Design and Analysis of Multiple Inlet-Multiple Outlet Piezoelectric Actuated Valveless Micropump for Micro Drug Delivery Application

被引:2
作者
Gayatri, M. K. S. L. [1 ]
Guha, Koushik [2 ]
Sateesh, Jasti [3 ]
机构
[1] Natl Inst Technol, Dept Mech Engn, Silchar 788010, Assam, India
[2] Natl Inst Technol, Dept Elect & Commun Engn, Silchar 788010, Assam, India
[3] Velagapudi Ramakrishna Siddhartha Engn Coll, Dept Elect & Commun Engn, Vijayawada, Andhra Pradesh, India
关键词
Valveless micropump; Piezoelectric actuator; Back pressure; Bi-directional pumping; Micromixer; ELECTROSTATIC MICROPUMP; FABRICATION; SIMULATION; SYSTEMS; MEMS; MICROFLUIDICS; PROGRESS; DEVICES;
D O I
10.1007/s40313-022-00961-8
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper deals with the design of multiple inlet- multiple outlet piezoelectric valveless micropump intending for applications in the medical field such as drug delivery, lab-on-chip devices, smart surgical tools, blood pressure sensor and glucose sensors. An extensive analysis is carried out using benchmark FEM tool COMSOL multiphysics to predict the accurate behavior of the micropump under different condition. The comparative analysis among different types of configuration has been performed and the best one among these is listed and explained with the help of parameters such as outlet flow velocity, flow rate and back pressure. It is evident from the results that 2-inlet 1-outlet piezoelectric valveless micropump seems to efficient as its maximum flow rate is reported as 35.85 mu l/min which is greater as compared to 2-inlet 2-outlet micropump whose maximum flow rate is 11.6 mu l/min. The flow is considered as laminar flow owing to Reynolds number is less than 2000. Different studies are conducted by changing the parameters which reveals that the flow rate is affected with the change in average inlet flow velocity of working fluid, diameter of the micropump, length of diffuser/nozzle. The study will help researches, scientists and medical personnel to establish a precise and controlled drug delivery using micropump in biomedical applications.
引用
收藏
页码:429 / 442
页数:14
相关论文
共 48 条
  • [11] Dynamic simulation of an electrostatic micropump with pull-in and hysteresis phenomena
    Francais, O
    Dufour, I
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 1998, 70 (1-2) : 56 - 60
  • [12] Microfluidics - a review
    Gravesen, Peter
    Branebjerg, Jens
    Jensen, Ole Sondergard
    [J]. Journal of Micromechanics and Microengineering, 1993, 3 (04) : 168 - 182
  • [13] Electronic MEMS for triggered delivery
    Grayson, ACR
    Shawgo, RS
    Li, YW
    Cima, MJ
    [J]. ADVANCED DRUG DELIVERY REVIEWS, 2004, 56 (02) : 173 - 184
  • [14] A ferrofluidic magnetic micropump
    Hatch, A
    Kamholz, AE
    Holman, G
    Yager, P
    Böhringer, KF
    [J]. JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2001, 10 (02) : 215 - 221
  • [15] Osmotic micropumps for drug delivery
    Herrlich, Simon
    Spieth, Sven
    Messner, Stephan
    Zengerle, Roland
    [J]. ADVANCED DRUG DELIVERY REVIEWS, 2012, 64 (14) : 1617 - 1627
  • [16] Designing Biomaterial Platforms for Cardiac Tissue and Disease Modeling
    House, Andrew
    Atalla, Iren
    Lee, Eun Jung
    Guvendiren, Murat
    [J]. ADVANCED NANOBIOMED RESEARCH, 2021, 1 (01):
  • [17] Peristaltic piezoelectric micropump system for biomedical applications
    Jang, Ling-Sheng
    Kan, Wai-Hong
    [J]. BIOMEDICAL MICRODEVICES, 2007, 9 (04) : 619 - 626
  • [18] Fabrication and test of a thermopneumatic micropump with a corrugated p plus diaphragm
    Jeong, OC
    Yang, SS
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2000, 83 (1-3) : 249 - 255
  • [19] [阚君武 Kan Junwu], 2004, [生物医学工程学杂志, Journal of Biomedical Engineering], V21, P297
  • [20] Design and test of a high-performance piezoelectric micropump for drug delivery
    Kan, JW
    Yang, ZG
    Peng, TJ
    Cheng, GM
    Wu, B
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2005, 121 (01) : 156 - 161