Estimation of radiation dose associated with bone SPECT/CT and establishing local diagnostic reference levels using size-specific dose estimate

被引:3
|
作者
Alrehily, Faisal [1 ,2 ]
Alshamrani, Abdullah F. [1 ]
机构
[1] Taibah Univ, Coll Appl Med Sci, Dept Diagnost Radiol Technol, Madinah, Saudi Arabia
[2] Taibah Univ, Coll Appl Med Sci, Bldg 115,Prince Naif Rd, Madinah 42353, Saudi Arabia
关键词
SPECT; CT; CTDIvol; DLP; SSDE; LDRL; Effective dose; NATIONAL-SURVEY; CT; SSDE;
D O I
10.1016/j.jrras.2023.100527
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Single-photon emission computed tomography (SPECT) and X-ray computed tomography (CT) is used to di-agnose and monitor the skeletal system. However, due to the combination of injected activity and CT radiation dose, its usage is linked to an increase in patient radiation exposure. This study aims to estimate the radiation dose received by patients while undergoing bone SPECT/CT and to establish its local diagnostic reference level (LDRL). In this retrospective study, data from 30 patients who had undergone bone scans were obtained from two hospital imaging devices. The effective doses (E) of the radiopharmaceutical agent and CT components were calculated. Furthermore, the patients' anteroposterior thickness at the midline and lateral widths were measured from their CT images to determine the size-specific dose estimate (SSDE). The mean age of the patients was 53.53 +/- 16 years. They were administered an injection activity of 740 MBq of 99mTc-Methyl diphosphonate for SPECT, yielding an E of 2.78 mSv. The mean CTDIvol, DLP, SSDE, and E values for the CT component were 8.76 +/- 3.25 mGy, 350 +/- 138 mGy cm, 13 +/- 4 mGy, and 4.90 +/- 1.94 mSv, respectively. Patients received a combined dose of 7.68 mSv from the radiopharmaceuticals and CT. The proposed LDRL for SPECT was comparable to international DRLs, but the CT LDRL was slightly higher. An SSDE-based LDRL is proposed. Moreover, the overall patient dose can be reduced even further by optimizing the injected activity and CT imaging factors.
引用
收藏
页数:4
相关论文
共 32 条
  • [31] The radiation dose from whole body hybrid imaging (PET-CT) procedures in Sri Lanka: The initial approach towards establishing national diagnostic reference level
    Lakmal, Tharange
    Lamabadusuriya, Manuja
    Satharasinghe, Duminda
    Thilakshika, Rasuni
    Pallewatte, Aruna
    Jeyasugiththan, Jeyasingam
    RADIATION PHYSICS AND CHEMISTRY, 2024, 215
  • [32] Radiation dose assessment and establishment of diagnostic reference levels for MDP Tc-99m bone scans: A multi-hospital study in Saudi Arabia
    Alhujaili, Sultan Fahad
    Sulieman, A.
    Alrowaili, Ziyad Awadh
    Alshabibi, Abdulaziz Saad A.
    Alafer, Feras
    Alqahtani, Mansour M.
    Salah, Hasan
    Alanezi, Abdullah J.
    Alrashidi, Abdullah S.
    Abuzaid, Mohamed
    APPLIED RADIATION AND ISOTOPES, 2025, 218