Implementing failure mode and effect analysis to improve the safety of volumetric modulated arc therapy for total body irradiation

被引:4
作者
Gray, Tara [1 ]
Antolak, Alexander [1 ]
Ahmed, Saeed [1 ]
Magnelli, Anthony [1 ]
Lu, Lan [1 ]
Cho, Young-Bin [1 ]
Qi, Peng [1 ]
Xia, Ping [1 ]
Cherian, Sheen [1 ]
Murphy, Erin [1 ]
Kilic, Sarah [1 ]
Zura, Kim [1 ]
Rankin, Emily [1 ]
Schulte, Jenna [1 ]
Kovacs, Tiffany [1 ]
Guo, Bingqi [1 ]
机构
[1] Cleveland Clin Fdn, Dept Radiat Oncol, CA Bldg, 10201 Carnegie Ave, Cleveland, OH 44106 USA
关键词
VMAT-TBI; FMEA; Quality Control; EFFECTS ANALYSIS FMEA; RADIATION-THERAPY; RANDOMIZED-TRIAL; MARROW; RISK; RADIOTHERAPY; BUSULFAN; LEUKEMIA; WORKFLOW; VMAT;
D O I
10.1002/mp.16466
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeVolumetric-modulated arc therapy for total body irradiation (VMAT-TBI) is a novel radiotherapy technique that has been implemented at our institution. The purpose of this work is to investigate possible failure modes (FMs) in the treatment process and to develop a quality control (QC) program for VMAT-TBI following TG-100 guidelines. MethodsWe formed a multidisciplinary team to map out the complete treatment process of VMAT-TBI following the AAPM TG-100 guidelines. This process map gives a visual representation of the VMAT-TBI workflow from the CT simulation, image processing, contouring, treatment planning, to treatment delivery. From the process map, potential FMs were identified. The occurrence (O), detectability (D), and severity of impact (S) of each FM were assigned according to scoring criteria (1-10) by the multidisciplinary team. A risk priority number (RPN) was calculated from average O, S, and D of each FM (RPN = O x S x D). High risk FMs were identified as 20% of the FMs having the highest RPN scores. After the FMEA analysis, fault-tree analysis (FTA) was performed for each major step of the treatment process to determine the effects of potential failures to the treatment outcome. Effective QC methods were identified to prevent the high risk failures and to improve the safety of the VMAT-TBI program. ResultsWe identified a total of 55 sub-processes and 128 FMs from the VMAT-TBI workflow. The top five high-risk FMs were: (1) Prescription and/or OAR constraints changed during planning and not communicated to the planner, (2) Patient moves or breathes too heavily during the upper body CT scan (3) Patient moves during the lower body CT scan, (4) Treatment planning system not calculating total body DVH metrics correctly for TBI, (5) Improper optimization criteria used or not sufficient optimization, resulting in suboptimal dose coverage, OAR sparing or excessive hotspots during treatment planning. Two FMs have average severity scores >= 8: Incorrect PTV subdivision/isocenter placement and Prescription and/or OAR constraints changed during planning and not communicated to the planner. Quality assurance and QC interventions including staff training, standard operating procedures, and quality checklists were implemented based on the FMEA and FTA. ConclusionFM and effect analysis was performed to identify high-risk FMs of our VMAT-TBI program. FMEA and FTA were effective in identifying potential FMs and determining the best quality management (QM) measures to implement in the VMAT-TBI program.
引用
收藏
页码:4092 / 4104
页数:13
相关论文
共 38 条
  • [1] Long-term follow-up of a randomized trial comparing the combination of cyclophosphamide with total body irradiation or busulfan as conditioning regimen for patients receiving HLA-identical marrow grafts for acute myeloblastic leukemia in first complete remission
    Blaise, D
    Maraninchi, D
    Michallet, M
    Reiffers, J
    Jouet, JP
    Milpied, N
    Devergie, A
    Attal, M
    Sotto, JJ
    Kuentz, M
    Ifrah, N
    Duriac, C
    Bordigoni, P
    Gratecos, N
    Guilhot, F
    Guyotat, D
    Gluckman, E
    Vernant, JP
    [J]. BLOOD, 2001, 97 (11) : 3669 - 3671
  • [2] Application of failure mode and effects analysis to optimization of linac quality controls protocol
    Bonfantini, Francesca
    Giandini, Tommaso
    Meroni, Silvia
    Cavallo, Anna
    Stucchi, Claudio
    Carrara, Mauro
    Mongioj, Valeria
    Veronese, Ivan
    Pignoli, Emanuele
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : 2541 - 2555
  • [3] Application of failure mode and effects analysis (FMEA) to pretreatment phases in tomotherapy
    Broggi, Sara
    Cantone, Marie Claire
    Chiara, Anna
    Di Muzio, Nadia
    Longobardi, Barbara
    Mangili, Paola
    Veronese, Ivan
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (05): : 265 - 277
  • [4] Randomized trial of busulfan vs total body irradiation containing conditioning regimens for children with acute lymphoblastic leukemia:: A Pediatric Blood and Marrow Transplant Consortium study
    Bunin, N
    Aplenc, R
    Kamani, N
    Shaw, K
    Cnaan, A
    Simms, S
    [J]. BONE MARROW TRANSPLANTATION, 2003, 32 (06) : 543 - 548
  • [5] Application of failure mode and effects analysis to treatment planning in scanned proton beam radiotherapy
    Cantone, Marie Claire
    Ciocca, Mario
    Dionisi, Francesco
    Fossati, Piero
    Lorentini, Stefano
    Krengli, Marco
    Molinelli, Silvia
    Orecchia, Roberto
    Schwarz, Marco
    Veronese, Ivan
    Vitolo, Viviana
    [J]. RADIATION ONCOLOGY, 2013, 8
  • [6] Total body irradiation and pneumonitis risk: a review of outcomes
    Carruthers, SA
    Wallington, M
    [J]. BRITISH JOURNAL OF CANCER, 2004, 90 (11) : 2080 - 2084
  • [7] A streamlined failure mode and effects analysis
    Ford, Eric C.
    Smith, Koren
    Terezakis, Stephanie
    Croog, Victoria
    Gollamudi, Smitha
    Gage, Irene
    Keck, Jordie
    DeWeese, Theodore
    Sibley, Greg
    [J]. MEDICAL PHYSICS, 2014, 41 (06)
  • [8] Total Body Irradiation (TBI) using Helical Tomotherapy in children and young adults undergoing stem cell transplantation
    Gruen, Arne
    Ebell, Wolfram
    Wlodarczyk, Waldemar
    Neumann, Oliver
    Kuehl, Joern Sven
    Stromberger, Carmen
    Budach, Volker
    Marnitz, Simone
    [J]. RADIATION ONCOLOGY, 2013, 8
  • [9] Image-guided volumetric-modulated arc therapy of total body irradiation: An efficient workflow from simulation to delivery
    Guo, Bingqi
    Sheen, Cherian
    Murphy, Erin
    Magnelli, Anthony
    Lu, Lan
    Cho, YoungBin
    Qi, Peng
    Majhail, Navneet S.
    Xia, Ping
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (10): : 169 - 177
  • [10] A method for evaluating quality assurance needs in radiation therapy
    Huq, M. Saiful
    Fraass, Benedick A.
    Dunscombe, Peter B.
    Gibbons, John P., Jr.
    Ibbott, Geoffrey S.
    Medin, Paul M.
    Mundt, Arno
    Mutic, Sassa
    Palta, Jatinder R.
    Thomadsen, Bruce R.
    Williamson, Jeffrey F.
    Yorke, Ellen D.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01): : S170 - S173