Estimation of radiation dose associated with bone SPECT/CT and establishing local diagnostic reference levels using size-specific dose estimate

被引:3
|
作者
Alrehily, Faisal [1 ,2 ]
Alshamrani, Abdullah F. [1 ]
机构
[1] Taibah Univ, Coll Appl Med Sci, Dept Diagnost Radiol Technol, Madinah, Saudi Arabia
[2] Taibah Univ, Coll Appl Med Sci, Bldg 115,Prince Naif Rd, Madinah 42353, Saudi Arabia
关键词
SPECT; CT; CTDIvol; DLP; SSDE; LDRL; Effective dose; NATIONAL-SURVEY; CT; SSDE;
D O I
10.1016/j.jrras.2023.100527
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Single-photon emission computed tomography (SPECT) and X-ray computed tomography (CT) is used to di-agnose and monitor the skeletal system. However, due to the combination of injected activity and CT radiation dose, its usage is linked to an increase in patient radiation exposure. This study aims to estimate the radiation dose received by patients while undergoing bone SPECT/CT and to establish its local diagnostic reference level (LDRL). In this retrospective study, data from 30 patients who had undergone bone scans were obtained from two hospital imaging devices. The effective doses (E) of the radiopharmaceutical agent and CT components were calculated. Furthermore, the patients' anteroposterior thickness at the midline and lateral widths were measured from their CT images to determine the size-specific dose estimate (SSDE). The mean age of the patients was 53.53 +/- 16 years. They were administered an injection activity of 740 MBq of 99mTc-Methyl diphosphonate for SPECT, yielding an E of 2.78 mSv. The mean CTDIvol, DLP, SSDE, and E values for the CT component were 8.76 +/- 3.25 mGy, 350 +/- 138 mGy cm, 13 +/- 4 mGy, and 4.90 +/- 1.94 mSv, respectively. Patients received a combined dose of 7.68 mSv from the radiopharmaceuticals and CT. The proposed LDRL for SPECT was comparable to international DRLs, but the CT LDRL was slightly higher. An SSDE-based LDRL is proposed. Moreover, the overall patient dose can be reduced even further by optimizing the injected activity and CT imaging factors.
引用
收藏
页数:4
相关论文
共 32 条
  • [1] Establishing local diagnostic reference levels for computed tomography examinations using size-specific dose estimates
    Alrehily, Faisal A.
    Alqahtani, Khalid S.
    Aljohani, Marwan H.
    Alharbi, Nawaf S.
    Alharbi, Reyan M.
    Abdulaal, Osamah M.
    Alshamrani, Abdullah F.
    Alsaedi, Abdulrahman S.
    Al-Murshedi, Sadeq H.
    Alhazmi, Fahad H.
    SAUDI MEDICAL JOURNAL, 2023, 44 (08) : 761 - 766
  • [2] A new approach to dose reference levels in pediatric CT: Age and size-specific dose estimation
    Rawashdeh, Mohammad
    Saade, Charbel
    Mousa, Dana S. Al
    Abdelrahman, Mostafa
    Kumar, Praveen
    McEntee, Mark
    RADIATION PHYSICS AND CHEMISTRY, 2023, 205
  • [3] DEVELOPMENT OF RADIATION DOSE CALCULATION SOFTWARE USING THE SIZE-SPECIFIC DOSE ESTIMATE
    Shohji, Tomokazu
    Tachibana, Atsushi
    Higuchi, Sousuke
    Nakata, Norio
    Hayashi, Daichi
    Katoh, Yo
    RADIATION PROTECTION DOSIMETRY, 2018, 182 (03) : 359 - 369
  • [4] Tailoring CT Dose to Patient Size: Implementation of the Updated 2017 ACR Size-specific Diagnostic Reference Levels
    Klosterkemper, Yan
    Appel, Elisabeth
    Thomas, Christoph
    Bethge, Oliver T.
    Aissa, Joel
    Kroepil, Patric
    Antoch, Gerald
    Boos, Johannes
    ACADEMIC RADIOLOGY, 2018, 25 (12) : 1624 - 1631
  • [5] Radiation dose of head and abdomen-pelvis computed tomography examinations using size-specific dose estimate
    Ahmadifard, M.
    Bakhshandeh, M.
    Mohammadbeigi, A.
    Khoshgard, K.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2022, 20 (01): : 185 - 189
  • [6] SIZE-specific dose estimate for lower-limb CT
    Kobayashi, Masanao
    Nishihara, Yusei
    Haba, Tomonobu
    Matsunaga, Yuta
    Asada, Yasuki
    Kobayashi, Shigeki
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (04) : 1183 - 1191
  • [7] SIZE-specific dose estimate for lower-limb CT
    Masanao Kobayashi
    Yusei Nishihara
    Tomonobu Haba
    Yuta Matsunaga
    Yasuki Asada
    Shigeki Kobayashi
    Physical and Engineering Sciences in Medicine, 2022, 45 : 1183 - 1191
  • [8] Introduction and evaluation of size-specific DLP for radiation dose estimation in CT examinations
    Dendo, Yutaka
    Abe, Keisuke
    Onodera, Shu
    Kayano, Shingo
    Ota, Hideki
    Takase, Kei
    JOURNAL OF RADIOLOGICAL PROTECTION, 2024, 44 (04)
  • [9] Estimation of size-specific dose estimate (SSDE) of CT scans using an effective diameter electron density
    Almuqbil, Khuloud
    Abuhaimed, Abdullah
    Abanomy, Ahmed
    Alashban, Yazeed
    Alhawas, Abdulrahman
    Martin, Colin J.
    EUROPEAN JOURNAL OF RADIOLOGY, 2024, 172
  • [10] Evaluation of direct point dose estimation based on the distribution of the size-specific dose estimate
    Anam, Choirul
    Sutanto, Heri
    Amilia, Riska
    Marini, Rini
    Barokah, Sinta Nur
    Osman, Noor Diyana
    Dougherty, Geoff
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (04) : 1525 - 1535