with optimal success rates (100%, each), the highest signal strengths (100 +/- 25% and 64 +/- 17%), and image qualities (r = 0.98 +/- 0.02 and 0.95 +/- 0.07). Iomeprol 400 mg/mL (non-ionic iodinated X-ray contrast medium) and Glucose 5% (non-ionic glucose solution) resulted in mostly well usable signals with above average success rates (87% and 89%), acceptable signal strength (32 +/- 8% and 16 + 3%), and sufficient image qualities (r = 0.80 +/- 0.19 and 0.72 +/- 0.21). Isotonic balanced crystalloid solution failed due to a poor success rate (42%), low signal strength (10 +/- 4%), and image quality (r = 0.43 +/- 0.28). While Iomeprol might enable simultaneous EIT and X-ray measurements, glucose might help to avoid sodium and chloride overload. Further research should address optimal doses to balance reliability and potential side effects.
引用
收藏
页数:12
相关论文
共 50 条
[31]   Conductivity Distribution Measurement at Different Low Frequencies Using a Modular 64 Electrode Electrical Impedance Tomography System [J].
Dimas, Christos ;
Sotiriadis, Paul P. .
2017 4TH PANHELLENIC CONFERENCE ON ELECTRONICS AND TELECOMMUNICATIONS (PACET), 2017, :133-136
[32]   Evaluation of Electrical Impedance Tomography for a Measurement of Tidal Volume in High-frequency Ventilation of the Pig - Case Study [J].
Rozanek, Martin ;
Kudrna, Petr ;
Rafl, Jakub ;
Ort, Vaclav ;
Kordik, Marek ;
Roubik, Karel .
2019 E-HEALTH AND BIOENGINEERING CONFERENCE (EHB), 2019,
[33]   Effects of different VV ECMO blood flow rates on lung perfusion assessment by hypertonic saline bolus-based electrical impedance tomography [J].
Zhang, Hongling ;
Wu, Yongran ;
Gao, Xuehui ;
Peng, Chengchao ;
Li, Ruirui ;
Wang, Azhen ;
Zhang, Jiancheng ;
Yuan, Shiying ;
Yang, Le ;
Zou, Xiaojing ;
Shang, You .
CRITICAL CARE, 2024, 28 (01)
[34]   Experimental Study on Early Detection of Acute Cerebral Ischemic Stroke Using Electrical Impedance Tomography Method [J].
Shi, Xuetao ;
You, Fusheng ;
Xu, Canhua ;
Wang, Liang ;
Fu, Feng ;
Liu, Ruigang ;
Dong, Xiuzhen .
WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 :510-513
[35]   Generation of Anatomically Inspired Human Airway Tree Using Electrical Impedance Tomography: A Method to Estimate Regional Lung Filling Characteristics [J].
Zamani, Majid ;
Kallio, Merja ;
Bayford, Richard ;
Demosthenous, Andreas .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2022, 41 (05) :1125-1137
[36]   Three broad classifications of acute respiratory failure etiologies based on regional ventilation and perfusion by electrical impedance tomography: a hypothesis-generating study [J].
He, Huaiwu ;
Chi, Yi ;
Long, Yun ;
Yuan, Siyi ;
Zhang, Rui ;
Yang, Yingying ;
Frerichs, Inez ;
Moller, Knut ;
Fu, Feng ;
Zhao, Zhanqi .
ANNALS OF INTENSIVE CARE, 2021, 11 (01)
[37]   Three broad classifications of acute respiratory failure etiologies based on regional ventilation and perfusion by electrical impedance tomography: a hypothesis-generating study [J].
Huaiwu He ;
Yi Chi ;
Yun Long ;
Siyi Yuan ;
Rui Zhang ;
Yingying Yang ;
Inéz Frerichs ;
Knut Möller ;
Feng Fu ;
Zhanqi Zhao .
Annals of Intensive Care, 11
[38]   Monitoring postoperative lung recovery using electrical impedance tomography in post anesthesia care unit: an observational study [J].
Nadine Hochhausen ;
Torsten Kapell ;
Martin Dürbaum ;
Andreas Follmann ;
Rolf Rossaint ;
Michael Czaplik .
Journal of Clinical Monitoring and Computing, 2022, 36 :1205-1212
[39]   Monitoring postoperative lung recovery using electrical impedance tomography in post anesthesia care unit: an observational study [J].
Hochhausen, Nadine ;
Kapell, Torsten ;
Duerbaum, Martin ;
Follmann, Andreas ;
Rossaint, Rolf ;
Czaplik, Michael .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2022, 36 (04) :1205-1212
[40]   Regional lung function assessment using electrical impedance tomography in COPD, PRISm, and normal spirometry subjects: insights into early diagnostic potential [J].
Li, Jiayi ;
Zhao, Zhanqi ;
He, Runze ;
Xie, Yuhong ;
Xu, Zhihao ;
Ni, Chunwei ;
Jiang, Ting ;
Ge, Huiqing .
BMC PULMONARY MEDICINE, 2025, 25 (01)